GSM28 Medical 28 Watt Global Performance Switchers GLOBAL PERFORMANCE SWITCHERS FEATURES: D C P o w e r S u p p l i e s Industrys smallest 28 W medically approved switcher Compact size (4.00 x 2.59 x 0.92) Wide-range ac input: 85-264 Vac Less than 25 A leakage current 120 Vac Approved to UL2601-1, EN60601-1 EMI to FCC, CISPR 11 Class B/IEC601-1-2 Overvoltage protection standard RoHS Compliant Models Available (G suffix) marked to LVD SPECIFICATIONS Temperature Coefficient Ac Input 0.03%/C typical on all outputs. 85-264 Vac, 47-63 Hz single phase. Environmental Input Current Designed for 0 to 50C operation at full rated output power derate output current Maximum input current at 120 Vac, 60 Hz with full rated output load is 0.85 A. and total output power by 2.5% per C above 50C. Hold-up Time Medical EMI/EMC Compliance 15 ms minimum from loss of ac input at full load, nominal line (120 Vac). All models include built-in EMI filtering to meet the following emissions require - Output Power ments: Normal continuous output power is 28 W, 32 W peak for 60 s maximum dura- tion, 10% duty cycle. Factory set to begin power limiting at approximately 35 EMI SPECIFICATIONS COMPLIANCE LEVEL W. Conducted Emissions EN55011 Class B FCC Class B Overload Protection Static Discharge EN61000-4-2, 6 kV contact, 8 kV air Fully protected against short circuit and output overload. Short circuit pro- RF Field Susceptibility EN61000-4-3, 3 V/meter tection is cycling type power limit. Fast Transients/Bursts EN61000-4-4, 2 kV, 5 kHz Output Noise Surge Susceptibility EN61000-4-5, 1 kV diff., 2 kV com. 0.5% rms, 1% pk-pk, 20 MHz bandwidth, differential mode. Measured with Line Frequency Harmonics EN61000-3-2 Class A scope probe directly across output terminals of the power supply with load Earth Leakage Current terminated with 0.1 F capacitor. Leakage current measured in the Gnd wire connection when measured per EN60601-1 or UL2601-1 is as follows: Overvoltage Protection Built in with firing point set per ratings table. OVP firing reduces voltage to Medical Normal SingleFault Test Test less than 50% of nominal voltage in 50 ms. Model Leakage Leakage Voltage Method Voltage Adjustment GSM28 25A 45 A 132 Vca/60 Hz UL2601-1 Factory set with fixed resistors to maximize reliability. GSM28 50 A 90 A 264 Vca/50 Hz IEC60601-1 Efficiency 70% minimum for the 5.1 V model at full rated load, nominal input voltage. Medical Medical Safety Efficiency increases as output voltage increases. SL Power Electronics Corp. declares under our sole responsibility that all GSM models are in conformity with the applicable requirements of UL2601-1 Patient Input Protection Care Equipment, CSA-C22.2 No. 234 (with additional tests to C22.2 No. 601.1 per Internal ac fuse provided on all units. Designed to open only if a catastrophic T.I.L. CA-08), EN60601-1. failure occurs in the unit. Inrush Current Inrush limited by internal thermistor. The inrush at 230 Vac, averaged over the first ac half-cycle under cold start conditions will not exceed 32 A. Minimum Load Requirement 5% of full load rating Transient Response 3.5% max. dev. 50% load step at 0.2 A/s. Recovery to 0.5% within 500 s. O All specifications are typical at nominal input, full load at 25 C unless otherwise statedMedical Model Output Current Load Regulation Initial Setpoint Tolerance OVP Setpoint Ripple and Noise GSM28-5 5.1 V 5.5 A 0.75% 2.5% 6.2 0.6 V 1.4% GSM28-12 12 V 2.3 A 0.75% 2.5% 14 1.0 V 1% GSM28-15 15 V 1.9 A 0.75% 2.5% 18.5 1.5 V 1% GSM28-24 24 V 1.2 A 0.75% 2.5% 28 2.5 V 1% GSM28-28 28 V 1.0 A 0.75% 2.5% 34 2.8 V 1% * Add G suffix to model number for RoHS compliant model. GSM28 MECHANICAL SPECIFICATIONS ENVIRONMENTAL SPECIFICATIONS OPERATING NON-OPERATING A. Units should be allowed to warm up/operate under non-condensing Temperature (A) See individual specs -40 to +85C conditions before application of power. Humidity (A) 0 to 95% RH 0 to 95% RH B. Shock testinghalf-sinusoidal, 10 3 ms duration, direction, 3 or- Shock (B) 20 g 40 g thogonal axes, total 6 shocks. pk pk Altitude -500 to 10,000 ft -500 to 40,000 ft C. , Random vibration10 to 2000Hz, 6dB/octave roll-off from 350 to 2000Hz 3 orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis 2 2 Vibration (C) 1.5 g 0.003 g /Hz 5 g 0.026 g /Hz rms rms non-operating. SL Power Electronics Corp., 6050 King Drive, Bldg. A, Ventura, CA 93003, USA. Phone:(805) 486 4565 Fax:(805) 487 8911 www.slpower.com Rev. 1/07. Data Sheet 2007 SL Power Electronics Corp. The information and specifications contained in this data sheet are believed to be correct at time of publication. However, Condor accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specifications are subject to change without notice.