The MBC75 Series of open frame medical power supplies feature a
wide universal AC input range of 85 V 264 VAC, offering 75 W of
output power in a compact footprint, with a variety of isolated single
output voltages.
The MBC series is designed and approved to the latest Medical
standards (EN/IEC 60601-1), providing 2 x MOPP isolation for Class I
& Class II applications.
These power supplies are ideal for medical, telecom, datacom,
industrial equipment and other applications.
3 x 2 x 1 Inch Form Factor
75 Watts with Convection Cooling
Approved to EN/IEC 60601-1
Efficiencies up to 93%
-40 To 70C Operating Temperature
Dual Fusing
Thermal Shut-Down Feature
2 Million Hours, Telcordia -SR332-Issue 3
Standby Power < 0.3 W
Suitable for BF Applications
Class II Option Available
RoHS Compliant
CE Marked
Diagnostic Home Health Care
Drug Pump Monitoring
Dialysis Portable Equipment
2
MBC75 Series
MBC75-1T12L Screw Terminal
12 V 6.25 A 0.0 A 1%
MBC75-1012L Molex Header
MBC75-1T15L Screw Terminal
5.00 A
15 V 0.0 A 1%
MBC75-1015L Molex Header
MBC75-1T24L Screw Terminal
3.12 A
24 V 0.0 A 1%
MBC75-1024L Molex Header
Screw Terminal
MBC75-1T30L
30 V 2.50 A 0.0 A 1%
MBC75-1030L
Molex Header
Screw Terminal
MBC75-1T48L
48 V 1.56 A 0.0 A 1%
MBC75-1048L
Molex Header
MBC75-1T58L Screw Terminal
58 V 1.29 A 0.0 A 1%
MBC75-1058L Molex Header
3
COVER-120-XBC metal cover kit accessory
Specifications are for nominal input voltage, 25C unless otherwise stated.
4
Universal 85-264 VAC / 390 VDC
Input Voltage
(Derate from 75 W @ 100 VAC to 65 W @ 85 VAC)
Input Frequency 47-63 Hz
115 VAC: 1 A max.
Input Current
230 VAC: 0.5 A max.
No Load Power Typical < 0.3 W
115 VAC: 25 A
Inrush Current 230 VAC: 45 A
264 VAC: 75 A
Typical (N.A. For Class II Option) 300 uA
Leakage Current
Touch current <100 uA
Power Factor @ Full Load, Active PFC > 0.95
Switching Frequency Typical 60 kHz
1
For Class II Option (without input Earth pin) add suffix: -2 (e.g.: MBC75-1012L-2).
2
Ripple is peak to peak with 20 MHz bandwidth and 10 F (Tantalum capacitor) in parallel with a 0.1 F capacitor at rated line voltage
and load ranges.
3
When used in Cover Kit, de-rate output power to 70 % under all operating conditions
4
Functional, not approved.