SGX Europe Sp. Z o.o. T: +48 (0) 32 438 4778 Building 11 Ligocka St. 103, E: sales.is sgxsensortech.com 40-568 Katowice, www.sgxsensortech.com Poland SGX-4NO2-2E Datasheet OUTLINE Industrial 2 electrode Nitrogen Dioxide (NO2) Sensor All dimensions are in mm All tolerances are 0.15mm PERFORMANCE Output signal 300 100 nA / ppm Typical Baseline Range 0.1ppm NO equivalent 2 (pure air) T90 Response Time < 60 seconds Measurement Range 0 - 20 ppm Maximum Overload 100 ppm Linearity Linear Repeatability < 2% NO equivalent 2 Recommended Load 10 ohms Resistor Resolution (Electronics 0.1 ppm typical dependent) OPERATING CONDITIONS Temperature Range -30C to +50C Pressure Range 800 to 1200 mbar Operating Humidity 15% to 90% RH Range LIFETIME INTRINSIC SAFETY DATA Long Term Output Drift <20% per annum Max at 2000ppm 0.3 mA Recommended Storage 0C to 20C Max o/c Voltage 1.3 V Temp Expected Operating Life > 24 months in air Max s/c Current <1.0 A SGX Sensortech 2012-2020 DS-0337 SGX-4NO2-2E, Issue 1, 27-Mar-2020, Page 1/2 SGX Europe Sp. Z o.o. T: +48 (0) 32 438 4778 Building 11 Ligocka St. 103, E: sales.is sgxsensortech.com 40-568 Katowice, www.sgxsensortech.com Poland CROSS SENSITIVITY DATA GAS CONCENTRATION SGX-4NO2-2E Carbon Monoxide 300ppm 0 ppm Suplhur Dioxide 20ppm 0 ppm Hydrogen 200ppm 0 ppm Nitric Oxide 50ppm 0 ppm Ammonia 50ppm 0 ppm Chlorine 1ppm <0.5 ppm * See Important Notes IMPORTANT NOTES 1) Do not glue or solder to the connector pins as this may damage the sensor and thereby invalidate the warranty, please use PSB sockets. 2) Details on recommended connector pins can be found in the Frequently Asked Questions within the Gas Sensor section of the SGX website. 3) All performance specifications are based upon the following environmental conditions: 20C, 50% relative humidity, 101kPa. 4) Cross Sensitivity Data is for information only. Calibration is recommended with target gas as the accuracy of calibration and measurement cannot be ensured. 5) The cross sensitivities are including but not limited to the gases stated in the table. It may respond to other gases. 6) The cross sensitivities may fluctuate between 30% and may differ from batch to batch or across sensors life time. 7) The device is designed to be RoHS compliant. 8) Poisoning - sensors are designed to operate in a wide range of harsh environments and conditions. However, it is important that exposure to high concentrations of solvent vapours is avoided, both during storage, fitting into instrument and operation. 9) When using sensors on printed circuit boards (PCBs), degreasing agents should be used prior to the sensor being fitted. 10) Output of SGX-4NO2-2E sensor is in negative polarity (compare to CO or H2S sensor) Warning: By the nature of the technology used, any electrochemical gas sensor offered by SGX Europe Sp. z o.o. can potentially fail to meet specification without warning. SGX Europe Sp. z o.o. makes every effort to ensure the reliability of our products of this type, where life safety is a performance requirement of the product, we recommend that all sensors and instruments using these sensors are checked for response to gas before use. SGX Europe Sp. z o.o reserves the right to make product changes without notice. No liability is accepted for any consequential losses, injury or damage resulting from the use of this document or from any omissions or errors herein. The data is given for guidance only. It does not constitute a specification or an offer for sale. The products are always subject to a program of improvement and testing which may result in some changes in the characteristics quoted. As the products may be used by the client in circumstances beyond the knowledge and control of SGX Europe Sp. z o.o, we cannot give any warranty as to the relevance of these particulars to an application. It is the clients responsibility to carry out the necessary tests to determine the usefulness of the products and to ensure their safety of operation in a particular application. Performance characteristics on this data sheet outline the performance of newly supplied sensors. Output signal can drift below the lower limit over. SGX Sensortech 2012-2020 DS-0337 SGX-4NO2-2E, Issue 1, 27-Mar-2020, Page 2/2