Additional Resources: Product Page date 06/20/2017 page 1 of 5 SERIES: ETMA 110W U DESCRIPTION: MEDICAL AC-DC POWER SUPPLY FEATURES up to 110 W power universal input (90~264 Vac) single regulated output from 12~24 V over voltage, overload and short circuit protections medical 60601-1 4th edition safety approvals designed for 2 x MOPP applications level V efficiency custom designs available RoHS output output output ripple efficiency MODEL 1 voltage current power and noise level max max max (Vdc) (A) (W) (mVp-p) ETMA120750U 12 7.5 90 120 V ETMA135667U 13.5 6.67 90 135 V ETMA150667U 15 6.67 100 150 V ETMA190580U 19 5.8 110 190 V ETMA200550U 20 5.5 110 200 V ETMA240460U 24 4.6 110 240 V Notes: 1. At full load, 100 ~ 240 Vac input, 20 MHz bandwidth oscilloscope, each output terminated with 10 F aluminum electrolytic and 0.1 F ceramic capacitors. PART NUMBER KEY ETMA120750U C - P51 - WP - CXX Base Number Reserved for Custom example of 12 Vdc, 7.5 A Configurations DC Plug Type Input Cable: Factory C = included Designation blank = no cable cui.comAdditional Resources: Product Page CUI Inc SERIES: ETMA 110W U DESCRIPTION: MEDICAL AC-DC POWER SUPPLY date06/20/2017 page 2 of 5 INPUT parameter conditions/description min typ max units voltage 90 264 Vac frequency 47 63 Hz current at 90 Vac, 50 Hz 2 A inrush current at 230 Vac, cold start 70 A no load power consumption 0.3 W OUTPUT parameter conditions/description min typ max units 19, 20, 24 V outputs 3 % total regulation all other outputs 5 % hold-up time at 115 / 230 Vac, full load 10 ms temperature coefficient 0.05 %/C PROTECTIONS parameter conditions/description 13.5 V output 118 150 % over voltage protection 24 V output 113 150 % all other outputs 120 150 % overload protection shutdown and auto restart 160 % short circuit protection output shutdown and auto restart SAFETY & COMPLIANCE parameter conditions/description min typ max units input to output for 2 seconds 5,656 Vdc isolation voltage input to frame ground for 2 seconds 2,121 Vdc input to output, at 500 Vdc 20 M insulation resistance input to frame ground, at 500 Vdc 20 M safety approvals medical UL/cUL UL 60601-1, EN 60601-1, CB IEC 60601-1 4th edition EMI/EMC FCC Part 18 Class B, EN 60601-1 Class B, EN 55011 Class B leakage current 0.1 mA MTBF at 25C, max. load 100,000 hours RoHS compliant yes ENVIRONMENTAL parameter conditions/description min typ max units operating temperature 0 40 C storage temperature -10 70 C humidity non-condensing 10 90 % cui.com