UL / IEC / EN 60601 3.1 Edition & UL / IEC / EN
60950 AM2 Safety Approvals
High power density: 500W in 3 x 5 footprint
Open Frame or Enclosed Versions Available
Remote ON/OFF Function
Built-in 12V/0.3A Auxiliary Output
Standby 5V @ 1A with Fan, @ 0.4A without
Fan
High Efficiency up to 93%
P.F.C. Function >0.95
Electrical Specifications
Input Environmental & Operating
Operating Temperature
-40C to +70C (with derating)
Storage Temperature -40C to +85C
Input Voltage 90-264 VAC or 120-370 VDC
Humidity 95% RH
Input Frequency 47-63 Hz
Operating altitude <3000m for medical use
Input Current (RMS) <6.0A @ 115 VAC; <3.00A @ 230 VAC
MTBF: >160K hours per MIL-HDBK-217F at full load
Power Factor >0.95 @ full load (230 VAC)
and 25C ambient
Inrush Current (<2ms) <40A @ 115 VAC; <80A @ 230 VAC
Earth Leakage Current < 0.1 mA max. (Input-Output)
Compliance
Safety Approvals
rd
USA/Canada UL60601-1 3 Edition , UL/cUL60950-1
Output
UL62368-1 (Pending)
Total Output 500W max. See table for details
Europe IEC/EN60601-1 3rd edition, TUV EN60950-1
Output Voltage See table
CB Report
Hold Up Time 8ms typical at full load and 115VAC
EN62368-1 (Pending)
(Note 4)
nominal line.
Efficiency Up to 93%. See table for details
Isolation: 4000VAC input to output, 2 x MOPP
1500VAC input to ground,1 x MOPP
Minimum Load 0%
1500VAC output to ground,1 x MOPP
EMC (IEC60601-1-2:2014) : FCC Class B Radiated & Conducted
EN55011/55022 Class B Radiated &
Conducted
Harmonic Currents IEC 61000-3-2:
Voltage Flicker IEC 61000-3-3
Electrostatic Discharge IEC 61000-4-2: 15kV Air, 8kV contact
Radiated Immunity IEC 61000-4-3: 10V/m
EFT/Burst IEC 61000-4-4: +/-2kV
Protection
Surge Immunity IEC 61000-4-5: 2005 1kV diff, 2kV com
Conducted Immunity IEC 61000-4-6: 10Vrms
Overvoltage 110-132%, Auto recovery
Magnetic Field IEC 61000-4-8: 30A/m
Dips / Interruptions IEC 61000-4-11: 30% reduction for 500ms,
Overload 145-170%, Auto recovery
100% reduction for 10ms.
Short Circuit Auto recovery
General
Dimensions 3.0W x 5.03L x 1.38H
Weight 1.058 pounds (480 g)
Your Partners in Power ... Power Partners, Inc. | 43 Broad Street | Hudson, MA 01749
1
Tel: (978)567-9600 | Fax: (978)567-9601
Specifications subject to change.
Website: www.powerpartners-inc.com
PDAM500: 11/12/2018
Medical
AC/DC Power
Models and Ratings
Output Max Load Max Load 30CFM Output Ripple & Efficiency Fan +5VSB
1
Model 2 6 7
Voltage Convection Forced Air Regulation Noise (230 VAC) Output Output
PDAM500-12A 12V 20A 41.5A 1.2% 160mV 90.5% 12V/0.3A 5V/1A
PDAM500-13A 15V 14.66A 33.3A 1.0% 160mV 90.5% 12V/0.3A 5V/1A
PDAM500-14A 24V 10A 20.8A 1.0% 240mV 92% 12V/0.3A 5V/1A
PDAM500-18A 48V 5A 10.41A 1.0% 480mV 93% 12V/0.3A 5V/1A
Note:
1. All models are available in an enclosed version (e.g. PDAM500-12A would be PDAM500-12C)
2. Listed values are taken at 230VAC. At 115VAC, 19.16A, 9.58A, and 4.8A for the models listed as shown.
3. Recommended to add Varistor 14S471K at L/N input side in parallel.
4. Hold-up Time measured at 90% Vout
5. Main Vout >3% Load, 12V (Aux) / 0.3A.
6. Measured at 20MHz bandwidth with a 47uF electrolytic capacitor and 0.1uF ceramic capacitor in parallel at the output connector.
7. 400mA convection rated.
Block Diagram
Your Partners in Power ... Power Partners, Inc. | 43 Broad Street | Hudson, MA 01749
2
Tel: (978)567-9600 | Fax: (978)567-9601
Specifications subject to change.
Website: www.powerpartners-inc.com
PDAM500: 11/12/2018
Medical
AC/DC Power