60W AC-DC Reliable Green Medical Adaptor G SM 60B series Users Manual Video + + ( for 5~9V) Features Description File Name:GSM60B-SPEC 2 0 2 1 - 1 2 - 1 560W AC-DC Reliable Green Medical Adaptor G SM 60B series SPECIFICATION ORDER NO. GSM60B05-P1J GSM60B07-P1J GSM60B09-P1J GSM60B12-P1J GSM60B15-P1J GSM60B18-P1J GSM60B24-P1J GSM60B48-P1J SAFETY MODEL NO. GSM60B05 GSM60B07 GSM60B09 GSM60B12 GSM60B15 GSM60B18 GSM60B24 GSM60B48 DC VOLTAGE Note.2 5V 7.5V 9V 12V 15V 18V 24V 48V RATED CURRENT 6A 6A 6A 5A 4A 3.33A 2.5A 1.25A CURRENT RANGE 0 ~ 6A 0 ~ 6A 0 ~ 6A 0 ~ 5A 0 ~ 4A 0 ~ 3.33A 0 ~ 2.5A 0 ~ 1.25A RATED POWER (max.) 30W 45W 54W 60W 60W 60W 60W 60W OUTPUT RIPPLE & NOISE (max.) Note.3 80mVp-p 80mVp-p 100mVp-p 100mVp-p 100mVp-p 120mVp-p 150mVp-p 200mVp-p VOLTAGE TOLERANCE Note.4 5.0% 5.0% 5.0% 3.0% 3.0% 3.0% 3.0% 2.5% LINE REGULATION Note.5 1.0% 1.0% 1.0% 1.0% 1.0% 1.0% 1.0% 1.0% LOAD REGULATION 5.0% 5.0% 5.0% 3.0% 3.0% 3.0% 3.0% 2.5% SETUP, RISE TIME Note.6 1000ms, 30ms / 230VAC 1500ms, 30ms / 115VAC at full load HOLD UP TIME (Typ.) 50ms / 230VAC 16ms / 115VAC at full load VOLTAGE RANGE Note.7 80 ~ 264VAC FREQUENCY RANGE 47 ~ 63Hz EFFICIENCY (Typ.) 81.5% 86% 87.5% 88% 88.5% 89% 90% 91.5% INPUT AC CURRENT (Typ.) 1.4A / 115VAC 1A / 230VAC INRUSH CURRENT (Typ.) Cold start 30A / 115VAC 60A / 230VAC LEAKAGE CURRENT(max.) Touch current < 50 A/264VAC 105 ~ 160% rated output power OVERLOAD Protection type : Hiccup mode, recovers automatically after fault condition is removed 5.2 ~ 7,0V 7.8 ~ 10.2V 9.4 ~ 12.2V 12.6 ~ 16.2V 15.7 ~ 20.3V 18.9 ~ 24.3V 25.2 ~ 32.4V 50.4 ~ 64.8V PROTECTION OVER VOLTAGE Protection type : Shut down o/p voltage, re-power on to recover Shut down o/p voltage, re-power on to recover OVER TEMPERATURE WORKING TEMP. -30 ~ +70 (Refer toDerating Curv) WORKING HUMIDITY 20% ~ 90% RH non-condensing -40 ~ +85, 10 ~ 95% RH non-condensing ENVIRONMENT STORAGE TEMP., HUMIDITY TEMP. COEFFICIENT 0.03% / (0 ~ 40) VIBRATION 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes OPERATING ALTITUDE Note.8 3000 meters SAFETY STANDARDS IEC60601-1/IEC60601-1-11, BS EN/EN60601-1/ BS EN/EN60601-1-11, ANSI/AAMI ES60601-1 / ES60601-1-11(3.1 version), CAN/CSA-C22.2 No. 60601-1:14 - Edition 3, EAC TP TC 004 approved ISOLATION LEVEL Primary-Secondary: 2xMOPP WITHSTAND VOLTAGE I/P-O/P:4KVAC I/P-O/P:100M Ohms / 500VDC / 25/ 70% RH ISOLATION RESISTANCE Parameter Standard Test Level / Note BS EN/EN55011 (CISPR11), FCC PART 15 / Conducted emission Class B CISPR22, CAN ICES-3(B)/NMB-3(B) EMC EMISSION BS EN/EN55011 (CISPR11), FCC PART 15 / Radiated emission Class B CISPR22, CAN ICES-3(B)/NMB-3(B) Harmonic current BS EN/EN61000-3-2 Class A SAFETY & Voltage flicker BS EN/EN61000-3-3 ----- EMC BS EN/EN60601-1-2, BS EN/EN61204-3 (Note. 9) Parameter Standard Test Level / Note ESD BS EN/EN61000-4-2 Level 4, 15KV air Level 4, 8KV contact Level 3, 10V/m( 80MHz~2.7GHz ) RF field susceptibility BS EN/EN61000-4-3 Table 9, 9~28V/m( 385MHz~5.78GHz ) EFT bursts BS EN/EN61000-4-4 Level 3, 2KV EMC IMMUNITY BS EN/EN61000-4-5 Surge susceptibility Level 3, 1KV/Line-Line Conducted susceptibility BS EN/EN61000-4-6 Level 3, 10V Magnetic field immunity BS EN/EN61000-4-8 Level 4, 30A/m 100% dip 1 periods, 30% dip 25 periods, Voltage dip, interruption BS EN/EN61000-4-11 100% interruptions 250 periods MTBF 720K hrs min. MIL-HDBK-217F(25) DIMENSION 125*50*31.5mm (L*W*H) OTHERS 0.32Kg 40pcs/13.8Kg/1.04CUFT PACKING PLUG See page 4~5 Other type available by customer requested CONNECTOR CABLE See page 4~5 Other type available by customer requested For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on