TECHNICAL DATASHEET
Medical AC-DC Enclosed Power Supply
18V 300W with 5V/0.5A Standby / MDS-300ADB18
MDS-300ADB18
Highlights & Features
Safety Approvals to IEC 60601-1 3.1rd ed. & IEC 60950-1
Compliant with IEC 60601-1-2 4th Ed. Requirements
Up to 190 Watt Convection in 3.5" x 5.5" x 1.75" Package
Up to 800 K Hrs MTBF
Suited for Type BF Medical Products.
2 x MOPP isolation
Safety Standards
CB Certified for worldwide use
Model Number: MDS-300ADB18
Unit Weight: 610 g (1.35 lb)
Dimensions (W x L x H): 88.9 x 140.0 x 45.0 mm
(3.5 x 5.5 x 1.75 inch)
General Description
3
Deltas new MDS-300ADB18 enclosed design offers a high energy density (8.8 Watts/in ) 3.5x5.5 inch design for Type BF patient
access medical products. With operating parameters of 90 to 264 Vac universal input voltage, temperatures of -10 to +70 degrees
centigrade, and altitudes of up to 5000 meters (16,400 feet), the design is well suited for a variety of both medical and non-medical
applications. Other features include input surge of 300 Vac, low leakage current, no-load input power < 0.5 Watt, and 800K hour MTBF.
This product is certified for EMC standards EN 55011 for industrial, scientific and medical (ISM) radio-frequency equipment, and EN
55032 for Industrial Technology Equipment (ITE) radio-frequency equipment. In addition, only recognized Japanese capacitors are
used to ensure long product life.
An enhanced feature set, includes a 5 Volt/0.5 amp standby output, remote on/off, remote sense, and a power good signal that are
included as part of the standard product.
The design, which has protection against shock compliant with 2XMOPP and Type BF requirements, has both medical (with risk report
available), and ITE safety approvals, including cURus (US & Canada)/TUV/GB (China); plus, CB certificates with all national deviations.
Full compliance with RoHS Directive 2011/65/EU for environmental protection is included.
Model Information (All with 5V/0.5A standby available):
Model Number Input Voltage Output Voltage Convection Current Output Forced Air Current Output
MDS-300ADB18 90-264Vac 18Vdc 0-10.55A 0-16.66A (with 10 CFM forced air)
Model Numbering
MDS 300 A D B 18
Max wattage in the
Delta Medical Power Family Code Product Type Input Type Code Output Voltage Revision code,
product series.
Supply A ~ Z D: Enclosed B: 3pin Class I 18 for 18V Refer to page 8
300 300W
All parameters are specified at 25C ambient unless otherwise noted.
1
www.DeltaPSU.com (June 2019, Rev. 06)
TECHNICAL DATASHEET
Medical AC-DC Enclosed Power Supply
18V 300W with 5V/0.5A Standby / MDS-300ADB18
Specifications
Input Ratings / Characteristics
Nominal Input Voltage 100-240Vac
Input Voltage Range 90-264Vac
Nominal Input Frequency 50-60Hz
Input Frequency Range 47-63Hz
Input Current (max) 4A
Input Surge Voltage (max) 300Vac for 100ms
Full load Efficiency (typ.) 92.5% @ 115Vac/60Hz
93.5% @ 230Vac/50Hz, Reference Fig.1
Standby Power (max) 0.5W (Inhibit signal high, 18 volt turned off)
@ 115Vac/60Hz, 230Vac/50Hz
Inrush Current (max) 40A @ 230Vac, cold start
1) 1)
Input-PE(protective earth) leakage current (max) 0.1mA @ NC , 0.3mA @ SFC
1) 1)
Output-PE(protective earth) leakage current for Type BF application (max) 0.1mA @ NC , 0.5mA @ SFC
Power Factor (min) 0.9 @ 115V/50Hz, 230V/50Hz, full load
1) NC: normal condition, SFC: single fault condition
Leakage Current
Input-PE Leakage Current 100Vac/60Hz (Typ) 264Vac/60Hz(Typ) Delta Limit IEC60601-1 Limit
Normal Condition 17.5uA 43.5uA 100uA max 5000uA max
Single Fault Condition 54.5uA 153uA 300uA max 10000uA max
Output-PE Leakage Current for Type BF application
Normal Condition 29.5uA 81.5uA 100uA max 100uA max
Single Fault Condition 43.7uA 129.4uA 500uA max 500uA max
Fig.1 Efficiency versus output load
All parameters are specified at 25C ambient unless otherwise noted.
2
www.DeltaPSU.com (June 2019, Rev. 06)