1~4 Output Medical Type 120W M edical series Features Universal AC input / Full range Low leakage current <180 A Protections: Short circuit / Overload / Over voltage UL60601-1 medical safety approved With power good and fail signal output (Optional) 100% full load burn-in test Fixed switching frequency at 45KHz 3 years warranty EN60601-1 IEC60601-1 Specification AC INPUT VOLTAGE TEMP. COEFFICIENT 90~264 VAC, 47~440Hz / 127~370VDC. 0.04%/ on all outputs at full load between 0~50 of ambient temperature. AC INPUT CURRENT (Typ.) VIBRATION Maximum input current 2.9A at 115VAC, 60Hz or 2G of acceleration, vibrating frequency adjust from 10Hz ~500Hz within a 1.7A at 230VAC, 60Hz with 100% output load. 10-minute cycle. 6 testing cycles (60 minutes) each along X, Y, Z axes. INRUSH CURRENT (Typ.) SAFETY STANDARDS Inrush current is less than 22A at 115VAC or less than Medical : UL60601-1, TUV EN60601-1, IEC60601-1 approved 45A at 230VAC under cold start conditions. Limiting Commercial : Also design refer to UL60950-1, TUV EN60950-1 provided by internal thermistors. WITHSTAND VOLTAGE SETUP, RISE TIME 4000VAC between input and output MPS-120: 800ms, 20ms / 230VAC at full load 1500VAC between input and F.G. 2000ms, 50ms / 115VAC at full load 500VAC between output and F.G. MPD,T,Q-120:500ms, 20ms / 230VAC at full load 1200ms, 50ms / 115VAC at full load ISOLATION RESISTANCE >100M Ohms for I/P-O/P, I/P-FG, O/P-FG by using 500VDC test voltage. HOLD-UP TIME (Typ.) 80ms / 230VAC at full load EMI COMPLIANCE 14ms / 115VAC at full load EMI Specifications Compliance Level Conducted & Radiation EN55011, Class B LEAKAGE CURRENT EN55022, Class B Leakage current is less than 180 A at 264VAC Harmonic distortion EN61000-3-2 Voltage flicker EN61000-3-3 DC OUTPUT ADJ. RANGE DC output voltage (or Ch1 of multiple output models) EMS COMPLIANCE can be adjusted between -5%~+10% rated output Compliance Level EMS Specification voltage by potential meter. EN61000-4-2, Level 3, 8KV ESD air EN61000-4-2, Level 2, 4KV ESD contact OVERLOAD PROTECTION EN61000-4-3, RF field susceptibility Level 2, 3V/m Fully protected against short circuit and output overload. Level 3, 10V/m The hiccup type protection will be activated at 110~150% EFT(Electrical Fast Transient)/Burst EN61000-4-4, Level 2, 1KV/5KHz (For MPD,T,Q-120),120~160%(For MPS-120) rated load Level 3, 2KV/5KHz and recovers automatically after fault condition is removed. Lightning/Surge EN61000-4-5, Level 4, 2KV/Line-Line 4KV/Line-Earth OVER VOLTAGE PROTECTION Conducted RF susceptibility EN61000-4-6, Level 2, 3Vrms/m Provided on output channel 1 only at 115%~135% Level 3, 10Vrms/m rated output voltage. (120%~140% for MPS-120-15/24/48). Magnetic field immunity EN61000-4-8, Level 2, 3A/m Output will be shut down when this protection is activated. Level 3, 10A/m Voltage dip, interruption EN61000-4-11,Compliance POWER GOOD / FAIL SIGNAL (OPTIONAL) Digital phone carrier immunity ENV50204, Level 2, 3V/m, 900MHz TTL logic high for power good and TTL low for Level 3, 10A/m, 900MHz power fail. When the output voltage reaches 90% of MTBF rated value, a +5V TTL signal will be sent out with a 262,100 hours min. at full load and 25 of ambient temperature, 10~500ms delay At least 1ms before the output calculated per MIL-HDBK-217F. voltage goes below 90% of the rated value, the TTL signal will be turned off. DIMENSION (L*W*H) * MPS-120-3.3 does not have this optional function. 