WWW.PHIHONG.COM 10W Medical Grade Adapter with Interchangeable Plugs* *Plugs sold separately Features DoE Level VI & CoC V5 Tier 2 LPS compliant Field Changeable AC Plugs IP21rating 2XMOPP Isolation OVP, OCP, SCP, OTP, brown-in and brown-out protections Class II Suitable for BF applications with system consideration Applications Home healthcare equipment ECG machine Patient monitor Respirator Portable medical devices Approvals and Certifications ANSI/AAMI ES 60601-1 IEC 60601-1:2005 + A1:2012 (Edition 3.1) CAN/CSA C22.2 No.60601-1 TH IEC60601-1-11:2015 IEC60601-1-2:2014(4 Edition) IEC62366:2007 + A1:2014 IEC60601-1-6:2010 + AMD1:2013 Mechanical Characteristics Length: 71.7mm (2.82in) Height: 40mm (1.57in) Width: 50mm (1.97in) Output Specifications 1 Load Ripple Regulation DC Output Model Voltage Min. Max. P-P (max.) Line & Load PMA10R-050AW-R 5V 0A 2A 100mV 5% Notes: 1. Test by 20Mhz bandwidth mode hi resolution with 10uF and 0.1uF capacitor 2. The characteristics defined are at ambient temperature of 25C unless otherwise specified Phihong is not responsible for any error, and reserves the right to make changes without notice. Please visit our website at www.phihong.com for the most up-to-date specifications and contact information. Revised 12/26/2019 2 PMA10R-050AW-R Characteristics WWW.PHIHONG.COM Input: AC Input Voltage Rating Over-Voltage Protection 115 to 230V AC 9V maximum AC Input Voltage Range Short-Circuit Protection 90 to 264V AC The power supply shall be protected when operating any output in overload condition, or when operating any output in a short. The power supply shall be AC Input Frequency self-recovering when the fault condition is removed 47 to 63Hz circuit condition. Input Current Over-Current Protection 0.3A max at 100VAC The power supply will be protected when output power at 2.3A(RMS) of all rated dc output Leakage Current <100uA at 264VAC/50Hz Over-Temperature Protection Over-temperature protection sensor Inrush Current 60A maximum at 240V AC Dielectric Withstand (Hi-pot) Test (Cold start & at input rated voltage +/-10%) Pri. to Sec. >4000VAC 10mA for 1 Minute Output: DC Output Connector Efficiency DC plug type 5.5 x 2.1 x 10mm (fork & groove) DOE Level VI DC plug pin assignment: CoC V5 tier 2 Inner (+V), Outer GND(-) Environmental: Medical AC Input Clips (sold separately): Temperature Operation 5 to +40C Part no: Description: Non-operation -25 to +70C RPA-MAB1W US white Relative Humidity 93% RH Max RPB- MAB1W Brazil white Emissions RPE- MAB1W Euro white Complies with CISPR 11 Class B RPH- MAB1W Korea white RPK- MAB1W UK white Immunity RPN-MAB1W Argentina white ESD: EN 61000-4-2, RPS- MAB1W AU white RS : EN61000-4-3 EFT/Burst : EN61000-4-4 Surge: EN 61000-4-5 CS : EN61000-4-6 Power Frequency Magnetic Field: EN61000-4-8 Voltage Dips : EN61000-4-11 Revised 12/26/2019