Tunnel Series AAT-085-24-22-00-00
Air-to-Air Thermoelectric Assembly
The Tunnel Series Air-to-Air thermoelectric assembly is a
thermoelectric based air conditioner designed to temperature
control small chambers used in analytical and medical
diagnostic instruments. The unique design offers premium
fans pushing air across-high density heat sinks to minimize the
number of air flow paths required to operate. The design
utilizes custom thermoelectric modules to maximize cooling
capacity with a high coefficient of performance. Moisture
resistant insulation is used to keep condensation from
penetrating the TEM cavity. The units operate on DC and are
designed for an indoor lab use environment.
Laird Part Number: 387000899
Americas: +1.919.597.7300
Europe: +46.31.420530
FEATURES APPLICATIONS
Asia: +86.755.2714.1166
Compact form factor Analytical storage
ets.sales@lairdtech.com
compartment temperature
Improved sealing for moisture
www.lairdtech.com
protection control
Medical diagnostic chamber
Reliable solid-state operation
RoHS compliant refrigeration
SPECIFICATIONS
TEA Model AAT,085,24,22,00,00
Heat Transfer, Cold Side Air
Heat Transfer, Hot Side Air
Cooling Power 85 W @ T=0C and Ta=35C, Tolerance 10%
TEM Input Power
Voltage, Nominal 24 VDC
Current, Nominal 4.2 Amps @ T=0C
Fan Input Power
Hot side Fan Voltage, Nominal 24 VDC
Hot side Fan Current, Nominal 0.46 Amps
Cold side Fan Voltage, Nominal 24VDC
Cold side Fan Current, Nominal 0.33Amps
Dimension (L x W x H) 248 x 82.5 x 153.5 mm
Weight 2.5kg
Operating Temperature -10C to 50C
Packaging Individual cardboard box
Page 1 of 4
AAT-085-24-22-00-00
PERFORMANCE CURVES
TEA performance at Th = 35C
Qc vs dT
90
80
70
60
50
40
30
20
10
0
0 5 10 15 20 25 30
dT (C)
Page 2 of 4
Qc (W)