M C E E T R S T Y I F S E T D N E T M O E ScopeMeter 120 and 190 Series incl. 225C and 215C Technical Data ScopeMeter 215C, Bus Health Test capability for 1,000 V CAT II and 600 V 225C and 190C Series: industrial buses (225C and 215C) CAT III safety certified Speed, performance High waveform resolution of Up to 1,000 V independently and analysis power 3000 datapoints per channel floating isolated inputs Frequency Spectrum using FFT 5000 count DMM and paperless For demanding applications, the analysis recorder built-in ScopeMeter 215C, 225C and 190C Series high-performance oscillo- Connect-and-View automatic Four hours rechargeable scopes offer specifications usually triggering, a full range of manual NiMH battery pack found on top-end bench instru- trigger modes plus external trig- ments. With up to 200 MHz band- ScopeMeter 120 Series: gering width, 2.5 GS/s real time sampling Three-in-one simplicity Digital Persistence for analyzing and a deep memory of 27,500 complex, dynamic signals like on The compact ScopeMeter 120 Series points per input, theyre ideal an analog oscilloscope is the rugged solution for industrial for engineers who need the full troubleshooting and installation Fast display update rate for capabilities of a high-performance applications. Its a truly integrated seeing dynamic behavior scope in a handheld, battery pow- test tool, with oscilloscope, multi- instantaneously ered instrument. meter and paperless recorder in Dual input - 200, 100 or Automatic capture and replay of one affordable, easy-to-use instru- 60 MHz bandwidth 100 screens ment. Find fast answers to problems Up to 2.5 GS/s real time 27,500 points and more per in machinery, instrumentation, sampling per input input record length using control and power systems. ScopeRecord mode A dual input 40 MHz or 20 MHz digital oscilloscope Two 5,000 counts true-rms digital multimeters Cursor measurements (Fluke 124, 125) A dual input TrendPlot recorder Connect-and-View trigger simplicity for hands-off operation Shielded test leads for oscillo- scope, resistance, continuity and capacitance measurements Full bandwidth, VPS40 10:1 40 MHz probe included standard with Fluke 124, 125 Up to seven hours battery operation 600 V CAT III safety certified Optically isolated RS-232 interface Rugged, compact case Bus Health test for industrial bus systems (Fluke 125) Power measurements and har- monics measurement (Fluke 125) G M A E N E A T M I S Y T O I L 9 A 0 0 U 1 QTechnical specifications 215C, 225C and 190C Series Horizontal Oscilloscope mode Pulse width triggering: Pulse width qualified by time. Vertical deflection Allows for triggering <t, >t, =t, t, where t is select- able in minimal steps of 0.01 div. or 50 nsec. Fluke 225C Fluke 215C Fluke 192C Fluke 199C Fluke 196C Time delay: One full screen of pre-trigger view or up Bandwidth 200 MHz 100 MHz 60 MHz to 100 screens (= 1200 divisions) of post-trigger delay Rise time 1.7 ns 3.5 ns 5.8 ns Dual slope triggering: Triggers on both rising and falling edges alike Bandwidth limiter: User selectable: 10 kHz, 20 MHz N-cycle triggering: Triggers on N-th occurrence of a or off trigger event N to be set in the range 2 to 99 Number of inputs: Two plus external trigger. Automatic capture of 100 screens All inputs references isolated from each other and ground. The instrument ALWAYS memorizes the last 100 Input coupling: AC or dc, with ground level indicator screens (no user setup required). When an anomaly Input sensitivity: 2 mV/div to 100 V/div occurs on screen, the REPLAY button can be pressed Normal/invert: On both input channels switched to review the full screen sequence over and over. separately Instrument can be set up for triggering on glitches or Variable attenuator: Variable gain on input channel A intermittent anomalies and will operate in baby-sit Input voltage: 1000 V CAT II, 600 V CAT III ratedsee mode capturing 100 events. general specifications