700P04 PRESSURE MODULE RANGE 15 PSID/G 100 kPa 1 bar 700P06 PRESSURE MODULE RANGE 100 PSIG 700 kPa 7 bar 700 Series Pressure Modules Instruction Sheet Low Differential High Gage yz1f.eps Figure 1. Introduction The Fluke 700 Series Pressure Modules allow you to measure pressure with the Fluke 700 Series Process Calibrators. Read this sheet before you use the pressure module. This sheet contains specifications and information about how to avoid damaging the pressure module. See the Calibrator Users Manual for operating instructions. The pressure module measures pressure using an internal microprocessor. It receives operating power from and sends digital information to the 700 Series calibrator. Gage pressure modules have one pressure fitting and measure pressure with respect to atmospheric pressure. Differential pressure modules have two pressure fittings and measure the difference between the applied pressure on the high fitting versus the low fitting. Figure 1 shows the two types. A differential pressure module functions like a gage module when the low fitting is open. Compatibility of Pressure Modules with Fluke Process Calibrators Fluke-701, 702, 741, 743, and 744 Documenting Process Calibrators: The Fluke-700P00 and P27 Pressure Modules are completely compatible with these calibrators. Fluke-716, 717, and 718 Pressure Calibrators: The Fluke-700P00 and P27 Pressure Modules will display pressure readings correctly on 716 Pressure Calibrators with firmware V1.2 or higher, or 717 and 718 Pressure Calibrators with firmware V1.3 or higher. To view the software revision, power-on the calibrator while holding down the MAX key. Read the REV number in the display. Contact your Fluke Service Center for a free update of calibrators with lower revision numbers. Reference PCN 4647. PN 942735 July 1994 Rev.12, 11/01 1994-2001 Fluke Corporation. All rights reserved. Printed in USA All product names are trademarks of their respective companiesFluke-725 Multifunction Process Calibrators: The Fluke-700P00 and Protecting Yourself from Pressure Releases P27 Pressure Modules will display correct pressure readings when To avoid a violent release of pressure in a pressurized system, shut off used with Fluke-725 Multifunction Process Calibrators at internal the isolation valve and slowly bleed off the pressure before you attach software revision 1.9 and later. To view the software revision, power-on or remove the pressure module from the pressure line. the calibrator while holding down the 0% key. Read the REV number in Avoiding Overpressure Damage the display. Contact your Fluke Service Center for a free update of Applying pressure in excess of the BURST PRESSURE specified on calibrators with lower revision numbers. Reference PCN 4578. the pressure module can destroy the pressure module. Burst pressure Box Contents is 3X full scale (2X for 700P09, 30X for 700P00). Pressure module, strap, 1/4 NPT to 1/4 ISO metric adapters, instruction sheet. 1 Table 1. Pressure Module Specifications GAGE OR ISOLATED OR REFERENCE STABILITY TEMP TOTAL UNCERTAINTY 2 RANGE 3 MODEL DIFFERENTIAL NONISOLATED UNCERTAINTY (1 YEAR) (0 to 50C) (% of Full Scale) 1 YEAR (23 3C) FLUKE-700P00 0 to 1.000 in H O Differential Nonisolated 0.300% 0.025% 0.025% 0.350% 2 0 to 0.249 kPa 0 to 0.002 bar FLUKE-700P01 0 to 10.00 in H O Differential Nonisolated 0.200% 0.050% 0.050% 0.300% 2 0 to 2.49 kPa 0 to 0.02 bar FLUKE-700P02 0 to 1.0000 psi Differential Nonisolated 0.150% 0.070% 0.080% 0.300% 0 to 6.8900 kPa 0 to 6.89 E2 bar FLUKE-700P03 0 to 5.0000 psi Differential Nonisolated 0.050% 0.020% 0.030% 0.100% 0 to 34.000 kPa 0 to 0.3400 bar FLUKE-700P04 0 to 15.000 psi Differential Nonisolated 0.025% 0.010% 0.015% 0.050% 0 to 100.00 kPa 0 to 1.0000 bar FLUKE-700P05 0 to 30.000 psi Gage Isolated 0.025% 0.010% 0.015% 0.050% 0 to 200.00 kPa 0 to 2.0000 bar FLUKE-700P06 0 to 100.00 psi Gage Isolated 0.025% 0.010% 0.015% 0.050% 0 to 700.00 kPa 0 to 7.0000 bar FLUKE-700P27 0 to 300.00 psi Gage Isolated 0.025% 0.010% 0.015% 0.050% 0 to 2000.0 kPa 0 to 20.000 bar FLUKE-700P07 0 to 500.00 psi Gage Isolated 0.025% 0.010% 0.015% 0.050% 0 to 3400.0 kPa 0 to 34.000 bar FLUKE-700P08 0 to 1000.0 psi Gage Isolated 0.025% 0.010% 0.015% 0.050% 0 to 7000.0 kPa 0 to 70.000 bar FLUKE-700P09 0 to 1500.0 psi Gage Isolated 0.025% 0.010% 0.015% 0.050% 0 to 10000.0 kPa 0 to 100.000 bar 1. Use of pressure zero function is required to achieve these specifications. Contact your Fluke Service Center for upgrade of your Fluke 701 or Fluke 702 V1.0, V1.1, or V1.2 calibrator. 2 2. Available pressure units (psi, kPa, bar, inHg, mmHg, inH 0, ftH 0., kg/cm , mmH 0, cmH 0) are determined by the calibrator being used. 2 2 2 2 3. Accuracy specifications apply for 0 to 100% of full scale, 0 to 50C. 1% FS typical, -10 to 0C. On 700P00 only, specification applies from 15 to 35C. 4. Maximum Non-Destructive Pressure: 3X maximum rated pressure, including common mode pressure. (700P09: 2X, 700P00: 30X) 5. Maximum Common Mode Pressure: 3X maximum rated pressure. (700P09: 2X, 700P00: 30X) 6. Specifications reflect a confidence interval of 95%.