TECHNICAL DATASHEET MDS Medical AC-DC Open Frame 15 Volt, 100 Watt / MDS-100BPS15 B 100BPS15 B Highlights & Features Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 62368-1 Compliant with IEC 60601-1-2 4th Ed. Requirements Compact size 2x4 Low touch current (<0.1 mA Normal & <0.3 mA single fault) Over-Voltage/Load/Temperature & Short Circuit protections 1 Million Hours MTBF 2 x MOPP (means of patient protection) Safety Standards CB Certified for worldwide use Model Number: MDS-100BPS15 B Unit Weight: 150 grams (5.3 ounces) Dimensions (W x L x H): 50.8 x 101.6 x 31.8 mm 2.0 x 4.0 x 1.25 in General Description The MDS series of embedded power supply come with universal AC input at 90 Vac to 264 Vac. Other features include low touch current, risk management report available and the electric shock protection comply with 2 x MOPP. The MDS series is certified for EMC standards according to EN/BS EN 55011 for industrial, scientific and medical (ISM) radio-frequency equipment and EN/BS EN 55032 for Information Technology Equipment (ITE) radio-frequency equipment. In addition, only recognized Japanese capacitors are used. The MDS series of embedded power supply come with both medical and ITE safety approvals including UL/cUL/CCC/CE and CB certification and are fully compliant with RoHS Directive for environmental protection. Model Information Medical AC-DC Open Frame Model Number Input Voltage Output Voltage Conversion Current Output Forced Air Current Output MDS-100BPS15 B 90-264 Vac 15 Vdc 5.3 A 6.7 A Model Numbering - MDS 100 BPS 15 B Output Voltage Max wattage in the product series. Delta Medical Family Code Maybe lower at some voltage. Single Output: power Supply 15 for 15 V 100 100 W All parameters are specified at 25C ambient unless otherwise noted. 1 www.DeltaPSU.com (November 2021, Rev. 10) TECHNICAL DATASHEET MDS Medical AC-DC Open Frame 15 Volt, 100 Watt / MDS-100BPS15 B Specifications Input Ratings / Characteristics Nominal Input Voltage 100-240 Vac Input Voltage Range 90-264 Vac Nominal Input Frequency 50-60 Hz Input Frequency Range 47-63 Hz Input Current (max) 2.5 A 100 Vac, 1.25 A 240 Vac Efficiency (typ.) 88.2%, Reference Fig. 1 Standby Power (max) 0.3 W Inrush Current (typ.) 40 A 115 Vac, 80 A 230 Vac 1) 2) Touch Current (max) 0.1 mA 264 Vac NC , 0.3 mA 264 Vac SFC 1) 2) Earth Leakage Current (max) 0.15 mA 264 Vac NC , 0.3 mA 264 Vac SFC 1) NC: normal condition 2) SFC: single fault condition 100 95 90 85 80 115Vac Efficiency 75 230Vac Efficiency 70 65 60 55 50 25% 50% 75% 100% Output load Fig.1 Efficiency versus output load Output Ratings / Characteristics Nominal Output Voltage 15 Vdc Total Regulation 1.5% Output Current 6.7 A with 10 CFM force air, 5.3 A for convection Output Power 100 W with 10 CFM force air, 80 W for convection Line Regulation (max) 0.5% Load Regulation (max) 1% Ripple & Noise (typ.) 51.2 mV pk-pk Full load, Reference Fig. 2 Start-up Time (max) 3000 ms 115 Vac Hold-up Time (min) 10 ms 115 Vac Dynamic Response (Overshoot & Undershoot O/P Voltage) 3% 50-100% load *Periodic and Random Deviation All parameters are specified at 25C ambient unless otherwise noted. 2 www.DeltaPSU.com (November 2021, Rev. 10) E fficiency (% )