DC-DC Converter XPPower IMB01 Series 1 Watt Single unregulated output 10% input range SIP7 package World wide medical approvals 4kVAC isolation reinforced 2 x MOPP at 300VAC 2A patient leakage current -40C to +105C operation Full load at +85C ambient Dimensions: MTBF 4.3MHrs IMB01: 3 year warranty 0.87 x 0.49 x 0.30 (22.0 x 12.5 x 7.5 mm) Designed for medical applications, the IMB01 series offers a 2 x MOPP and IEC60601-1 compliant solution. Included in the compact SIP7 package is short circuit protection, Iow leakage current of 2A and operation to +85C. Models & Ratings Input Current Maximum Capacitive Input Voltage Output Voltage Output Current Efficiency Model Number Load F No Load Full Load 5 V 200 mA 253 mA 79% IMB0105S05 5V 12 V 84 mA 50 mA 252 mA 220 F 80% IMB0105S12 (4.5-5.5) 15 V 68 mA 252 mA 81% IMB0105S15 5 V 200 mA 105 mA 79% IMB0112S05 12V 12 V 84 mA 35 mA 104 mA 220 F 81% IMB0112S12 (10.8-13.2) 15 V 68 mA 108 mA 79% IMB0112S15 5 V 200 mA 55 mA 76% IMB0124S05 24V 12 V 84 mA 20 mA 53 mA 220 F 79% IMB0124S12 (21.6-26.4) 15 V 68 mA 54 mA 79% IMB0124S15 Notes Input currents measured at nominal input voltage. Mechanical Details 0.87 (22.0) 0.83 (21.0) 0.01 (0.25) Bottom View 0.30 0.49 (7.5) (12.5) 1 2 6 7 0.02 (0.5) 0.02 (0.5) 0.08 (2.0) Pin Connections Notes 0.02 Pin Single 1. All dimensions are in inches (mm) (0.5) 0.13 2. Weight: 0.009 lbs (4.1 g) approx. 1 +Vin (3.2) 2 -Vin 3. Pin diameter: 0.020.002 (0.50.05) 0.40 0.14 6 -Vout 4. Pin pitch tolerance: 0.014 (0.25) (3.5) (10.16) 5. Case tolerance: 0.02 (0.5) 7 +Vout 0.10 0.10 (2.54) (2.54) www.xppower.com 1DC-DC Converter XPPower IMB01 Series Input Characteristic Minimum Typical Maximum Units Notes & Conditions 4.5 5 5.5 5 V nominal 10.8 12 13.2 12 V nominal Input Voltage Range VDC 21.6 24 26.4 24 V nominal -0.7 9 VDC for 1s 5 V models -0.7 18 VDC for 1s 12 V nominal Input Surge -0.7 30 VDC for 1s 24 V nominal Input Filter Internal Capacitor Output Characteristic Minimum Typical Maximum Units Notes & Conditions Output Voltage 5 15 VDC See Models and Ratings table Patient Leakage Current 2 A Initial Set Accuracy 1 3 % % V nominal Minimum Load 10 % Minimum load required for regulation Line Regulation 1.2 1.5 % From minimum to maximum input Load Regulation 10 % From 10% to full load Ripple & Noise 75 mV pk-pk 20 MHz bandwidth. Measured using 0.1 F ceramic capacitor Short Circuit Protection Continuous, with auto recovery Maximum Capacitive Load See Models and Ratings table Temperature Coefficient 0.01 0.02 %/ C General Characteristic Minimum Typical Maximum Units Notes & Conditions Efficiency 80 % Isolation: Input to Output 4000 VAC 60s reinforced, 300 Vrms 2 x MOPP Switching Frequency 60 kHz May enter burst mode frequency of 12-28 kHz at light load 9 Isolation Resistance 10 500 VDC Isolation Capacitance 20 100 kHz, 1V pF 3 Power Density 7.80 W/in Mean Time Between Failure 4.3 MHrs MIL-HDBK-217F, +25 C GB Weight 0.009 (4.1) lb (g) Case Material Non conductive black plastic, UL94V-0 rated Pin Material Tinned Copper Environmental Characteristic Minimum Typical Maximum Units Notes & Conditions Operating Altitude 5000 m Transport Altitude 10,000 m Derate linearly from 100% load at +85 C to 0% at Operating Temperature -40 +105 C +105 C Storage Temperature -50 +125 C Case Temperature +105 C Humidity 95 %RH Non-condensing Cooling Natural convection EMC: Emissions Phenomenon Standard Test Level Notes & Conditions Conducted EN55011 Class A See Application Note Radiated EN55011 Class A www.xppower.com 2