V10
CFM130M SERIES
130 WATT MEDICAL
AC-DC POWER SUPPLY
WITH PFC
Features
* Universal Input 80~264Vac
* 2x 3 Open Frame Compact Size
* 100W with Natural Convection
* 130W with Fan-Cooled
* No Load Input Power Consumption<150mW
* Active PFC Function
* High Efficiency up to 94%
* Continuous Short Circuit Protection
* Meets 2 MOPP IEC/EN60335-1
* EMI Safety Meets Class & Class
* Operating Altitude 5000m
Ordering information
CFM130MXXX- X
Blank: WAFER
B: Base Cooling
C: with Cover
OUTPUT CURRENT RIPPLE& VOLTAGEA LINE LOAD % EFF.
OUTPUT
MODEL NOISE CCURACY REGULATION REGULATION (typ.)
Natural Fan Cooled
VOLTAGE
NOTE 2 NOTE 1 NOTE 3 NOTE 4 NOTE 5
Convection NOTE 7
CFM130M120 12V 8.34A 10.8A 1% 2% 0.5% 1% 93%
CFM130M240 24V 4.2A 5.4A 1% 2% 0.5% 1% 93%
CFM130M360 36V 2.8A 3.6A 1% 2% 0.5% 1% 94%
CFM130M480 48V 2.1A 2.7A 1% 2% 0.5% 1% 94%
Specifications are subject to change without notice.
1 V10
Specifications Specifications CFM130M Series Derating Curve
INPUT SPECIFICATIONS:
Voltage .................................................................... 80~264Vac
Frequency ....... 47 to 63Hz
Inrush Current . Cold start @25100A max. @240Vac
Input Current .... 100Vac/1.5A max., 240Vac/0.8Amax.
Leakage Current ...... 100uA max.
OUTPUT SPECIFICATIONS:
Holdup Time 20ms min. @115Vac
Short Circuit Protection ....... Hiccup Mode (Auto Recover)
Over Voltage Protection . Auto Recover
Temperature Coefficient ..................... 0.05%/ max.
GENERAL SPECIFICATIONS:
Isolation Input to output = 4000VAC
Operating Temperature ............................ -30 ~ 70 (See Derating Curve)
Storage Temperature ............... -40 ~ 85
Humidity .................. 93% RH max. Non condensing
Cooling .. Natural Convection@100W, 10CFM Air Flow@130W
Switching Frequency 105 KHz Typical
MTBF ..... MIL-HDBK-217F, GB, 25/115VAC . 400 Khrs min.
Altitude 5000m
Dimensions . 3.000x2.000x1.20 inches (76.2x50.8x30.5mm)
-B .... 3.598x2.000x1.299 inches (91.4x50.8x33.0mm)
-C ..... 3.598x2.520x1.358 inches (91.4x64.0x34.5mm)
Weight .. 135g, 170g(-B), 218g(-C)
SAFETY AND EMC:
Emission and Immunity (Ed. 4.0) . EN55011 Class B, IEC61000-3-2
IEC61000-4-2, 3, 4, 5, 6, 8, 11, IEC61000-3-3, FCC Part 18 Class B(note 7)
Safety (Ed. 3.1) .. Class , Class II, IEC60601-1, EN60601-1
NOTE:
UL ANSI/AAMI ES60601-1 1. Voltage accuracy is set at full load.
2. Add a 0.1uF ceramic capacitor and a 10uF E.L. capacitor to output for
ripple & noise measuring @20MHz BW.
3. Line regulation is measured from 100Vac to 240Vac with full load.
4. Load regulation is measured from 10% to 100% full load.
5. Typical efficiency at 230 VAC and full load at 25C.
6. Standard input and output connectors (CN1 and CN2) wafer with
TAIWAN KING PIN TERMINAL PVHI series and mate with JST
housing XH series or equivalent.
7. Requires 10CFM.
Mechanical Specification
All Dimensions In Inches[mm]
Tolerance Inches:x.xxx= 0.02
Millimeters: x.xx = 0.5
CFM130M CFM130M-B CFM130M-C
3.000 [76.20]
3.598 [91.40]
3.598 [91.40]
2.700 [68.58]
0.150 [3.80]
3.354 [85.20]
0.150 [3.81]
1 1
1
2
2
2
6
6
6
5 5
5
4 4
4
3
3
PIN CONNECTION
3
Pin Function
1 ACL
PE
PE
Air Flow PE 2 ACN
Direction
Air Flow
3 -Vout
Direction
4 -Vout
Air Flow
Direction 5 +Vout
+Vout
6
Typical at 25, nominal line and 75% load, unless otherwise Specified
2
5CM
0.150 [3.81]
0.098 [2.50]max 1.700 [43.18]
1.040 [26.40]max 2.000 [50.80]
1.201 [30.50]max
5CM
0.197 [5.00]max
0.157 [4.00]
1.042 [26.46]max
1.380 [35.05]
1.302 [33.06]max
2.000 [50.80]
5CM
1.382 [35.10]
1.358 [34.50]
2.000 [50.80]
2.520 [64.00]