36W AC-DC Reliable Green Medical Adaptor G SM 36U series Users Manual Video + + ANSI/AAMI ES60601-1/-1-11 IEC60601-1/-1-11 TPTC004 File Name:GSM36U-SPEC 2021-12-1536W AC-DC Reliable Green Medical Adaptor G SM 36U series SPECIFICATION ORDER NO. GSM36U05-P1J GSM36U07-P1J GSM36U09-P1J GSM36U12-P1J GSM36U15-P1J GSM36U18-P1J GSM36U24-P1J GSM36U48-P1J GSM36U05 GSM36U07 GSM36U09 GSM36U12 GSM36U15 GSM36U18 GSM36U24 GSM36U48 SAFETY MODEL NO. DC VOLTAGE Note.2 5V 7.5V 9V 12V 15V 18V 24V 48V RATED CURRENT 4.5A 4.32A 4A 3A 2.4A 2A 1.5A 0.75A CURRENT RANGE 0 ~ 4.5A 0 ~ 4.32A 0 ~ 4A 0 ~ 3A 0 ~ 2.4A 0 ~ 2A 0 ~ 1.5A 0 ~ 0.75A RATED POWER (max.) 22.5W 32.4W 36W 36W 36W 36W 36W 36W OUTPUT 80mVp-p 80mVp-p 80mVp-p 120mVp-p 120mVp-p 150mVp-p 180mVp-p 240mVp-p RIPPLE & NOISE (max.) Note.3 VOLTAGE TOLERANCE Note.4 6.0% 5.0% 5.0% 3.0% 3.0% 3.0% 2.0% 2.0% LINE REGULATION Note.5 1.0% 1.0% 1.0% 1.0% 1.0% 1.0% 1.0% 1.0% LOAD REGULATION 6.0% 5.0% 5.0% 3.0% 3.0% 3.0% 2.0% 2.0% SETUP, RISE TIME Note.6 500ms, 30ms / 230VAC 1000ms, 30ms / 115VAC at full load HOLD UP TIME (Typ.) 16ms / 230VAC 10ms / 115VAC at full load VOLTAGE RANGE Note.7 80 ~ 264VAC 113 ~ 370VDC FREQUENCY RANGE 47 ~ 63Hz EFFICIENCY (Typ.) 80% 83% 84% 86% 87% 87% 87% 88% INPUT 0.9A / 115VAC 0.45A / 230VAC AC CURRENT (Typ.) INRUSH CURRENT (Typ.) 55A / 230VAC 30A / 115VAC Touch current < 50 A/264VAC LEAKAGE CURRENT(max.) 105 ~ 170% rated output power OVERLOAD Protection type : Hiccup mode, recovers automatically after fault condition is removed PROTECTION 5.25 ~ 7.5V 7.88 ~ 10.5V 9.45 ~ 13V 12.6 ~ 17.2V 15.75 ~ 20.25V 18.9 ~ 25.2V 25.2 ~ 32.4V 50.4 ~ 64.8V OVER VOLTAGE Protection type : Shut down o/p voltage, re-power on to recover WORKING TEMP. -25 ~ +60 (Refer toDerating Curv) WORKING HUMIDITY 20% ~ 90% RH non-condensing ENVIRONMENT STORAGE TEMP., HUMIDITY -40 ~ +85, 10 ~ 95% RH non-condensing TEMP. COEFFICIENT 0.03% / (0~40) VIBRATION 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes SAFETY STANDARDS ANSI/AAMI ES60601-1, ES60601-1-11, EAC TP TC 004 approved ISOLATION LEVEL Primary-Secondary: 2xMOPP SAFETY & WITHSTAND VOLTAGE I/P-O/P:4KVAC EMC ISOLATION RESISTANCE I/P-O/P:100M Ohms / 500VDC / 25/ 70% RH (Note. 8) Standard Test Level / Note Parameter EMC EMISSION Conducted emission FCC PART 15 / CISPR22 Class B Radiated emission FCC PART 15 / CISPR22 Class B MTBF 657.4K hrs min. MIL-HDBK-217F(25) OTHERS DIMENSION 79*54*33mm (L*W*H) 215g 60pcs / 13.9Kg / CARTON PACKING See page 4~5 Other type available by customer requested PLUG CONNECTOR CABLE See page 4~5 Other type available by customer requested 1. All parameters are specified at 115VAC input, rated load, 25 70% RH ambient. 2. DC voltage: The output voltage set at point measure by plug terminal & 50% load. 3. Ripple & noise are measured at 20MHz by using a 12 twisted pair terminated with a 0.1 f & 47 f capacitor. 4. Tolerance: includes set up tolerance, line regulation, load regulation. 5. Line regulation is measured from low line to high line at rated load. NOTE 6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time. 7. Derating may be needed under low input voltage. Please check the derating curve for more details. 8. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on