45W Reliable Green Medical Power Supply RPS-45 series Users Manual + + 0 5.5 4 & () * + ,) - ../ % 0 Suitable for BF application with appropriate system consideration Cooling by free air convection EMI class B for class configuration No load power consumption<0.1W Extremely low leakage current Protections: Short circuit / Overload / Over voltage 1 + 2 3 K hours 4 & Description RPS-45 is a 45W highly reliable green PCB type medical power supply with a high power density on the 3 by 2 footprint. It accepts 80~264VAC input and offers various output voltages between 3.3V and 48V. The working efficiency is up to 91% and the extremely low no load power consumption is down below 0.1W. RPS-45 is able to be used for Class(no FG) system design. The extremely low leakage current is less than 100 A. In addition, it conforms to international medical regulations (2*MOPP) and EMC BS EN/EN55011, perfectly fitting all kinds of BF rated patient contact medical system equipment. File Name:RPS-45-SPEC 2021-09-1545W Reliable Green Medical Power Supply RPS-45 series SPECIFICATION ORDER NO. RPS--45 3.3 RPS--45 5 RPS--45 75. RPS--45 12 RPS--45 15 RPS--45 24 RPS--45 48 DC VOLTAGE 3.3V 5V 7.5V 12V 15V 24V 48V RATED CURRENT 8A 8A 5.4A 3.8A 3A 1.9A 0.94A CURRENT RANGE 0 ~ 8.8A 0 ~ 8.8A 0 ~ 5.95A 0 ~ 4.18A 0 ~ 3.3A 0 ~ 2.1A 0 ~ 1.03A RATED POWER 26.4W 40W 40.5W 45.6W 45W 45.6W 45.1W OUTPUT PEAK LOAD(10sec.) Note.2 29W 44W 44.6W 50.2W 49.5W 50.2W 49.4W RIPPLE & NOISE (max.) Note.3 60mVp-p 60mVp-p 80mVp-p 100mVp-p 100mVp-p 120mVp-p 120mVp-p VOLTAGE ADJ.RANGE 3.1~3.6V 47.~.V55 7.12~8.3V 11.4~13.2V 13.5~16.5V 22.8~27.6V 45.6~52.8V VOLTAGE TOLERANCE Note.4 2.0% 2.0% 2.0% 2.0% 1.0% 1.0% 1.0% LINE REGULATION 0.5% 0.5% 0.5% 0.5% 0.5% 0.5% 0.5% LOAD REGULATION 2.0% 2.0% 2.0% 2.0% 1.0% 1.0% 1.0% SETUP, RISE TIME 500ms, 30ms / 230VAC 500ms, 30ms / 115VAC at full load HOLD UP TIME (Typ.) 30ms / 230VAC 16ms / 115VAC at full load VOLTAGE RANGE Note.5 80 ~ 264VAC FREQUENCY RANGE 47 ~ 63Hz INPUT EFFICIENCY (Typ.) 80.5% 83% 85% 88% 89% 90% 91% AC CURRENT (Typ.) 1.2A / 115VAC 1A / 230VAC INRUSH CURRENT (Typ.) OLD STAR 30A/115VAC 60A/230VAC LEAKAGE CURRENT(max.) Note.6 Touch current< 100 A/264VAC 115 ~ 150% rated output power OVERLOAD Protection type : Hiccup mode, recovers automatically after fault condition is removed PROTECTION 3.8~5V 5.7~6.8V 8.6~11.3V 13.8~16.2V 17.2~20.3V 28.4~32.4V 55.2~64.8V OVER VOLTAGE Protection type : Shut down o/p voltage, re-power on to recover WORKING TEMP. -30 ~ +70 (Refer toDerating Curv) WORKING HUMIDITY 20% ~ 90% RH non-condensing STORAGE TEMP., HUMIDITY -40 ~ +85, 10 ~ 95% RH non-condensing ENVIRONMENT TEMP. COEFFICIENT 0.03% /(0 ~ 50 ) VIBRATION 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes OPERATING ALTITUDE Note.7 4000 meters IEC60601-1, TUV BS EN/EN60601-1, EAC TP TC 004,UL ANSI / AAMI ES60601-1 (3.1 version), SAFETY STANDARDS CAN/CSA-C22.2 No. 60601-1:14 - Edition 3 approved Design refer to BS EN/EN60335-1 ISOLATION LEVEL Primary-Secondary: 2xMOPP WITHSTAND VOLTAGE I/P-O/P: 4KVAC ISOLATION RESISTANCE I/P-O/P:100M Ohms / 500VDC / 25/ 70% RH Parameter Standard Test Level / Note BS EN/EN55011 (CISPR11) Class B Conducted emission EMC EMISSION Radiated emission BS EN/EN55011 (CISPR11) Class B SAFETY & Harmonic current BS EN/EN61000-3-2 Class A EMC BS EN/EN61000-3-3 ----- (Note. 8) Voltage flicker BS EN/EN60601-1-2 Parameter Standard Test Level / Note ESD BS EN/EN61000-4-2 Level 4, 15KV air Level 4, 8KV contact Level 3, 10V/m( 80MHz~2.7GHz ) RF field susceptibility BS EN/EN61000-4-3 Table 9, 9~28V/m( 385MHz~5.78GHz ) EFT bursts BS EN/EN61000-4-4 Level 3, 2KV EMC IMMUNITY Surge susceptibility BS EN/EN61000-4-5 Level 4, 2KV/Line-Line BS EN/EN61000-4-6 Conducted susceptibility Level 3, 10V Magnetic field immunity BS EN/EN61000-4-8 Level 4, 30A/m 100% dip 1 periods, 30% dip 25 periods, Voltage dip, interruption BS EN/EN61000-4-11 100% interruptions 250 periods MTBF 726.2Khrs min. MIL-HDBK-217(25) DIMENSION (L*W*H) 76.2*50.8*24mm or 3 * 2 *0.945 inch OTHERS 0.11Kg 120pcs/14.2Kg/0.94CUFT PACKING 1. All parameters NOT specially mentioned are measured at 230VAC input, rated load and 25 of ambient temperature. 2. 33% Duty cycle maximum within every 30 seconds. Average output power should not exceed the rated power. 3. Ripple & noise are measured at 20MHz of bandwidth by using a 12 twisted pair-wire terminated with a 0.1 f & 47 f parallel capacitor. 4. Tolerance : includes set up tolerance, line regulation and load regulation. 5. Derating may be needed under low input voltages. Please check the derating curve for more details. 6. Touch current was measured from primary input to DC output. NOTE 7. 5% 6 + &73 % + + 3 % + + % % % ) , 3 + /7 8. The power supply is considered a component which will be installed into a final equipment.All the EMC tests are been executed by mounting the unit on a 360mm*360mm metal plate with 1mm of thickness The final equipment must be re-confirmed that it still meets EMC directives. For guidance on how to perform these EMC tests, please refer to EMI testing of component power supplies. (as available on