D C C S S a O O R R L E E M M p T T P P O R R L L p I I E E A A H V V l N N N N i T T E Medical certified 2MOPP module, BF ready Features Class II installations (without FG) AC/DC Converter IP68 waterproof encapsulation Operation altitude certified up to 5000m Regulated No external components necessary Converter Energy Efficiency Level VI RACM30-ER Description 30 Watt The RACM30-ER series comprises reliable and highly efficient power conversion modules in a 3 Round potted IP68 waterproof encapsulation to withstand harsh operating conditions. With a certified operation up to 5000m altitude and less than 2x2 of required board space, these modules are built to power compact applications in medical healthcare, household, sanitary, smart building, and Shape Single automation appliances. For easy integration the series are covered by certifications for medical, household, and ITE safety standards and designed to pass class B limits by more than 6dB margin Output without the need for any external components. Selection Guide Part Input Output Output Efficiency (1) (2) Number Voltage Range Voltage Current typ. VAC VDC A % (3) RACM30-12SER 90-264 12 2.5 88 (3) RACM30-24SER 90-264 24 1.25 89.5 Notes: 2+ RoHS REACH Note1: Other output voltages on request ERP compliant compliant compliant 10 from 10 Note2: Efficiency is tested at nominal input (115/230VAC) and full load at +25C ambient CERTIFIED EN-60335-1 TEBA XEBA 2MOPP 250VAC Model Numbering (3) RACM30- SER PREFERRED ALTERNATIVES Please consider this alternatives: Output Power Single Output Voltage RACM40-K Series Notes: Note3: Other connection types on reqeuest IEC/EN60950-1 certified UL60950-1 certified ANSI/AAMI ES60601-1 certified IEC/EN60601-1 certified IEC/EN60335-1 certified IEC/EN61558-1 certified IEC/EN61558-2-16 certified IEC/EN60601-1-2 certified EN55024/32 certified EN55014-1 (-2) certified www.recom-power.com REV.: 4/2020 MED-1 O O a P P C C S R R n O O M M U D D c O O U U O C C C C e E E T T R R H s S SRACM30-ER AC/DC Converter Series Specifications (measured Ta= 25C, nom. Vin (115/230VAC), full load after warm-up unless otherwise stated) BASIC CHARACTERISTICS Parameter Condition Min. Typ. Max. Internal Input Filter Pi type Input Voltage Range 90VAC 230VAC 264VAC 115VAC 1000mA Input Current 230VAC 290mA 115VAC 60A Inrush Current 230VAC 95A No load Power Consumption 75mW Input Frequency Range 47Hz 63Hz Minimum Load 0% Power Factor 0.55 115VAC 75ms Start-up Time 230VAC 150ms Rise Time 115VAC / 230VAC 10ms 115VAC 15ms Hold-up Time 230VAC 55ms Internal Operating Frequency 100% load at nominal Vin 100kHz Output Ripple and Noise 75mVp-p Efficiency vs. Load RACM30-12SER RACM30-24SER 100 100 90 90 80 80 70 70 60 60 50 50 40 40 30 30 85VAC 85VAC 20 20 115VAC 115VAC 10 10 230VAC 230VAC 264VAC 264VAC 0 0 0 25 50 75 100 0 25 50 75 100 Output Load % Output Load % REGULATIONS Parameter Condition Value Output Accuracy 3.0% max. Line Regulation low line to high line 1.0% max. Load Regulation 0% to 100% load 1.0% max. Transient Response 100% load step change 3.0% max. continued on next page www.recom-power.com REV.: 4/2020 MED-2 Efficiency % Efficiency %