EFE300M / EFE400M Isolation Input to Output Reinforced 2 x MOPPs (3rd edition 60601) 4kVac, 5.7kVdc type tested to 4kVac (equivalent to 5.7kVdc), production tested to 4.3kVdc. Input to Earth Basic 1 x MOPP (3rd edition 60601), 1.5kVac, 2.3kVdc Output to Earth Basic 1 x MOPP (3rd edition 60601), 1.5kVac Output Specification EFE300M EFE400M 300W 400W Continuous (including fan supply) or RMS (including Peak power) Output Power EFE400M derates below 100Vac input and units fitted with fan, vertical output connector, ORing FET derate. See handbook for details. 400W 530W EFE300M - for 10 seconds. Outputs above 36V, 350W. Peak Power EFE400M - for 10 seconds. Outputs 47V and above, 470W. better than 4% Including Line (for 90-264Vac input change), Load (for 0-100% load change) Total Regulation and temperature (0-50C) Ripple & Noise 1.5% pk-pk, using EIAJ test method & 20MHz bandwidth Voltage Setting Accuracy 1% at 50% load Turn on Time 1.5s max at 90 Vac & 100% rated output power. EFE400M 2s max. Efficiency up to 90% Hold up >16ms at 90 Vac, 75% load Min Load None <5% of set voltage for 50% load change Transient Response (in 50s within the range 25 - 100% load) Recovery <1ms for recovery to 2% of set voltage Short circuit protection Yes Auto recovery after removal of short circuit Over Temperature protection Yes Primary - auto recovers, secondary - cycle power to restart Over Voltage Protection Yes Latching, need to cycle ac to restart unit. Fan supply 12V / 1A Depending on Case/Fan Option selected. See previous page for details Global Signals Remote on/off Enable - TTL logic level low (relative to Standby 0V) enables channel 1 and fan supply Inhibit - TTL logic level low (relative to Standby 0V) inhibits channel 1 and fan supply Standby Supply 5V / 2A or 12V / 1A, isolated supply, not affected by remote on/off. Power Good Logic high indicates ac supply is good and Ch1 is within regulation ORing FET Allows redundant connection of power supplies with no additional diodes required. Environment 0C to 50C operational, -40C to 70C storage (max 12 months). Temperature Full load, with 2m/s air blown from input to output (approximately 10CFM) Derating 50C to 70C derate each output by 2.5% per C Low Temp Startup -20C Humidity 5 - 95% RH non condensing 3 x 30g shocks in each plane, total 18 shocks 30g shock = 11ms (+/-0.5msec), half sine Shock Conforms to EN60068-2-27, EN60068-2-47, IEC68-2-27, IEC68-2-47, JIS C0041-1987. Conforms to MIL-STD-810E/F, Method 516.5, Pro IV, VI Single axis 10 - 500 Hz at 2g (sweep and endurance at resonance) in all 3 planes Vibration Conforms to EN60068-2-6, IEC68-2-6 Conforms to MIL-STD-810E, Method 514.4, Pro I, Cat 1,9 Medical approval = -200 to 3000 metres operational (-200 to 5000m storage/transportation) Altitude Non medical approval = -200 to 5000 metres operational (-200 to 5000m storage/transportation) a a - non open frame EFE400M units = -200 to 4000 metres Pollution Degree 2, Material group IIIb Emissions EN61000-6-3:2007, EN60601-1-2:2007 Radiated Electric Field EN55011, EN55032 (as per CISPR.11/22) Class B, FCC47 part 15 subpart B see application note for details Conducted Emissions EN55011, EN55032 (as per CISPR.11/22) Class B, FCC47 part 15 subpart B Conducted Harmonics EN61000-3-2 Class A Class C - EFE300M at 100W and above Flicker EN61000-3-3 Compliant - d only max EFE300M 400M Datasheet 69930 - v32.0 EFE300M / EFE400M 2