8 April 2015
Page 1
FEATURES 260W Medical Power Supply SP-M260 Series
9VDC to 48VDC in a 1VDC increments
Active Power Factor & Harmonic Current Correction
Short Circuit, Over Load and Over Voltage Protection
90-264Vac Wide Range Input
Over Temperature Protection Option
nd nd
UL/cUL 60601 2 Ed, DEMKO (EN60601 2 ED), CB and CE Certified
RoHS Compliant
Low Profile 1U
SPECIFICATION
Common Models SP-M250-09-B SP-M250-12-B SP-M250-20-B SP-M250-24-B SP-M250-48-B SP-M250-58-B*
Voltage (1) 9V 12V 20V 24V 48V 58V
Current Range 0 ~ 20A 0 ~ 21.67A 0 ~ 13A 0 ~ 10.83A 0 ~ 5.42A 0 ~ 4.5A
Rated Power 180W 260W 260W 260W 260W 260W
Ripple & Noise
1% Pk to Pk
(Max) (2)
Voltage
Adjustment 10%
Range
Voltage
5%
Tolerance
Line Regulation 0.5%
Load
1%
Regulation
Setup, Rise
All output shall reach their steady state values within 3 Sec. of turn on
Time
Hold Time 16mS min.
Rated Voltage 100V ~ 240VAC
Frequency 47Hz / 63Hz
Efficiency
(Typ. @ 115V 75% 85% 85% 85% 85% 85%
AC)
AC Current 4.0A @ 120VAC (max.)
Inrush Current Will not exceed 30A / 115VAC, Cold Start, 25C
Leakage
<100 uA
Current
110-130 % rated output power
Overload
Hiccup mode, recovers automatically after fault condition is removed
110% ~ 150% rated output voltage
Over Voltage
Hiccup mode, recovers automatically after fault condition is removed
Short Circuit Hiccup mode, recovers automatically after fault condition is removed
Operational 0C ~ 50C with 290 LFM forced air-cooling and or 80 Watts convection cooled.
Temperature
Range
Storage - 40 ~ +80C
Operational 20 ~ 90% RH non-condensing
Working
Humidity
Storage 5 ~ 95% R.H
Temperature
0.05%/C typical on all outputs
Coefficient
Vibration 3G, 10-25Hz, 15min. Per cycle for each axis (X, Y, Z)
Autec Power Systems Tel: (805)522-0888
www.autec.com
667 E Cochran St, Simi Valley, CA93065
Fax: (805)522-8777
Protection
Environmental Conditions Input Requirements Output Requirements
Requirement 8 April 2015
Page 2
Model SP-M250-09-B SP-M250-12-B SP-M250-20-B SP-M250-24-B SP-M250-48-B SP-M250-58-B*
Safety
nd nd
UL/cUL (60601-1) 2 Ed, DEMKO (EN60601 2 ED), CE and CB - Safety Pending on 58V model
Standards
Withstand
I/P - O/P: 3KVAC (4242 DC), 10mA for 1 minute
Voltage
Isolation
I/P - O/P, I/P - case: 10M Ohms / 500VDC
Resistance
EMI
Conduction EN55022 , FCC Class B
& Radiation
EMS
EN61204, EN61000-3-2, EN61000-3-3
Immunity
MTBF 30K HRS Compliance: MIL-HDBK-217F
Dimension
127*83.82*38.1 (mm) - 5.0*3.3*1.5 (inch)
(L*W*H)
Weight 0.18kg - 0.40Lb
PIN ASSIGNMENT SHIPPING SPECIFICATIONS
AC DC UNITSP PER CARTON CARTON WEIGHT
Series
Connector
SP-M250
Input Output 20.0 12.34 Kgs 27.21 Lbs
PIN 1 Line +V
PIN 2 NC +V
SHIPPING CARTON
WIDTH LENGTH HEIGHT
PIN 3 Neutral +V
DIMENSIONS
Series Inch 14.00 15.00 10.00
PIN 4 NC +V
SP-M250 MM 355.6 381.0 254.0
PIN 5 Ground +V
PIN 6 -V
-V
PIN 7
PIN 8 -V
PIN 9 -V
PIN 10 -V
PIN 11 +S (remote Sense)
PIN 12 -S
Connector Molex 26-60-4050 Molex 26-60-4120
Autec Power Systems Tel: (805)522-0888
www.autec.com
667 E Cochran St, Simi Valley, CA93065
Fax: (805)522-8777
Others Safety & EMC