60W Dual Output Medical Type RPD-60 series 4 - +) ) % 1 94 ( < & > ( A B - = 9 1 9 1 1 ( ( 1 1 8 0 BB & : < C * 963 C * 93 C * 6 3 1 4 0 & 2 D % = E4 & & .. SPECIFICATION MODEL RPD-60A RPD-60B OUTPUT NUMBER CH1 CH2 CH1 CH2 DC VOLTAGE 5V 12V 5V 24V RATED CURRENT 5A 2A 3.5A 1.5A CURRENT RANGE 0.5 ~ 5.5A 0.1 ~ 2.2A 0.5 ~ 3.85A 0.1 ~ 1.65A RATED POWER 49W 53.5W PEAK LOAD(10sec.) Note.4 53.9W 58.85W OUTPUT RIPPLE & NOISE (max.) Note.2 80mVp-p 80mVp-p 80mVp-p 100mVp-p VOLTAGE TOLERANCE Note.3 +3,-2% 6.0% +3,-2% +8,-4% LINE REGULATION 0.5% 1.0% 0.5% 1.0% LOAD REGULATION 1.5% 2.0% 1.5% 2.0% SETUP, RISE TIME 300ms, 15ms/230VAC 300ms, 15ms/115VAC at full load HOLD UP TIME (Typ.) 70ms/230VAC 14ms/115VAC at full load VOLTAGE RANGE 90 ~ 264VAC 127 ~ 370VDC FREQUENCY RANGE 47 ~ 63Hz EFFICIENCY (Typ.) 78% 82% INPUT AC CURRENT (Typ.) 1.1A/115VAC 0.7 A/230VAC INRUSH CURRENT (Typ.) COLD START 60A/230VAC 30A/115VAC LEAKAGE CURRENT Note.8 Earth leakage current < 150A/264VAC , Touch current < 100A/264VAC 115 ~ 150% rated output power OVERLOAD Protection type : Hiccup mode, recovers automatically after fault condition is removed PROTECTION CH1: 5.75 ~ 6.75V OVER VOLTAGE Protection type : Shut down o/p voltage, re-power on to recover -20 ~ +65 (Refer toDerating Curv) WORKING TEMP. 20 ~ 90% RH non-condensing WORKING HUMIDITY STORAGE TEMP., HUMIDITY -40 ~ +85, 10 ~ 95% RH ENVIRONMENT TEMP. COEFFICIENT 0.03%/ (0 ~ 45) VIBRATION 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes UL60950-1, TUV EN60950-1, ANSI/AAMI ES60601-1, TUV EN60601-1, IEC60601-1 approved SAFETY STANDARDS ISOLATION LEVEL Primary-Secondary: 2xMOPP, Primary-Earth:1xMOPP, Secondary-Earth:1xMOPP SAFETY & WITHSTAND VOLTAGE I/P-O/P:4KVAC I/P-FG:2KVAC O/P-FG:1.5KVAC EMC (Note 5) ISOLATION RESISTANCE I/P-O/P, I/P-FG, O/P-FG:100M Ohms / 500VDC / 25/ 70% RH EMC EMISSION Compliance to EN55011(CISPR11),EN55022 (CISPR22) Class B, EN61000-3-2,-3 EMC IMMUNITY Compliance to EN61000-4-2,3,4,5,6,8,11, EN55024, EN60601-1-2, EN61204-3, medical level, criteria A MTBF 677.8K hrs min. MIL-HDBK-217F (25) OTHERS DIMENSION 101.6*50.8*29mm (L*W*H) 0.15Kg 96pcs/15.4Kg/0.89CUFT PACKING NOTE % & ( ) & *& & +, - ( ( + . / 0 & 1 1 ( & 0 & & & & 2 2 * 3 1 4 ( & 3 53 6 ( & 7 8 1 -99&&& & : < ( ( 9 44 & , = > = = & / File Name:RPD-60-SPEC 2014-12-1360W Dual Output Medical Type RPD-60 series Mechanical Specification Unit:mm 101.6 95.25 3.175 FS1 AC FUSE HS2 T2.5/250V 3 1 CN1 2 2 1 3 4 FS2 5 AC FUSE 6 T2.5/250V CN2 HS1 4-3.3 AC Input Connector (CN1) : JST B3P-VH or equivalent DC Output Connector (CN2) : JST B6P-VH or equivalent Pin No. Assignment Mating Housing Terminal Pin No. Assignment Mating Housing Terminal 1 AC/N 1,2 V1 JST VHR JST SVH-21T-P1.1 2 No Pin 3,4 COM JST VHR JST SVH-21T-P1.1 or equivalent or equivalent or equivalent or equivalent 3 AC/L 5 V2 6 NC = = : Grounding required Block Diagram fosc : 100KHz RECTIFIERS V2 & FILTER RECTIFIERS RECTIFIERS EMI POWER +V1 I/P & & FILTER SWITCHING COM FILTER FILTER DETECTION FG CIRCUIT CONTROL O.L.P. O.V.P. Derating Curve Output Derating VS Input Voltage 100 100 90 80 80 60 70 60 40 50 20 40 -20 0 10 20 30 40 50 60 65 70 (HORIZONTAL) 90 95 100 115 120 140 160 180 200 220 240 264 AMBIENT TEMPERATURE () INPUT VOLTAGE (VAC) 60Hz File Name:RPD-60-SPEC 2014-12-13 LOAD (%) 50.8 3 max. 29 44.45 3.175 LOAD (%)