for further details Replay: Manual or continuous replay. Displays the Vertical resolution: 8 bit captured 100 screens as a live animation, or Accuracy: (1.5 % of reading + 0.04 x range/div) under manual control. Each screen has date and Input impedance: 1 MW 1 % // 15 pF 2 pF time-stamp. Replay storage: Up to 2 sets of 100 screens each Fluke 225C Fluke 215C Fluke 192C can be saved for later recall and analysis Fluke 199C Fluke 196C Maximum 2.5 GS/s 1 GS/s 500 MS/s FFT Frequency spectrum analysis real-time sample rate Shows frequency content of oscilloscope waveform Number of 2 2 2 using Fast Fourier Transform digitizers Window: Automatic, Hamming, Hanning or None Time base 5 ns/div 10 ns/div Automatic window: Digitally re-samples acquired range to 5 s/div to 5 s/div waveform to get optimum frequency resolution in FFT Maximum record length: 3000 points per input in resultant Scope mode 27,500 points per input in ScopeRecord Vertical scale: Linear/logarithmic, in volts roll mode (5 ms/div to 2 min/div) Frequency axis: Logarithmic frequency range Accuracy: (0.01 % of reading + 1 pixel) automatically set as function of timebase range of Glitch capture: 50 nsec (5 sec/div to 1 min/div) oscilloscope Display and acquisition Waveform compare and pass/fail testing Display: 144 mm full-color LCD, with backlight Waveform compare: Provides storage and display Display modes: Input A, input B, dual, average, of a reference waveform for visual comparison with Replay newly acquired waveforms. Reference is derived from Visible screen width: 12 divisions an acquired waveform and can be modified in the Waveform mathematics: A + B, A - B, A * B, all ScopeMeter or externally using FlukeView Software. with user selectable scaling of resultant A versus B Pass/Fail Testing: In waveform compare mode, the (X - Y mode) Frequency spectrum using FFT analysis Color Scopemeter can be set up to store only matching Acquisition modes: Normal, auto, single shot, (Pass) or only non-matching (Fail) acquired wave- ScopeRecord roll, glitch capture, waveform forms in the replay memory bank for further analysis. compare, waveform compare with automatic Automatic scope measurements Pass / Fail testing, Bus Health test mode (225C and 215C only), Eyepattern Display of single ended or V dc, V ac rms, V ac+dc, Vpeak max, Vpeak min, Vpeak differential bus signal (Fluke 225C and 215C only). to peak, A ac, A dc, A ac+dc, frequency (Hz), risetime, falltime, power factor, Watts, VA, VA reactive, phase, Trigger and delay pulse width (pos/neg), dutycycle (pos/neg), temperature Source: Input A, input B, external trigger input. C, temperature F, dBV, dBm into 50 W and 600 W All input references isolated from each other and Vpwm ac, Vpwm ac+dc for measurement on pulse from ground width modulated motordrives and frequency inverters Modes: Automatic Connect-and-View, free run, single Cursor measurements shot, edge, delay, video, video line, selectable pulsewidth, dual slope, N-cycle Source: Input A, input B or the mathematical result Connect-and-View : Advanced automatic triggering trace (excluding A vs B curve) that recognizes signal patterns, automatically sets up Dual horizontal lines: Voltage at cursor 1 and 2, and continuously adjusts triggering, time base and voltage between cursors amplitude. Automatically displays stable waveforms Dual vertical lines: Time between cursors, 1/T of complex and dynamic signals like motor drive and between cursors (in Hz), voltage between markers, control signals. Can be switched off if desired. risetime with markers, falltime with markers Vrms Video triggering: NTSC, PAL, PAL+, SECAM. Includes between cursors, Watts between cursors field 1, field 2 and line select Single vertical line: Min-Max and average voltage at cursor position frequency and rms-value of individual frequency component in FFT result 2 Fluke Corporation ScopeMeter 215C, 225C and 190C Series, 120 Series