AC/DC Medical Power Supply TPP 65E-D Series, 65 Watt Compact encapsulated 65 Watt PCB module (3.2 x2.2 ) I/O isolation 4000 VAC rated for 250 VAC working voltage Certification according to IEC/EN/ES 60601-1 3rd edition for 2 x MOPP Risk management process according to ISO 14971 incl. risk management file Acceptance criteria for electronic assemblies acc. to IPC-A-610 Level 3 Low leakage current <100 A rated for BF applications ES 60601-1 IEC 60601-1 UL 62368-1 IEC 62368-1 Operating temperature 40C to 80C The TPP65E-D is a 65 Watt encapsulated AC/DC modules in a PCB package. It EMC compliance according to IEC features a reinforced double I/O isolation system according to latest medical 60601-1-2 4th edition and EN55032 safety standards (60601-1 3rd edition, 2 x MOPP). The leakage current is be- class B low 100 A and makes the module suitable for BF (body floating) applications. Operating up to 5000m altitude The excellent efficiency of up to 94% allows a high power density and compact design (3.2 x 2.2). The operating temperature range is 40C to +80C with 5-year product warranty derating above 60C. The EMC characteristic is dedicated for applications in in- dustrial and medical fields. High reliability is provided by use of high quality components and an excellent thermal management making the TPP 65E-D an ideal solution for medical devices and for demanding safety and space critical applications. Models Order Code Output Power Output Voltage Output Current Efficiency max. nom. (adjustable) max. typ. TPP 65-105E-D 50 W 5 VDC (4.5 - 5.5 VDC) 10 000 mA 90 % TPP 65-112E-D 12 VDC (10.8 - 13.2 VDC) 5 420 mA 93 % TPP 65-115E-D 15 VDC (13.5 - 16.5 VDC) 4 340 mA 94 % TPP 65-124E-D 65 W 24 VDC (21.6 - 26.4 VDC) 2 710 mA 94 % 36 VDC (32.4 - 39.6 VDC) 1 810 mA 93 % TPP 65-136E-D TPP 65-148E-D 48 VDC (43.2 - 52.8 VDC) 1 360 mA 93 % Page 1 / 5 www.tracopower.com November 23, 2021TPP 65E-D Series, 65 Watt Input Specifications Input Voltage - AC Range Operational Range: 85 - 264 VAC (Full Range) Rated Range: 100 - 240 VAC (Full Range) - DC Range Operational Range: 120 - 370 VDC (Designed for, no certification) Polarity: +DC: L / DC: N Input Frequency 47 - 63 Hz Input Current - Full Load & Vin = 230 VAC 900 mA max. - Full Load & Vin = 115 VAC 1 500 mA max. Power Consumption - At no load 210 mW max. (Ready to meet ErP directive) Input Inrush Current - At 230 VAC 80 A max. - At 115 VAC 30 A max. Input Protection T 3.15 A / 250 VAC (Internal Fuse L & N) Recommended Input Fuse 3 150 mA (slow blow) (The need of an external fuse has to be assessed in the final application.) Output Specifications Output Voltage Adjustment 10% (By external trim resistor) See application note: www.tracopower.com/overview/tpp65e-d Output power must not exceed rated power Voltage Set Accuracy 1% max. Regulation - Input Variation (Vmin - Vmax) 0.2% max. - Load Variation (0 - 100%) 0.5% max. Ripple and Noise 5 VDC model: 100 mVp-p max. (w/ 10uF) (20 MHz Bandwidth) 12 VDC model: 100 mVp-p max. (w/ 10uF) 15 VDC model: 100 mVp-p max. (w/ 10uF) 24 VDC model: 120 mVp-p max. (w/ 1uF) 36 VDC model: 120 mVp-p max. (w/ 1uF) 48 VDC model: 200 mVp-p max. (w/ 0.1uF) 5 VDC model: 75 mVp-p typ. (w/ 10uF) 12 VDC model: 75 mVp-p typ. (w/ 10uF) 15 VDC model: 75 mVp-p typ. (w/ 10uF) 24 VDC model: 75 mVp-p typ. (w/ 1uF) 36 VDC model: 75 mVp-p typ. (w/ 1uF) 48 VDC model: 150 mVp-p typ. (w/ 0.1uF) Capacitive Load 5 VDC model: 20 000 F max. 12 VDC model: 4 520 F max. 15 VDC model: 2 900 F max. 24 VDC model: 1 130 F max. 36 VDC model: 520 F max. 48 VDC model: 285 F max. Minimum Load Not required Temperature Coefficient 0.02 %/K max. Hold-up Time - At 230 VAC 75 ms min. - At 115 VAC 16 ms min. Start-up Time - At 230 VAC 1 000 ms max. - At 115 VAC 1 000 ms max. Short Circuit Protection Continuous, Automatic recovery Output Current Limitation 120 - 180% of Iout max. 145% typ. of Iout max. Overvoltage Protection 130% typ. of Vout nom. 125 - 140% of Vout nom. Transient Response - Response Deviation 3% max. (50% to 75% Load Step at 2.5 A/s) - Response Time 600 s typ. (50% to 75% Load Step at 2.5 A/s) All specifications valid at nominal voltage, resistive full load and +25C after warm-up time, unless otherwise stated. Page 2 / 5 www.tracopower.com November 23, 2021