TR100M SERIES 100 WATT MEDICAL SWITCH ADAPTER Features Universal Input Range 90~264Vac High Efficiency up to 89% Class II No Load Input Power Consumption < 300mW Approval IEC/EN/UL 60601-1 2 MOPP Approval EN55011 and CISPR/FCC Class B Operating Altitude 3000m Continuous Short Circuit Protection Over Voltage Protection RIPPLE & VOLTAGE LINE LOAD %EFF. MODEL OUTPUT OUTPUT NOISE ACCURACY REGULATION REGULATION (Typ.) NUMBER VOLTAGE CURRENT NOTE1 NOTE2 NOTE3 NOTE4 NOTE5 TR100M120 12 V 8.34 A 120mV 2% 1% 4% 88% TR100M150 15 V 6.67 A 150mV 2% 1% 3% 88% TR100M180 18 V 5.56 A 180mV 2% 1% 2% 88% 190mV TR100M190 19 V 5.27 A 2% 1% 2% 88% 200mV TR100M200 20 V 5.0 A 2% 1% 2% 88% 240mV TR100M240 24 V 4.17 A 2% 1% 2% 88% TR100M480 48 V 2.1 A 480mV 2% 1% 2% 89% Note: 1. Add a 0.1uF ceramic capacitor and a 10uF E.L. capacitor to output for ripple & noise measuring 20MHz BW. 2. Voltage accuracy is set at 60% full load. 3. Line regulation is measured from 100V to 240V with full load. ac ac 4. Load regulation measured from 60% to 100% full load and from 60% to 20% full load (60%40% full load). 5. Typical efficiency at 230V and 75% full load at 25. ac PART NUMBER Series Output Voltage DC Plug Type Cable Type Cable Length TR100M XXX -XX E XX 12012V 150 15V 180 11720mm with Ferrite Core 18V 100W 190 E 121220mm with Ferrite Core Medical 19V See Page 6 UL1185 with OVP Adapter 1800mm with Ferrite Core 20020V 13 24024V See page 6 for restrictions 480 48V Part Number Example: TR100M120-11E36, 100W, Class II, 12V Output, DC Jack Type, Cable Length 1500mm with Ferrite Core dc www.cincon.com 1 REVISION: V17 TR100M Series TECHNICAL SPECIFICATIONS (All specifications are typical at nominal input, full load at 25 unless otherwise noted.) ABSOLUTE MAXIMUM RATINGS PARAMETER NOTES and CONDITIONS Device Min. Typ. Max. Units 90 264 V ac Input Voltage All 120 370 V dc Operating Temperature See Derating Curve All -20 70 Storage Temperature All -20 85 Input/Output Isolation Voltage 1 minute All 4800 V ac Operating Altitude All 3000 m INPUT CHARACTERISTICS PARAMETER NOTES and CONDITIONS Device Min. Typ. Max. Units Operating Voltage Range All 100 240 V ac Input Frequency Range All 47 63 Hz Maximum Input Current 100% Load, V =100V All 1.5 A in ac Leakage Current (Touch) All 100 uA Under Voltage Protection All 65 70 V ac Power Factor 230V /50Hz Full load All 0.9 ac V =240V , Cold start at 25. Inrush Current in ac All 100 A OUTPUT CHARACTERISTICS PARAMETER NOTES and CONDITIONS Device Min. Typ. Max. Units TR100M120 11.76 12 12.24 TR100M150 14.7 15 15.3 TR100M180 17.64 18 18.36 V =115V and 230V , I =60% Full load in ac ac o Output Voltage Set Point TR100M190 18.62 19 19.38 V dc T =25 c TR100M200 19.6 20 20.4 TR100M240 23.52 24 24.48 TR100M480 47.04 48 48.96 TR100M120 0 8.34 TR100M150 0 6.67 TR100M180 0 5.56 V =115V and 230V ,T =25 Operating Output Current Range in ac ac c TR100M190 0 5.27 A TR100M200 0 5 TR100M240 0 4.17 TR100M480 0 2.1 Holdup Time V =115V All 16 ms in ac Output Voltage Regulation TR100M120 4.0 TR100M150 3.0 TR100M180 2.0 Load Regulation 60%40% Full load change % TR100M190 2.0 TR100M200 2.0 TR100M240 2.0 TR100M480 2.0 Line Regulation V =High line to low line, full load All 1.0 % in www.cincon.com 2 REVISION: V17