D C C S S a O O R R L E E M M p T T P P O R R L L p I I E E A A H V V l N N N N i T T E Medical certified 2MOPP module, BF ready Features Class II installations (without FG) AC/DC Converter IP68 waterproof encapsulation Operation altitude certified up to 5000m Regulated No external components necessary Converter Energy Efficiency Level VI RACM18-ER Description 18 Watt The RACM18-ER series comprises highly reliable power conversion modules in a potted IP68 3 Round waterproof encapsulation to withstand harsh operating conditions. With a certified operation up to 5000m altitude and a temperature range from -20C up to +80C these modules are built to power medical healthcare, household, sanitary, smart building and automation process appliances. For easy Shape Single integration the product line is certified to medical, household, as well as to ITE safety standards and designed to pass class B limits by more than 6dB margin without the need for any external components. Output Selection Guide Part Input Output Output Efficiency (1) (2) Number Voltage Range Voltage Current typ. VAC VDC A % (3) RACM18-05SER 90-264 5 2.5 81 (3) RACM18-12SER 90-264 12 1.5 82 (3) RACM18-24SER 90-264 24 0.75 83.5 2+ Notes: RoHS REACH ERP compliant compliant compliant 10 from 10 Note1: Other output voltages on request Note2: Efficiency is tested at nominal input (115/230VAC) and full load at +25C ambient CERTIFIED EN-60335-1 REBA VEBA 2MOPP 280VAC Model Numbering (3) RACM18- SER PREFERRED ALTERNATIVES Please consider this alternatives: Output Power Single Output Voltage RACM40-K Series Notes: Note3: Other connection types on reqeuest IEC/EN60950-1 certified UL60950-1 certified ANSI/AAMI ES60601-1 certified IEC/EN60601-1 certified IEC/EN60335-1 certified IEC/EN61558-1 certified IEC/EN61558-2-16 certified IEC/EN60601-1-2 certified EN55024/32 certified EN55014-1 (-2) certified CISPR32 certified www.recom-power.com REV.: 3/2020 MED-1 O O a P P C C S R R n O O M M U D D c O O U U O C C C C e E E T T R R H s S SRACM18-ER AC/DC Converter Series Specifications (measured ta= 25C, nom. Vin (115/230VAC), full load after warm-up unless otherwise stated) BASIC CHARACTERISTICS Parameter Condition Min. Typ. Max. Internal Input Filter Pi type Input Voltage Range 90VAC 230VAC 264VAC 115VAC 500mA Input Current 230VAC 150mA 115VAC 24A Inrush Current 230VAC 46A No load Power Consumption 40mW 75mW Input Frequency Range 47Hz 63Hz Minimum Load 0% 0.46 Power Factor 115VAC 180ms Start-up Time 230VAC 200ms Rise Time 115VAC/230VAC 15ms 115VAC 15ms Hold-up Time 230VAC 65ms Internal Operating Frequency 100% load at nominal Vin 100kHz 20MHz BW 140mVp-p Output Ripple and Noise Efficiency vs. Load RACM18-12SER RACM18-24SER 100 100 90 90 80 80 70 70 60 60 50 50 40 40 30 30 85VAC 85VAC 20 20 115VAC 115VAC 10 10 230VAC 230VAC 264VAC 264VAC 0 0 0 25 50 75 100 0 25 50 75 100 Output Load % Output Load % REGULATIONS Parameter Condition Value Output Accuracy 3.0% max. Line Regulation low line to high line 1.0% max. Load Regulation 0% to 100% load 1.0% max. Transient Response 100% load step change 3.0% max. continued on next page www.recom-power.com REV.: 3/2020 MED-2 Efficiency % Efficiency %