The MBC201 Series of open-frame medical power supplies, with its wide universal input range 90-264 VAC, is available at 200 W of output power with forced-air cooling and a variety of single output voltages. The MBC Series is designed and approved to the latest medical standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I & Class II applications. These power supplies are ideal for medical, telecom, datacom, industrial equipment and other applications. 5 x 3 x 1.5 Inch Form Factor 200 W with Forced-Air Cooling Approved to EN / IEC 60601-1 High Efficiency > 88% 12 V Fan Output 5 V Standby Output Remote Sense IEC Protection Class Options: Class I: Earthing Tab J4 (no suffix) Class II: No Earthing Tab (-2 suffix) Medical Safety Agency Approvals CE Marked Diagnostic Dialysis Drug Pump Home Health Care Monitoring Portable Equipment 2 MBC201 Series 2 MAX LOAD OUTPUT 1 MODEL CONNECTOR MIN LOAD POWER VOLTAGE CONVECTION 300 LFM MBC201-1005G JST (Header) 26.0 A 130 W 5 VDC 16.67 A 0.0 A MBC201-1T05G Screw Terminal 35.0 A 175 W MBC201-1012G JST (Header) 12 VDC 13.33 A 16.67 A 0.0 A 200 W MBC201-1T12G Screw Terminal MBC201-1015G JST (Header) 15 VDC 10.67 A 13.33 A 0.0 A 200 W MBC201-1T15G Screw Terminal MBC201-1024G JST (Header) 24 VDC 6.67 A 8.33 A 0.0 A 200 W MBC201-1T24G Screw Terminal MBC201-1030G JST (Header) 5.33 A 6.67 A 0.0 A 30 VDC 200 W MBC201-1T30G Screw Terminal MBC201-1048G JST (Header) 48 VDC 3.33 A 4.17 A 0.0 A 200 W MBC201-1T48G Screw Terminal 3 COVER-201-XBC Metal Cover Kit (accessory) 1 Class I products have an Earthing tab. For Class II version (without Earthing tab) add suffix -2 (e.g.: MBC201-1012G-2). 2 Combined output power from V1, VSTBY and VFAN should not exceed the total output power rating. 3 When used in Cover Kit, de-rate output power to 70 % under all operating conditions. Specifications are for nominal input voltage, 25C unless otherwise stated. PARAMETER DESCRIPTION / CONDITION SPECIFICATION Input Voltage Universal 90 264 VAC / 120 390 VDC Input Frequency 47 63 Hz 120 VAC: 2.4 A max. Input Current 230 VAC: 1.2 A max. 120 VAC: 35 A max. Inrush Current 230 VAC: 65 A max. 120 VAC: < 150 A Leakage Current 230 VAC: < 300 A No Load Power 0.8 W PFC converter (variable) 35 to 250 kHz, 90 kHz typical Switching Frequency Resonant converter (variable) 35 to 250 kHz, 90 kHz typical tech.support psbel.com