V19 TR15RAM SERIES 15W Medical SWITCHING ADAPTER Features * Universal Input Range 90~264VAC * Meets EN60601-1 and EN55011 Class B * Continuous Short Circuit Protection * Interchangeable AC Plugs * Over Voltage Protection * Efficiency & Standby Power Meet Level V (Output Cable Length1800mm) * Meets 2MOPP Ordering information TR15RAMXXX - XX E XX - XX -BK Model No. DC Plug Type DC Cable Length and Type Color of Overmold Case 01: 720mm BE: Blue 02: 1220mm GY: Gray 03: 1800mm RD: Red 11: 7200mm with Ferrite Core PE: Purple 12: 1220mm with Ferrite Core OR: Orange 13: 1800mm with Ferrite Core * 18AWG / UL1185 VOLTAGE LINE LOAD OUTPUT OUTPUT RIPPLE & NOISE % EFF. (Typ.) MODEL ACCURACY REGULATION REGULATION VOLTAGE CURRENT NOTE 1 NOTE 5 NOTE 2 NOTE 3 NOTE 4 TR15RAM050 5 V 2.0 A 1% 3% 1% 4% 73% TR15RAM120 82% 12 V 1.1 A 1% 2% 1% 2% TR15RAM150 15 V 1.0 A 1% 2% 1% 2% 81% TR15RAM240 82% 24 V 625 mA 1% 2% 1% 2% Specifications are subject to change without notice. 1 V19 Specifications TR15RAM Series Derating Curve INPUT SPECIFICATIONS: 100% Voltage ....................................................... 90~264Vac Frequency ...... 47 to 63Hz 75% Inrush Current ...................... 50A max. 240Vac 20LFM(Natural Convection) 50% Leakage Current . 0.1mA max. 25% OUTPUT SPECIFICATIONS: Hold-up Time 10ms typ. 115Vac 0% Temperature Coefficient ............................................. 0.05%/C 0 10 20 30 40 50 60 70 o Short Circuit Protection . Continuous (Auto Recovery) Ambient Temperature( C) Over Voltage Protection ..TVS Component to Clamp GENERAL CHARACTERISTICS: NOTE: Isolation ..... Input to output = 5,656VDC 1. Add a 0.1uF ceramic capacitor and a 10uF E.L. capacitor to output for Switching Frequency . 95KHz typical Ripple & Noise measuring 20MHz BW. Operating Temperature .................................. 0 - 70(see derating curve) 2. Voltage setpoint at 60% full load. 3. Line regulation measured from 100Vac to 240Vac full load. Storage Temperature ..................... -20 - 85 Cooling . Natural Convection 4. Load regulation measured from 60% to full load and from 60% to 20% MTBF . MIL-HDBK-217F, GB, at 25/115VAC 200Khrs min. load (60% +/- 40% load). Humidity .................................................... 93% RH max. Non condensing 5. Typical efficiency at 230VAC and full load at 25 . 6. Various TR Series adapters are PSE certified. PSE certification alone Altitude ...... 3000m is not sufficient for importation into Japan. A valid PSE mark must Dimensions .... 3.173x1.885x1.704 inches (80.60x47.90x43.30mm) Weight 130g(0.29 Pounds) contain the name of the importer as shown in the example below. If PSE mark is required, the name of the registered importer must be supplied SAFETY AND EMC: Emission and Immunity (Ed. 4.0) ..... EN55011 Class B, FCC47 CFR Part 15 to Cincon on order placement. Product labels will not contain PSE mark if importer name is not supplied. Consult factory or local EN60601-1-2, EN61000-3-2, 3, IEC61000-4-2, 3, 4, 5, 6, 8, 11 Safety (Ed. 3.1) ....................................................... IEC60601-1:2005+A1 representative for details. EN60601-1:2006+A11:2011+A1+A12, UL ANSI/AAMI ES60601-1: 2005 Mechanical Specification All Dimensions are in inches mm Tolerance:Inches:X.XXX0.02 Millimeters:X.XX0.5 DC Plug type: V+ ) V- DC Plug :Straight(5.52.1) L12mm 18AWG1800mm 3.173 80.60 70.866 1800.00 1.969 50.00 + - Typical at 25C, nominal line and 75% load, unless otherwise Specified 2 0.248 6.30 1.885 47.90 1.515 38.50 0.157 4.00 0.712 18.10 Max 0.629 16.00 0.874 22.20 0.005 0.13 1.704 43.30 0.669 17.00 0.031 0.70 TR30RAM/TR15RA-A 1.102 28.00 0.492 12.50 0.002 0.05 Rated Load(%) 0.727 18.47 0.002 0.05 1.102 28.00 0.748 19.00 0.019 0.50 0.665 16.90 0.011 0.30 0.343 8.70 0.019 0.50 1.275 32.40 1.137 28.90