DC/DC Converter THM 6 Series, 6 Watt Wide 2:1 input voltage 6 W DC/DC converter in a compact DIP-24 plastic case I/O isolation 5000 VACrms rated for 250 VACrms working voltage Certification according to IEC/EN/ES 60601-1 3rd edition for 2xMOPP CB Scheme Risk management process according to IEC 606010-1 ES 60601-1 ISO 14971 including risk management file The THM-6 series is a range of medical 6 Watt DC/DC converters in DIP-24 Acceptance criteria for electronic plastic package and with wide 2:1 input voltage range. They provide a reinforced assemblies according to IPC-A-610 Level 3 isolation system for 5000 VACrms isolation and a very low leakage current of less than 2 A. The units are approved to IEC/EN/ES 60601-1 3rd edition for 2 Low leakage current < 2A x MOPP (Means Of Patient Protection) and come along with an ISO 14971 risk Extended operating temperature range management file. Design and production conform to the quality management 40C to 90C. system ISO 13485. With a high efficiency of up to 89% and highest grade com - EMC compliance to IEC 60601-1-2 4th ponents the converters can reliably operate in an ambient temperature range edition and EN55032 class A of 40C up to +90C. They constitute a reliable solution not only for medical Operating up to 5000m altitude equipment but also for demanding ranges of application such as transportation, 5 year product warranty control & measurement or IGBT drivers. Models Order code* Input voltage range Output voltage Output current max. Efciencfi y typ. THM 6-0510 3.3 VDC 1800 mA 81.5 % THM 6-0511 5.0 VDC 1200 mA 86.0 % THM 6-0512 12 VDC 500 mA 86.0 % THM 6-0513 15 VDC 400 mA 87.5 % 4.5 9 VDC THM 6-0515 (5 VDC nominal) 24 VDC 250 mA 87.0 % THM 6-0521 5.0 VDC 600 mA 84.0 % THM 6-0522 12 VDC 250 mA 86.5 % THM 6-0523 15 VDC 200 mA 87.5 % THM 6-1210 3.3 VDC 1800 mA 83.5 % THM 6-1211 5.0 VDC 1200 mA 86.0 % THM 6-1212 12 VDC 500 mA 89.0 % THM 6-1213 15 VDC 400 mA 88.5 % 9. 0 18 VDC THM 6-1215 (12 VDC nominal) 24 VDC 250 mA 88.5 % THM 6-1221 5.0 VDC 600 mA 85.0 % THM 6-1222 12 VDC 250 mA 89.0 % THM 6-1223 15 VDC 200 mA 88.0 % THM 6-2410 3.3 VDC 1800 mA 83.0 % THM 6-2411 5.0 VDC 1200 mA 86.0 % THM 6-2412 12 VDC 500 mA 89.0 % THM 6-2413 15 VDC 400 mA 89.0 % 18 36 VDC THM 6-2415 (24 VDC nominal) 24 VDC 250 mA 88.5 % THM 6-2421 5.0 VDC 600 mA 85.0 % THM 6-2422 12 VDC 250 mA 88.5 % THM 6-2423 15 VDC 200 mA 88.5 % THM 6-4810 3.3 VDC 1800 mA 82.5 % THM 6-4811 5.0 VDC 1200 mA 86.5 % THM 6-4812 12 VDC 500 mA 88.0 % THM 6-4813 15 VDC 400 mA 88.5 % 36 75 VDC THM 6-4815 (48 VDC nominal) 24 VDC 250 mA 88.0 % THM 6-4821 5.0 VDC 600 mA 85.0 % THM 6-4822 12 VDC 250 mA 88.0 % THM 6-4823 15 VDC 200 mA 87.0 % * suffix -B2 for trim option with adjustable output suffix -B3 for remote control option suffix -B4 for trim + remote-control option www.tracopower.com Page 1 of 5 THM 6 Series, 6 Watt Input Specifications Input current no load 5 Vin models: 20 mA typ. 12 Vin models: 10 mA typ. 24 Vin models: 6 mA typ. 48 Vin models: 4 mA typ. Surge voltage (3 s max.) 5 Vin models: 16 V max. 12 Vin models: 25 V max. 24 Vin models: 50 V max. 48 Vin models: 100 V max. Start-up voltage 5 Vin models: 4.5 VDC (or lower) 12 Vin models: 9 VDC (or lower) 24 Vin models: 18 VDC (or lower) 48 Vin models: 36 VDC (or lower) Startup time 30 ms Under voltage shut down 5 Vin models: 4 VDC typ. 12 Vin models: 8 VDC typ. 24 Vin models: 16 VDC typ. 48 Vin models: 33 VDC typ. EMC emissions Conducted & Radiated input suppression EN 55011 limits to IEC 60601-1-2 4th editon EN 55032 class A (internal filter) Application note for filter class B proposal www.tracopower.com/overview/thm6 EMC immunity Generic for Medical equipment IEC/EN 60601-1-2 4th edition ESD (electrostatic discharge) EN 61000-4-2, air 8 kV, contact 6 kV, perf. criteria A Radiated immunity EN 61000-4-3, 10 V/m, perf. criteria A Fast transient / surge EN 61000-4-4, 2 kV, perf. criteria A (with external input capacitor / diode) EN 61000-4-5, 2 kV perf. criteria A 5 Vin models: Nippon chemi-con KY 1000 F/ 25 V and reverse diode (Vishay V10P45) in parallel 12 & 24 Vin models: Nippon chemi-con KY 470 F/ 50 V 48 Vin models: Nippon chemi-con KY 330 F/ 100 V Conducted immunity EN 61000-4-6, 10 Vrms, perf. criteria A Magnetic field immunity EN 61000-4-8 100 A/m, continuous, perf. criteria A 1000 A/m, 1 sec., perf. criteria A External input fuse required 5 Vin models: 2.5 A (recommended values, slow blow type) 12 Vin models: 1.25 A 24 Vin models: 0.63 A 48 Vin models: 0.315 A www.tracopower.com Page 2 of 5