V14 CFM200M SERIES 200 WATT MEDICAL AC-DC POWER SUPPLY WITH PFC Features * Universal Input Range 90~264Vac * 2x 4 Open Frame/CFM200M * 180W with Natural Convection 220Vac/CFM200M * 200W with Natural Convection 220Vac/CFM200MXXXC * Active PFC Meets EN61000-3-2 * No Load Power Consumption<0.3W 3 * High Power Density Up to 16.9W/Inch /CFM200M CFM200MXXXC (Covered type) * +12V Fan Output * High Efficiency up to 93.5% * Meet Class II & Class I * Meets 2 MOPP CFM200MXXX (Open Frame type) RIPPLE VOLTAGE LINE LOAD % EFF. VOLTAGE OUTPUT MODEL & NOISE ACCURACY REG. REG. (Typ.) CURRENT OUTPUT NOTE 1 NOTE 2 NOTE 3 NOTE 4 NOTE 5 Main Output Voltage +12V 16.67A 2.0% 0.5% 1% 92% CFM200M120 150mVp-p CFM200M240 +24V 8.33A 240mVp-p 2.0% 0.5% 1% 93.5% CFM200M480 +48V 4.17A 480mVp-p 2.0% 0.5% 1% 93% Fan Output Voltage 0.5A All +12V NOTE 6 Typical at 25, 230Vac and 60% rated load, unless otherwise specified 1 V14 CFM200M Series De-rating Curve Specifications INPUT SPECIFICATIONS: AC Input Voltage .................................................. 90~264Vac Frequency ..... 47 to 63Hz Inrush Current ........................ 100A max. 240Vac Leakage Current (Earth) 260uA typ., 300uA max. Touch Current (CFM200MXXXC) .. 75uA typ., 100uA max. OUTPUT SPECIFICATIONS: Total Rated Output Power 200W Hold-up Time . 10ms typ. Over Voltage Protection ... Recycle AC input to restart Short Circuit Protection ....... Hiccup mode(Auto Recovery) Temperature Coefficient .... 0.05%/ GENERAL SPECIFICATIONS: Isolation ...... Input to output = 4000VAC Over Temperature Protection Auto Recovery Operating Temperature ..................... - 20 ~ 80 (see Derating Curve) Storage Temperature .................. -40~85 Humidity ................ 93% RH max. Non condensing Switching Frequency .............................................. 85KHz Typical MTBF ..... MIL-HDBK-217F, GB, 25 /115VAC 279Khrs typ. Altitude 3000m Dimensions Open frame versions . 4.000x2.000x1.480 Inches(101.60x50.80x37.60mm) Covered versions . 4.606x2.441x1.575 Inches (117.00x62.00x40.00mm) Weight Open frame versions ..... 253g (0.558 Pounds) Covered versions ..... 314g (0.692 Pounds) SAFETY AND EMC: Emission and Immunity (Ed. 4.0) . EN55011 Class B, FCC CFR 47 pat 18 IEC61000-3-2, 3, IEC61000-4-2, 3, 4, 5, 6, 8, 11 Safety (Ed. 3.1) . IEC60601-1:2005+A1:2012 EN60601-1:2006+A11:2011+A1+A12, UL ANSI/AAMI ES60601-1 NOTE: 1. Add a 0.1uF ceramic capacitor and a 47uF E.L. capacitor to output for ripple &noise measuring 20MHz BW. 2. Voltage accuracy is set at 60% rated load and 25 Ta. 3. Line regulation is measured from High Line to Low Line with rated load. 4. Load regulation is measured from full to 10% rated. 5. Typical efficiency at 230 VAC and full load at 25. 6. 12V/0.3A with nature convection 12V/0.5A with 10 CFM air flow, tolerance 10% at main output 10-100% full load. 7. Need an external 1mH choke at input for Class II type to pass EN55011 Class B. 8. Input connectors (CN1) wafer with TAIWAN KING PIN TERMINAL PVHI series and mate with JST Housing VHR series or equivalent. 9. Fan output connector wafer with TOWNES ENTERPRISE 2001BW series and mate with JST Housing PHR series and JST SPH series crimp terminal or equivalent. 10. Output connectors (Vo+ & Vo- with M3 screw) mate with round terminal and round terminal of the max outer diameter is 6.75mm, max inner diameter is 3.9mm. 2