GPC40 Commercial/GPM40 Medical
40 WATT GLOBAL PERFORMANCE SWITCHERS
GLOBAL PERFORMANCE SWITCHERS
Summary:
Wide-range ac input 85-264 Vac
D C P o w e r S u p p l i e s
2-year warranty
Conducted EMI exceeds FCC Class B and CISPR 22
Class B (Commercial models) and CISPR 11
Class B (Medical models)
Single and multiple outputs
Commercial Approved to UL1950,
EN60950 E3:2000 and CSA-C22.2 No. 234
Medical Approved to UL2601-1, EN60601-1 and
CSA22.2 No. 601
Single and multiple outputs
RoHS compliant models available (G suffix)
marked to LVD
SPECIFICATIONS
Ac Input Output Noise
85-264 Vac, 47-63 Hz single phase. 0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential mode. Measured with noise
probe directly across output terminals of the power supply.
Input Current
Transient Response
Maximum input current at 120 Vac, 60 Hz with full rated output load: 1.3 A
Main output500 s typical response time for return to within 0.5% of final value
Hold-Up Time
for a 50% load step change. i/ t<0.2 A/s. Maximum voltage deviation is 3.5%.
20 ms minimum from loss of ac input at full load, nominal line (115 Vac).
Startup/ shutdown overshoot less than 3%.
Output Power
Remote Sense
40 W continuous, 50 W peak. Peak ratings are for 60 s maximum duration,
Provided as a standard feature on single-output models.
10% duty cycle. During peak load condition, output regulation may exceed
Voltage Adjustment
total regulation limits.
Built-in potentiometer adjusts main output voltage 5%.
Overload Protection
EMI/EMC Compliance
Fully protected against short circuit and output overload.
All models include built-in EMI filtering to meet the
Short circuit protection is cycling type power limit on outputs 1 & 2; foldback
following emissions requirements:
type on output 3. Recovery after fault is automatic. See output ratings chart
for additional notes or conditions.
EMI SPECIFICATIONS COMPLIANCE LEVEL
Overvoltage Protection. Conducted Emissions EN55022 Class B; FCC Class B
Main outputs: 124% 12%.
Conducted Emissions EN55011 Class B; FCC Class B
Static Discharge EN61000-4-2, 6 kV contact, 8 kV air
Efficiency
RF Field Susceptibility EN61000-4-3, 3 V/meter
70% at full rated load, nominal input voltage, depending on model and load
Fast Transients/Bursts EN61000-4-4, 2 kV, 5 kH
distribution.
Surge Susceptibility EN61000-4-5, 1 kV diff., 2 kV com.
Turn-on Time
Line Frequency Harmonics EN61000-3-2
Less than 1 second at 120 Vac, 25C (inversely proportional to input voltage
Commercial Safety
and thermistor temperature).
Approved to UL1950, CSA22.2 No. 950, IEC950, EN60950. All dc outputs are SELV
Input Protection
under normal and single fault conditions.
Internal ac fuse provided. Designed to blow only if a catastrophic failure oc-
curs in the unit.
Medical Leakage Current
Inrush Current
The maximum leakage current is as follows:
Inrush is limited by internal thermistors. Inrush at 240 Vac under cold start
Test Condition Normal Single Fault
conditions will not exceed 34 A.
120 Vac @ 60 Hz input 25 A 35 A
264 Vac @ 60 Hz input 40 A 60 A
Temperature Coefficient
Medical Safety
0.03%/C typical on all outputs.
Approved to UL2601-1, CSA22.2 No. 601 Level 3 and EN60601-1. UL file E116994;
Environmental
CSA #LR46516. The output(s) are intended for safety earthed Signal Output
Designed for 0 to 50C operation at full rated output power; derate output
and Intermediate Circuits only. The output(s) are not acceptable for patient con-
current and total output power by 2.5% per C above 50C.
nection without additional isolation. All dc outputs are SELV under normal and
single fault conditions.
O
All specifications are typical at nominal input, full load at 25 C unless otherwise statedCommercial Medical Output No. Output Output Output Output Noise P-P Total Notes
Model Model Minimum Maximum Peak Regulation
GPC40A GPM40A 1 +5.1 V 0.4 A 4 A 5 A 50 mV 2%
2 +12 V 0 A 2 A 3 A 120 mV 5% B,C
3 -12 V 0 A 0.4 A 0.7 A 120 mV 3% C
GPC40B GPM40B 1 +5.1 V 0.4 A 4 A 5 A 50 mV 2%
2 +15 V 0 A 2 A 3 A 150 mV 5% B,C
3 -15 V 0 A 0.4 A 0.7 A 150 mV 3% C
GPC40D GPM40D 1 +5.1 V 0.4 A 4 A 5 A 50 mV 2%
2 +24 V 0.A 1 A 1.5 A 240 mV 5% B,C
3 -12 V 0.A 0.4 A 0.7 A 120 mV 3% C
GPC40-3.3 GPM40-3.3 1 3.3 V 0 8 A 10 A 35 mV 2%
GPC40-5 GPM40-5 1 5 V 0 8 A 10 A 50 mV 2%
GPC40-9 GPM40-9 1 9 V 0 A 4.5 A 6 A 90mV 2%
GPC40-12 GPM40-12 1 12 V 0 A 3.3 A 4.2 A 120 mV 2%
GPC40-15 GPM40-15 1 15 V 0 A 2.7 A 3.3 A 150 mV 2%
GPC40-24 GPM40-24 1 24 V 0 A 1.7 A 2.1 A 240 mV 2%
GPC40-28 GPM40-28 1 28 V 0 A 1.4 A 1.8 A 280 mV 2%
Note: For 48 volt model, GPC40-48 or GPM40-48, contact factory for availability.
A. Total regulation is defined as the maximum deviation from the nominal voltage for all steady-state conditions of initial voltage setting, input line voltage and output load.
B. To maintain these regulation conditions, the 5.1V current must be at least 1/4 of V2 and not greater than 5 times the V2 current.
C. Requires +5.1V to be adjusted within 1% with at least a 0.4A load to maintain regulation on this output.
D. Add G suffix to model number for RoHS compliant model.
GPC40/GPM40 MECHANICAL SPECIFICATIONS
3
3
A. Units should be allowed to warm up/operate under non-condensing
ENVIRONMENTAL SPECIFICA-
OPERATING NON-OPERATING
onditions before application of power.
TIONS
B. Shock testinghalf-sinusoidal, 10 3 ms duration, direction,
Temperature (A) See individual specs. -40 to +85C
3 orthogonal axes, total 6 shocks.
C. Random vibration10 to 2000Hz, 6dB/octave roll-off from 350 to
Humidity (A) 0 to 95% RH 0 to 95% RH
2000Hz, 3 orthogonal axes. Tested for 10 min./axis operating and
Shock (B) 20 g 40 g
pk pk
1 hr./axis non-operating.
Altitude -500 to 10,000 ft -500 to 40,000 ft
2 2
Vibration (C) 1.5 g 0.003 g /Hz 5 g 0.026 g /Hz
rms rms
SL Power Electronics Corp, 6050 King Drive, Bldg. A, Ventura, CA 93003, USA. Phone:(805) 486 4565 Fax:(805) 487 8911 www.slpower.com. Rev. 1/07.
Data Sheet 2007 SL Power Electronics Corp. The information and specifications contained in this data sheet are believed to be correct at time of publication.
However, Condor accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.