60W AC-DC High Reliability Medical Adaptor G SM 60B series ( for 5~9V) F e a t u r e s D e s c r i p t i o n File Name:GSM60B-SPEC 2016-03-16 60W AC-DC High Reliability Medical Adaptor G SM 60B series SPECIFICATION ORDER NO. GSM60B05-P1J GSM60B07-P1J GSM60B09-P1J GSM60B12-P1J GSM60B15-P1J GSM60B18-P1J GSM60B24-P1J GSM60B48-P1J SAFETY MODEL NO. GSM60B05 GSM60B07 GSM60B09 GSM60B12 GSM60B15 GSM60B18 GSM60B24 GSM60B48 DC VOLTAGE Note.2 5V 7.5V 9V 12V 15V 18V 24V 48V RATED CURRENT 6A 6A 6A 5A 4A 3.33A 2.5A 1.25A CURRENT RANGE 0 ~ 6A 0 ~ 6A 0 ~ 6A 0 ~ 5A 0 ~ 4A 0 ~ 3.33A 0 ~ 2.5A 0 ~ 1.25A RATED POWER (max.) 30W 45W 54W 60W 60W 60W 60W 60W OUTPUT RIPPLE & NOISE (max.) Note.3 100mVp-p 100mVp-p 100mVp-p 100mVp-p 100mVp-p 150mVp-p 180mVp-p 240mVp-p VOLTAGE TOLERANCE Note.4 5.0% 5.0% 5.0% 3.0% 3.0% 3.0% 3.0% 2.5% LINE REGULATION Note.5 1.0% 1.0% 1.0% 1.0% 1.0% 1.0% 1.0% 1.0% LOAD REGULATION 5.0% 5.0% 5.0% 3.0% 3.0% 3.0% 3.0% 2.5% SETUP, RISE TIME Note.6 1000ms, 30ms / 230VAC 1500ms, 30ms / 115VAC at full load HOLD UP TIME (Typ.) 50ms / 230VAC 15ms / 115VAC at full load VOLTAGE RANGE Note.7 80 ~ 264VAC 120 ~ 370VDC FREQUENCY RANGE 47 ~ 63Hz EFFICIENCY (Typ.) 81.5% 86% 87.5% 88% 88.5% 89% 90% 91.5% INPUT AC CURRENT (Typ.) 1.4A / 115VAC 1A / 230VAC INRUSH CURRENT (Typ.) 30A / 115VAC 65A / 230VAC LEAKAGE CURRENT(max.) Touch current < 50 A/264VAC 105 ~ 160% rated output power OVERLOAD Protection type : Hiccup mode, recovers automatically after fault condition is removed 5.25 ~ 6.75V 7.88 ~ 10.13V 9.45 ~ 12.15V 12.6 ~ 16.2V 15.75 ~ 20.25V 18.9 ~ 24.3V 25.2 ~ 32.4V 50.4 ~ 64.8V PROTECTION OVER VOLTAGE Protection type : Shut down o/p voltage, re-power on to recover OVER TEMPERATURE Shut down o/p voltage, re-power on to recover WORKING TEMP. -30 ~ +60 (Refer toDerating Curv) 20% ~ 90% RH non-condensing WORKING HUMIDITY STORAGE TEMP., HUMIDITY -40 ~ +85, 10 ~ 95% RH ENVIRONMENT TEMP. COEFFICIENT 0.03% / (0 ~ 40) VIBRATION 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes SAFETY STANDARDS ANSI/AAMI ES60601-1 / ES60601-1-11, TUV EN60601-1 / 60601-1-11 approved ISOLATION LEVEL Primary-Secondary: 2xMOPP SAFETY & WITHSTAND VOLTAGE I/P-O/P:4KVAC EMC (Note. 8) ISOLATION RESISTANCE I/P-O/P:100M Ohms / 500VDC / 25/ 70% RH EMC EMISSION Compliance to EN55011(CISPR11) class B, EN61000-3-2,3, FCC PART 15 class B,CAN ICES-3(B)/NMB-3(B) Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3 medical level, criteria A EMC IMMUNITY MTBF 720K hrs min. MIL-HDBK-217F(25) DIMENSION 125*50*31.5mm (L*W*H) OTHERS 0.32Kg 40pcs/13.8Kg/1.05CUFT PACKING PLUG See page 3 Other type available by customer requested CONNECTOR CABLE See page 3 Other type available by customer requested For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on