45W Triple Output Medical Type MPT-45 series Users Manual % & % ( & ) * + , % - ( % . / 012 3 =5 +-+ A= +-+ D SPECIFICATION MODEL MPT-45A MPT-45B MPT-45C OUTPUT NUMBER CH1 CH2 CH3 CH1 CH2 CH3 CH1 CH2 CH3 DC VOLTAGE 5V 12V -5V 5V 12V -12V 5V 15V -15V RATED CURRENT 3A 2A 0.3A 3A 2A 0.3A 3A 1.6A 0.3A CURRENT RANGE 0.4 ~ 5A 0.2 ~ 2.5A 0 ~ 0.5A 0.4 ~ 5A 0.2 ~ 2.5A 0 ~ 0.5A 0.4 ~ 5A 0.2 ~ 2.3A 0 ~ 0.5A RATED POWER 40.5W 42.6W 43.5W OUTPUT POWER (max.) 52W with 18CFM min. Forced air convection OUTPUT RIPPLE & NOISE (max.) Note.2 60mVp-p 120mVp-p 60mVp-p 60mVp-p 120mVp-p 100mVp-p 60mVp-p 120mVp-p 100mVp-p VOLTAGE ADJ. RANGE CH1:4.5 ~ 5.5V VOLTAGE TOLERANCE Note.3 4.0% 7.0% 5.0% 4.0% 7.0% 5.0% 4.0% 7.0% 5.0% LINE REGULATION 1.0% 2.0% 1.0% 1.0% 2.0% 1.0% 1.0% 2.0% 1.0% LOAD REGULATION 3.0% 4.0% 1.0% 3.0% 4.0% 1.0% 3.0% 4.0% 1.0% SETUP, RISE TIME 800ms, 20ms/230VAC 800ms, 20ms/115VAC at full load HOLD UP TIME (Typ.) 80ms/230VAC 12ms/115VAC at full load VOLTAGE RANGE 90 ~ 264VAC 127 ~ 370VDC FREQUENCY RANGE 47 ~ 63Hz EFFICIENCY(Typ.) 73% 75% 75% INPUT AC CURRENT (Typ.) 1.2A/115VAC 0.7A/230VAC INRUSH CURRENT (Typ.) COLD START 20A/115VAC 40A/230VAC LEAKAGE CURRENT Note.7 Earth leakage current < 250 A/264VAC , Touch current < 60 A/264VAC 53 ~ 75W rated output power OVERLOAD Protection type : Hiccup mode, recovers automatically after fault condition is removed PROTECTION 5.75 ~ 6.75VDC on CH1 OVER VOLTAGE Protection type : Hiccup mode, recovers automatically after fault condition is removed -10 ~ +60 (Refer toDerating Curv) WORKING TEMP. 20 ~ 90% RH non-condensing WORKING HUMIDITY STORAGE TEMP., HUMIDITY -20 ~ +85, 10 ~ 95% RH ENVIRONMENT TEMP. COEFFICIENT 0.04%/ (0 ~ 50) on +5V output 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes VIBRATION ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1 approved SAFETY STANDARDS ISOLATION LEVEL Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP SAFETY & WITHSTAND VOLTAGE I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:0.5KVAC 1min. EMC (Note 4) ISOLATION RESISTANCE I/P-O/P, I/P-FG:100M Ohms / 500VDC / 25/ 70% RH EMC EMISSION Compliance to EN55011 (CISPR11) Class B, EN61000-3-2,-3 EMC IMMUNITY Compliance to EN61000-4-2,3,4,5,6,8,11, EN60601-1-2, medical level, criteria A MTBF 271.5Khrs min. MIL-HDBK-217F (25) OTHERS DIMENSION 127*76*28mm (L*W*H) PACKING 0.21Kg 72pcs/18Kg/1.28CUFT +4 ) 5 6 ) ) 3 7 8 % ) ) 4 NOTE 4 9 : ) &12 % + - ) 4+ % : /< % 4 34 6 8 4 /4 6 ) % . ) 4 =& ( ) 3 ))>3 )) ) +)) % 4 6 % . ) ) - % ) ) =& 4 % ) =& 8 % =&A % ) 4B (as available on 4-3.8 45W Triple Output Medical Type MPT-45 series Mechanical Specification Unit:mm 127 115.8 5.6 HS1 M1 CN2 1 FS1 2 AC FUSE 3 T2A/250V 4 1 5 HS2 6 2 CN1 FS2 AC FUSE SVR1 T2A/250V AC Input Connector (CN1) : Molex 5277-02 or equivalent DC Output Connector (CN2) : Molex 5273-06 or equivalent Pin No. Assignment Mating Housing Terminal Pin No. Assignment Mating Housing Terminal 1 AC/L 1 +V Molex 5194 Molex 5195 or equivalent or equivalent 2 AC/N 2,3 +5V Molex 5195 Molex 5194 or equivalent or equivalent 4,5 COM 1 +81 6 -V : Grounding Required 1.HS1,HS2 cannot be shorted 2.M1 is safety ground Block Diagram fosc : 45KHz RECTIFIERS & +V RECTIFIERS FILTER POWER EMI I/P & RECTIFIERS SWITCHING FILTER FILTER & +5V FILTER RECTIFIERS COM & -V FG FILTER DETECTION CIRCUIT PWM O.L.P. CONTROL O.V.P. Derating Curve Static Characteristics 100 100 90 80 80 70 60 60 40 50 20 40 -10 0102030405060 (HORIZONTAL) 90 100 110 135 155 230 264 AMBIENT TEMPERATURE () INPUT VOLTAGE (V) 60Hz File Name:MPT-45-SPEC 2021-09-15 LOAD (%) 76 3 max. 28 64.8 5.6 LOAD(%)