48W Single Output SLE48 Desktop Family Medical/Industrial Grade FEATURES AND BENEFITS Medical Medical & ITE Safety Approvals CoC V5 Tier 2(2016) 2 x MOPP Input to Output Isolation 0.075W Standby Power Industrial ( ) Suitable for Medical Equipment Up to Class BF 9V to 54V Outputs, Up to 48W Low Leakage Current 100uA 5,000m Operating Altitude DOE Efficiency Leve l VI AC Inlet IEC60320 C8 (class II) or C14 (class I) input VI IP22 Class I, II MODEL SELECTION Voltage Line & Load Efficiency Model Number Volts Current Rated Power Ripple & Noise Start Up Delay Tolerance Regulation (Average) 120mV SLE48S0903N01 9.0V 5.0A 45.0W 5% 87.80% 3s pk-pk 120mV SLE48S1203N01 12.0V 4.0A 48.0W 5% 87.80% 3s pk-pk 120mV 3s SLE48S1503N01 15. 0V 3.2A 48.0W 5% 87.80% pk-pk Line: 1% Load: 5% 120mV 3s 87.80% SLE48S1803N01 18.0V 2.66A 47.88W 5% pk-pk 150mV 3s SLE48S2403N01 24.0V 2.0A 48.0W 5% 87.80% pk-pk 300mV 3s SLE48S4803N01 48.0V 1.0A 48.0W 5% 87.80% pk-pk 120mV SLE48S0903F01 9.0V 5.0A 45.0W 5% 3s 87.80% pk-pk 120mV SLE48S1203F01 12.0V 4.0A 48.0W 5% 3s 87.80% pk-pk 120mV SLE48S1503F01 15.0V 48.0W 5% 3.2A 87.80% 3s pk-pk Line: 1% Load: 5% 120mV 2.66A 47.88W 87.80% 3s SLE48S1803F01 18.0V 5% pk-pk 150mV SLE48S2403F01 24.0V 2.0A 48.0W 5% 3s 87.80% pk-pk 300mV SLE48S 4803F01 48.0V 1.0A 48.0W 5% 3s 87.80% pk-pk Notes: 1. Measured at output connector with 20MHz bandwidth and 0.1F ceramic in parallel with 10F electrolytic capacitors. 2. Other output voltages in the range of 9V through 54V are available, contact us for details. 3. N in the model number (SLE48S2403N01) indicates IEC60320 C8 inlet (class II). indicates C14 (class I). C6, (class I) and C18 (class II) inputs are available, Contact SL Power for details. 4. 03 in the model number indicates 2.5 x 5.5 x 9.5mm straight barrel type connector. Other output connector options are available, please contact our sales representative for details. 5. Power supply is not provided with a line cord. 6. Power supplies are not medical equipment (applied parts), medical product manufacturers shall take responsibility for further evaluation of class B/BF/CF compliance of their end product. SLE48 Datasheet v0621 Copyright 2021 SL Power Electronics Corp. All rights reserved. Page 148W Single Output SLE48 Desktop Family Medical/Industrial Grade INPUT SAFETY Input Voltage and 90VAC264VAC Medical: IEC60601-1, ANSI/AAMI ES60601-1 CB Frequency 47Hz63Hz ITE: IEC62638-1, UL62638-1 Input Current 1.1A at 90VAC Medical: CAN/CSA C22.2 NO. 60601-1 UL ITE: CAN/CSA C22.2 NO. 62638-1 Inrush Current 70A max at 240VAC cold start TUV Medical: EN60601-1 ITE: EN62638-1 Touch Leakage 100A at 264VAC (max) CCC China GB4943 Current ENVIRONMENT EMI/EMC COMPLIANCE Medical: IEC/EN 60601-1-2, CISPR 11 Operating Temperature 0C to 40C Conducted Emissions ITE: EN55032, CISPR 22 Storage Temperature -20C to 60C Medical: IEC/EN 60601-1-2, CISPR 11 Radiated Emissions Weight 265 grams ITE: EN55032, CISPR 22 Dimensions 122.0mm x 51.0mm x 31.5mm Medical: IEC/EN 60601-1-2 Immunity ITE: EN55024, CISPR 24 Operating Humidity 10% RH to 90% RH, non-condensing Electro-Static Discharge Storage Humidity 5% RH to 90% RH EN61000-4-2, 15kV air, 8kV contact (ESD) Immunity on Power Ports Operating Altitude 5,000m Radiated RF EM Fields EN61000-4-3, 10V/m, 3V/m Dielectric Withstand 5,656VDC input to output 3 Susceptibility (80MHz2.7GMHz) Voltage Electrical Fast Transients EN61000-4-4, 2kV on AC port, 1kV on Insulation Resistance 10M Ohms, 500VDC input to output (EFT)/Bursts signal ports Surges, Line to Line (DM) EN61000-4-5, 1kV line to line (diff mode) PROTECTION and Line to Ground (CM) EN61000-4-6, 3Vrms, 6Vrms Overload Protection 120% 180% rated output power, Auto-recovery Conducted RF Immunity (0.15MHz80MHz) 120%1 80% rated output voltage input Overvoltage Protection Power Frequency Magnetic to reset EN61000-4-8, 30A/m Field Immunity Short circuit Protection Trip and restart - Hiccup Mode Voltage Dip Immunity EN61000-4-11, 0%, 70%, 0% of UT Harmonic Current EN61000-3-2, Class A RELIABILITY Emissions Flicker Test EN61000-3-3 MTBF >190,000 hours MIL-HDBK-217 at 25C PACKAGING ISOLATION Weight, Packing Qty 13kg, Individual Box: 4 0/carton Isolation Safety Rating 4,000VAC Input to Output (2 x MOPP) MECHANICAL DRAWING SLE48 Datasheet v0621 Copyright 2021 SL Power Electronics Corp. All rights reserved. Page 2