TR220M SERIES 220 WATT MEDICAL SWITCH ADAPTER Features Universal Input Range 80~264Vac High Efficiency up to 94% Class I (TR220MA), Class II (TR220MB) No Load Input Power Consumption < 150mW Approval IEC/EN/UL 60601-1 2 MOPP Approval IEC/EN 60601-1-11 (TR220MB) Approval EN55011 and CISPR/FCC Class B Approval IP22 Operating Altitude 5000m Continuous Short Circuit Protection Over Voltage Protection Meets CoC Tier 2 and DOE Level VI RIPPLE & VOLTAGE LINE LOAD %EFF. MODEL OUTPUT OUTPUT NOISE ACCURACY REGULATION REGULATION (Typ.) NUMBER VOLTAGE CURRENT NOTE1 NOTE2 NOTE3 NOTE4 NOTE5 TR220MA120 12 V 16.67 A 120mV 2% 1% 4% 91% TR220MA240 24 V 9.17 A 200mV 2% 1% 3% 93% 200mV TR220MA360 36 V 6.11 A 2% 1% 2% 93% TR220MA480 48 V 4.58 A 200mV 2% 1% 2% 94% TR220MA560 56 V 3.93 A 200mV 2% 1% 2% 94% TR220MB120 12 V 16.67A 120mV 2% 1% 4% 91% TR220MB240 24 V 9.17 A 200mV 2% 1% 3% 93% TR220MB360 36 V 6.11 A 200mV 2% 1% 2% 93% TR220MB480 48 V 4.58 A 200mV 2% 1% 2% 94% 200mV TR220MB560 56 V 3.93 A 2% 1% 2% 94% Note: 1. Add a 0.1uF ceramic capacitor and a 10uF E.L. capacitor to output for ripple & noise measuring 20MHz BW. 2. Voltage accuracy is set at 60% full load. 3. Line regulation is measured from 100V to 240V with full load. ac ac 4. Load regulation measured from 60% to 100% full load and from 60% to 20% full load (60%40% full load). 5. Typical efficiency at 230V and 75% full load at 25. ac PART NUMBER Series Output Voltage DC Plug Type Cable Type Cable Length XXX -XX E XX TR220M X 120 12V 240 24 V 471 9 50mm with Ferrite Core A 220W Class I Medical BCla ss II 360 36V See Page 6 EUL2 464 with OVP 12 1 220mm with Ferrite Core Adapter 480 13 48V 1800mm with Ferrite Core 560 56V See page 6 restrictions Part Number Example: TR220MA240-1538E12, 220W, Class I, 24Vdc Output, Din Power Plug Molded Type, Cable Length 1220mm with Ferrite Core TR220MB360-11E13, 220W, Class II, 36V Output, DC Jack Type, Cable Length 1800mm with Ferrite Core dc www.cincon.com 1 REVISION: V10 TR220M Series TECHNICAL SPECIFICATIONS (All specifications are typical at nominal input, full load at 25 unless otherwise noted.) ABSOLUTE MAXIMUM RATINGS PARAMETER NOTES and CONDITIONS Device Min. Typ. Max. Units Input Voltage All 80 264 V ac Operating Temperature See Derating Curve All -30 70 Storage Temperature All -40 85 Operating Altitude All 5000 m INPUT CHARACTERISTICS PARAMETER NOTES and CONDITIONS Device Min. Typ. Max. Units Operating Voltage Range All 100 240 V ac Input Frequency Range All 47 63 Hz Maximum Input Current 100% Load, V =100V All 2.5 A in ac Leakage Current (Earth) All 300 uA Leakage Current (Touch) All 100 uA Under Voltage Protection All 50 60 65 V Power Factor 230V /50Hz Full load All 0.92 ac OUTPUT CHARACTERISTICS PARAMETER NOTES and CONDITIONS Device Min. Typ. Max. Units TR220MA/B120 11.76 12 12.24 TR220MA/B240 23.52 24 24.48 V =115V and 230V , I =60% Full load in ac ac o Output Voltage Set Point TR220MA/B360 35.28 36 36.72 V dc T =25 c TR220MA/B480 47.04 48 48.96 TR220MA/B560 54.88 56 57.12 TR220MA/B120 0 16.67 TR220MA/B240 0 9.17 V =115V and 230V ,T =25 Operating Output Current Range in ac ac c TR220MA/B360 0 6.11 A TR220MA/B480 0 4.58 TR220MA/B560 0 3.93 Holdup Time V =115V All 20 22 ms in ac Output Voltage Regulation TR220MA/B120 4 TR220MA/B240 3 Load Regulation 60%40% Full load change TR220MA/B360 2 % TR220MA/B480 2 TR220MA/B560 2 Line Regulation V =High line to low line All 1 % in TR220MA/B120 13.5 TR220MA/B240 30 Over Voltage Protection Latch Off TR220MA/B360 42 V dc TR220MA/B480 56 TR220MA/B560 59.2 TR220MA/B120 120 1. Add a 0.1uF ceramic capacitor and a TR220MA/B240 200 10uF aluminum electrolytic capacitor to output Output Ripple and Noise TR220MA/B360 200 mV 2. Oscilloscope is 20MHz band width TR220MA/B480 200 3. Ambient temperature=25 TR220MA/B560 200 www.cincon.com 2 REVISION: V10