AC/DC Medical Power Supply TPP 15-D Series, 15 Watt High power density power supply (encapsulated) Certification according to IEC/EN/ES 60601-1 3rd edition for 2 x MOPP Low leakage current <75 A rated for BF applications EMC compliance to IEC 60601-1-2 4th edition Risk management process according to ISO 14971 incl. risk management file Acceptance criteria for electronic assemblies acc. to IPC-A-610 Level 3 Protection class ll ES 60601-1 IEC 60601-1 Operating up to 5000 m altitude UL 62368-1 IEC 62368-1 Ready to meet ErP directive, no load The TPP 15-D AC/DC power supplies feature a reinforced double I/O isolation power consumption <75 mW system according to medical safety standards IEC/EN/ES 60601-1 3rd edition for 2 x MOPP approved for an operating altitude of 5000 m. The earth leakage 5-year product warranty current is below 75 A what makes the units suitable for BF (body floating) ap- plications. The excellent efficiency of up to 88.5% offers a high power density in the packaging format 1.1 x 1.65. The full load operating temperature range covers 40C to +70C while it goes up to 85C with 50% load derating. The units operate in compliance to the medical EMC emission and immunity levels according to latest standard IEC 60601-1-2 4th edition. Models Order Code Output Power Output Voltage Output Current Efficiency max. nom. (adjustable) max. typ. TPP 15-103-D 13.2 W 3.3 VDC (2.97 - 3.63 VDC) 4 000 mA 84 % TPP 15-105-D 5 VDC (4.5 - 5.5 VDC) 3 000 mA 86 % TPP 15-109-D 9 VDC (8.1 - 9.9 VDC) 1 670 mA 86 % 12 VDC (10.8 - 13.2 VDC) 1 250 mA 87 % TPP 15-112-D TPP 15-115-D 15 W 15 VDC (13.5 - 16.5 VDC) 1 000 mA 87 % TPP 15-124-D 24 VDC (21.6 - 26.4 VDC) 625 mA 88 % TPP 15-136-D 36 VDC (32.4 - 39.6 VDC) 417 mA 88 % TPP 15-148-D 48 VDC (43.2 - 52.8 VDC) 313 mA 89 % Page 1 / 5 www.tracopower.com October 14, 2020TPP 15-D Series, 15 Watt Input Specifications Input Voltage - AC Range 85 - 264 VAC (Full Range) - DC Range 120 - 370 VDC (Designed for, no certification) Input Frequency 47 - 63 Hz Input Current - Full Load & Vin = 230 VAC 300 mA max. - Full Load & Vin = 115 VAC 450 mA max. Power Consumption - At no load 75 mW max. (Ready to meet ErP directive) Input Inrush Current - At 230 VAC 40 A max. Input Protection T 1.6 A / 250 VAC (Internal Fuse) Recommended Input Fuse (The need of an external fuse has to be assessed in the final application.) Output Specifications Output Voltage Adjustment 10% (By external trim resistor) See application note: www.tracopower.com/overview/tpp15-d Output power must not exceed rated power Voltage Set Accuracy 1% max. Regulation - Input Variation (Vmin - Vmax) 0.2% max. - Load Variation (0 - 100%) 0.7% max. (3.3 and 5 VDC model) 0.5% max. (other output models) Ripple and Noise 3.3 VDC model: 40 mVp-p typ. (w/ 10 F X5R) (20 MHz Bandwidth) 5 VDC model: 40 mVp-p typ. (w/ 10 F X5R) 9 VDC model: 70 mVp-p typ. (w/ 10 F X5R) 12 VDC model: 70 mVp-p typ. (w/ 10 F X5R) 15 VDC model: 70 mVp-p typ. (w/ 10 F X5R) 24 VDC model: 100 mVp-p typ. (w/ 10 F X5R) 36 VDC model: 100 mVp-p typ. (w/ 10 F X5R) 48 VDC model: 140 mVp-p typ. (w/ 1 F X7R) Capacitive Load 3.3 VDC model: 6 000 F max. 5 VDC model: 4 000 F max. 9 VDC model: 1 860 F max. 12 VDC model: 1 200 F max. 15 VDC model: 820 F max. 24 VDC model: 470 F max. 36 VDC model: 220 F max. 48 VDC model: 150 F max. Minimum Load Not required Temperature Coefficient 0.02 %/K max. Hold-up Time - At 115 VAC 8 ms min. Start-up Time - At 230 VAC 500 ms max. Short Circuit Protection Continuous, Automatic recovery Output Current Limitation 120 - 200% of Iout max. 145% typ. of Iout max. Overvoltage Protection 125 - 140% of Vout nom. Transient Response - Response Deviation 8% max. (75% to 100% Load Step) - Response Time 500 s typ. (75% to 100% Load Step) All specifications valid at nominal voltage, full load and +25C after warm-up time unless otherwise stated. Page 2 / 5 www.tracopower.com October 14, 2020