AC/DC Medical Power Supply TPP 30A-J Series, 30 Watt High power density power supply (open frame) Certification according to IEC/EN/ES 60601-1 3rd edition for 2 x MOPP Low leakage current <75 A rated for BF applications EMC compliance to IEC 60601-1-2 4th edition Risk management process according to ISO 14971 incl. risk management file Acceptance criteria for electronic assemblies acc. to IPC-A-610 Level 3 Protection class l and ll ES 60601-1 IEC 60601-1 Operating up to 5000 m altitude UL 62368-1 IEC 62368-1 Ready to meet ErP directive, no load The TPP 30A-J AC/DC power supplies feature a reinforced double I/O isolation power consumption <60 mW system according to medical safety standards IEC/EN/ES 60601-1 3rd edition for 2 x MOPP approved for an operating altitude of 5000 m. The earth leakage 5-year product warranty current is below 75 A what makes the units suitable for BF (body floating) ap- plications. The excellent efficiency of up to 92% offers a high power density in the packaging format 1.36 x 3.34. The full load operating temperature range covers 40C to +60C while it goes up to 85C with 50% load derating. The units operate in compliance to the medical EMC emission and immunity levels according to latest standard IEC 60601-1-2 4th edition. Models Order Code Output Power Output Voltage Output Current Efficiency max. nom. (adjustable) max. typ. TPP 30-103A-J 20 W 3.3 VDC (2.97 - 3.63 VDC) 6 000 mA 84 % TPP 30-105A-J 5 VDC (4.5 - 5.5 VDC) 6 000 mA 87 % TPP 30-109A-J 9 VDC (8.1 - 9.9 VDC) 3 340 mA 88 % 12 VDC (10.8 - 13.2 VDC) 2 500 mA 91 % TPP 30-112A-J TPP 30-115A-J 30 W 15 VDC (13.5 - 16.5 VDC) 2 000 mA 91 % TPP 30-124A-J 24 VDC (21.6 - 26.4 VDC) 1 250 mA 90 % TPP 30-136A-J 36 VDC (32.4 - 39.6 VDC) 840 mA 90 % TPP 30-148A-J 48 VDC (43.2 - 52.8 VDC) 630 mA 92 % Page 1 / 4 www.tracopower.com November 3, 2020TPP 30A-J Series, 30 Watt Input Specifications Input Voltage - AC Range 85 - 264 VAC (Full Range) - DC Range 120 - 370 VDC (Designed for, no certification) Input Frequency 47 - 63 Hz Input Current - Full Load & Vin = 230 VAC 400 mA max. - Full Load & Vin = 115 VAC 800 mA max. Power Consumption - At no load 60 mW max. (Ready to meet ErP directive) Input Inrush Current - At 230 VAC 40 A max. Input Protection T 1.6 A / 250 VAC (Internal Fuse in L & N) Recommended Input Fuse (The need of an external fuse has to be assessed in the final application.) Output Specifications Output Voltage Adjustment 10% (By trim potentiometer) Output power must not exceed rated power Voltage Set Accuracy 1% max. Regulation - Input Variation (Vmin - Vmax) 0.2% max. - Load Variation (0 - 100%) 0.7% max. (3.3 and 5 VDC model) 0.5% max. (other output models) Ripple and Noise 50 mVp-p typ. (w/ 10 F X7R) (20 MHz Bandwidth) Capacitive Load 3.3 VDC model: 10 000 F max. 5 VDC model: 12 000 F max. 9 VDC model: 3 720 F max. 12 VDC model: 2 085 F max. 15 VDC model: 1 350 F max. 24 VDC model: 520 F max. 36 VDC model: 235 F max. 48 VDC model: 130 F max. Minimum Load Not required Temperature Coefficient 0.02 %/K max. Hold-up Time - At 115 VAC 16 ms min. Start-up Time - At 230 VAC 1 500 ms max. Short Circuit Protection Continuous, Automatic recovery Output Current Limitation 110 - 170% of Iout max. 140% typ. of Iout max. Overvoltage Protection 125 - 140% of Vout nom. Transient Response - Response Deviation 3% max. (50% to 75% Load Step) - Response Time 500 s typ. (50% to 75% Load Step) Safety Specifications Safety Standards EN 62368-1 - IT / Multimedia Equipment IEC 62368-1 UL 62368-1 - Household EN 60335-1 IEC 60335-1 - Medical Equipment EN 60601-1 IEC 60601-1 ANSI/AAMI ES 60601-1 2 x MOPP (Means Of Patient Protection) - Power Transformers IEC 61558-1 IEC 61558-2-16 - Certification Documents www.tracopower.com/overview/tpp30a-j Protection Class Class I & II (Prepared): Reinforced Insulation Pollution Degree PD 2 Over Voltage Category OVC II All specifications valid at nominal voltage, full load and +25C after warm-up time unless otherwise stated. Page 2 / 4 www.tracopower.com November 3, 2020