177.8x107.95x35.5mm or 7 x4.25 x1.4 WORKING TEMP. PACKING Whole series can operate from -20~70 . Please refer 0.55Kg 24pcs/14.5Kg/0.99CUFT to the derating curves. WORKING HUMIDITY 20~90% RH non-condensing. STORAGE TEMP., HUMIDITY -40~+85 , 10~90% RH File Name:120W MEDICAL-SPEC 2011-03-311~4 Output Medical Type 120W M edical series Output Chart OUTPUT CURRENT RIPPLE & NOISE VOLTAGE OUTPUT RATED LINE LOAD PEAK LOAD MODEL (Max.) TOLERANCE EFFICIENCY VOLTAGE CURRENT MINIMUM CONVECTION WITH FAN REGULATION REGULATION WITH 25CFM FAN (Note 2) (Note 3) LOAD (25CFM) (max.) (Note 4) MPS-120-3.3 3.3V 24A 0A 16A 24A 26A 80mVp-p 3.0% 1.0% 3.0% 68% MPS-120-5 5V 22A 0A 14.7A 22A 26A 80mVp-p 3.0% 1.0% 3.0% 73% MPS-120-12 12V 10A 0A 6.7A 10A 11A 100mVp-p 2.0% 1.0% 2.0% 77% MPS-120-15 15V 8A 0A 5.3A 8A 8.8A 100mVp-p 2.0% 1.0% 2.0% 79% MPS-120-24 24V 5A 0A 3.3A 5A 5.5A 120mVp-p 2.0% 1.0% 2.0% 81% MPS-120-48 48V 2.5A 0A 1.7A 2.5A 2.8A 120mVp-p 2.0% 1.0% 2.0% 82% MPD-120A 5V 10A 2A 7.3A 10A 12A 80mVp-p 2.0% 0.5% 0.5% 75% 12V 5A 0.5A 3.6A 5A 6A 120mVp-p 7.0% 2.0% 3.5% MPD-120B 5V 10A 2A 7A 10A 12A 80mVp-p 2.0% 0.5% 0.5% 76% 24V 2.9A 0.3A 1.9A 2.9A 3.2A 250mVp-p 8.0% 2.0% 4.0% MPT-120A 5V 10A 2A 7.3A 10A 12A 80mVp-p 2.0% 0.5% 1.0% 12V 4.8A 0.4A 3.5A 4.8A 5.8A 120mVp-p +8,-6% 1.5% 3.5% 72% -5V 0.6A 0A 0.3A 0.6A 1A 80mVp-p 5.0% 0.5% 1.0% MPT-120B 5V 10A 2A 7.3A 10A 12A 80mVp-p 2.0% 0.5% 1.0% 12V 4.4A 0.4A 3.2A 4.4A 5.3A 120mVp-p 6.0% 1.5% 3.5% 73% -12V 0.6A 0A 0.4A 0.6A 1A 80mVp-p 5.0% 0.5% 1.0% MPT-120C 5V 10A 2A 7A 10A 11A 80mVp-p 2.0% 0.5% 1.0% 15V 4A 0.4A 2.6A 4A 4.4A 150mVp-p +6,-7% 2.0% 3.5% 72% -15V 0.6A 0A 0.4A 0.6A 1A 80mVp-p 5.0% 0.5% 1.0% MPT-120D 5V 10A 2A 7.3A 10A 12A 80mVp-p 2.0% 0.5% 1.0% 24V 2.2A 0.4A 1.6A 2.2A 2.64A 300mVp-p +8,-6% 3.0% +4,-3% 74% 12V 0.6A 0A 0.4A 0.6A 1A 120mVp-p 5.0% 0.5% 1.0% MPQ-120B 5V 10A 2A 7.3A 10A 11A 80mVp-p 2.0% 0.5% 0.5% 12V 4.2A 0.5A 3.1A 4.2A 5A 120mVp-p 6.0% 1.5% 3.5% 71% -5V 0.6A 0A 0.4A 0.6A 1A 80mVp-p 5.0% 0.5% 1.0% -12V 0.6A 0A 0.4A 0.6A 1A 80mVp-p 5.0% 0.5% 1.0% MPQ-120C 5V 10A 2A 7.3A 10A 11A 80mVp-p 2.0% 0.5% 0.5% 15V 3.2A 0.5A 2.4A 3.2A 3.8A 150mVp-p +6,-7% 2.0% 3.5% 71% -5V 0.6A 0A 0.4A 0.6A 1A 80mVp-p 5.0% 0.5% 1.0% -15V 0.6A 0A 0.4A 0.6A 1A 80mVp-p 5.0% 0.5% 1.0% MPQ-120D 5V 10A 2A 7A 10A 11A 80mVp-p 2.0% 0.5% 0.5% 12V 1A 0.2A 0.7A 1A 1.1A 150mVp-p +8,-6% 2.0% 3.5% 74% 24V 2.1A 0.3A 1.4A 2.1A 2.3A 300mVp-p 8.0% 2.0% 3.5% -12V 0.6A 0A 0.3A 0.6A 1A 80mVp-p 5.0% 0.5% 1.0% MPQ-120E 5V 10A 2A 7.3A 10A 11A 80mVp-p 2.0% 0.5% 0.5% 12V 3A 0.5A 2.3A 3A 3.3A 120mVp-p 6.0% 2.0% 3.0% 73% 15V 0.6A 0A 0.4A 0.6A 1A 80mVp-p 8.0% 2.0% 3.0% 24V 0.6A 0A 0.4A 0.6A 1A 80mVp-p 5.0% 0.5% 1.0% Notes : 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature. 2. Ripple & noise are measured at 20MHz of bandwidth by using a 12 twisted pair-wire terminated with a 0.1 f & 47 f parallel capacitor. 3. Tolerance : includes set up tolerance, line regulation and load regulation. 4. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power. 5. The power supply is considered a component which will be installed into a final equipment. The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on