R w a A r E r Reinforced Insulation for 250VAC Working Voltage Features Clearance and Creepage Distance: 8mm DC/DC Converter 5kVAC I/P to O/P 2MOPP Isolation 2A Patient Leakage Current Regulated Industry Standard Pinout Converters 2:1 and 4:1 Wide Input Range REM6 Description 6 Watt 5 The REM6 series of medical grade regulated DC/DC converters features reinforced 5kVAC/1 minute 2:1 & 4:1 isolation with low 2A leakage and are 60601-1 3rd Ed. certified for 250VAC continuous working. The compact DIP24 package offers tightly regulated single and dual outputs, even under no-load conditions. The outputs are short circuit and overload protected. The converters are available in two different pin- DIP24 ning options and optionally with an external control pin for standby consumption as low as 12.5mW. The converters are fully certified to CB, IEC/EN and ANSI/AAMI standards and carry the CE and UL marks. Single and Dual Selection Guide Output Part Input Output Output Efficiency Max. Capacitive Number Voltage Range Voltage Current typ. Load (VDC) (VDC) (mA) (%) (F) (3,4) REM6-xx3.3S/ 4.5-9 / 9-18 / 18-36 / 36-75 3.3 1800 81.5 / 83.5 / 83 / 82.5 2100 (3,4) REM6-xx05S/ 4.5-9 / 9-18 / 18-36 / 36-75 5 1200 86 / 86 / 86 / 86.5 1500 (3,4) REM6-xx12S/ 4.5-9 / 9-18 / 18-36 / 36-75 12 500 86 / 89 / 89 / 88 260 (3,4) REM6-xx15S/ 4.5-9 / 9-18 / 18-36 / 36-75 15 400 87.5 / 88.5 / 88.5 / 88.5 210 (3,4) REM6-xx24S/ 4.5-9 / 9-18 / 18-36 / 36-75 24 250 87 / 88.5 / 88.5 / 88 75 (3,4) REM6-xxx05D/ 4.5-9 / 9-18 / 18-36 / 36-75 5 600 84 / 85 / 85 / 85 860 (3,4) REM6-xx12D/ 4.5-9 / 9-18 / 18-36 / 36-75 12 250 86.5 / 89 / 88.5 / 88 150 (3,4) REM6-xx15D/ 4.5-9 / 9-18 / 18-36 / 36-75 15 200 87.5 / 88 / 88.5 / 87 110 (3,4) REM6-xx3.3SW/ 9-36 / 18-75 3.3 1800 83 / 82.5 2100 (3,4) REM6-xx05SW/ 9-36 / 18-75 5 1200 86 / 86.5 1500 (3,4) REM6-xx12SW/ 9-36 / 18-75 12 500 89 / 88 260 (3,4) REM6-xx15SW/ 9-36 / 18-75 15 400 89 / 88.5 210 (3,4) REM6-xx24SW/ 9-36 / 18-75 24 250 88.5 / 88 75 (3,4) REM6-xx05DW/ 9-36 / 18-75 5 600 85 / 85 860 E314885 (3,4) REM6-xx12DW/ 9-36 / 18-75 12 250 88.5 / 88 150 2MOPP (3,4) REM6-xx15DW/ 9-36 / 18-75 15 200 88.5 / 87 110 250VAC Model Numbering IEC-60601-1 Certified REM 6- / ES-60601-1 Certifed (4) Output Power CTRL Option EN-55011 Certified (1) (3) EN-55022 Certified Input Voltage Pinning (2) Output Voltage Input Type Single or Dual Notes: Note1: for 4:1 Input Voltage Type add W, see Note 2. 2:1 4:1 W xx= 4.5-9 Vin = 05 xx= 9-36Vin = 24 xx= 9-18Vin = 12 xx= 18-75Vin = 48 xx= 18-36Vin = 24 xx= 36-75Vin = 48 Note2: Blank for Standard 2:1 Input Voltage Range W suffix for 4:1 Input Voltage Range Note3: A suffix for A pinning C suffix for C pinning, for more details refer to Package Style and Pinning Note4: CTRL suffix for control pin option, for A pinning only, for C pinning not aviable Examples: REM6-0512D/A = 2:1 Input, 4.5-9Vin, 12Vout, pinout A, without control pin REM6-1215S/C = 2:1 Input, 9-18Vin, 15Vout, pinout C, without control pin REM6-4815SW/A/CTRL = 4:1 Input, 36-75Vin, 15Vout, pinout A with control pin REM6 -243.3SW/C = 4:1 Input, 9-36Vin, 3.3Vout, pinout C, without control pin Refer to Applications Notes www.recom-power.com REV.: 05/2016 MED-1 a Y n t 5 yREM6 DC/DC Converter Series Specifications (measured at T = 25C, nominal input voltage, full load and after warm-up) A BASIC CHARACTERISTICS Parameter Condition Min. Typ. Max. 2:1 5Vin nom. 16VDC 12Vin nom. 25VDC Absolute Maximum Input Voltage 24Vin nom. 50VDC ( 3sec max.) 48Vin nom. 100VDC 4:1 24Vin nom. 50VDC 48Vin nom 100VDC 2:1 5Vin nom. 4VDC 4.5VDC 12Vin nom. 8VDC 9VDC 24Vin nom. 16VDC 18VDC Under Voltage Lockout 48Vin nom. 33VDC 36VDC 8VDC 9VDC 4:1 24Vin nom. 16VDC 18VDC 48Vin nom Start-up Time constant resistive load, Power up or Remote ON/OFF 30ms Remote ON/OFF DC-DC ON Open or 0-1.2VDC (referenced to -Vin Pin) DC-DC OFF 2.2-12VDC Current of CTRL Pin -0.5mA 1mA Remote OFF Input Current 2.5mA Operating Frequency Range 225kHz 250kHz 275kHz Output Ripple and Noise 10F/25V X7R MLCC for 3.3, 5Vout 30mVp-p (20MHz BW limited) 10F/25V X7R MLCC for 12, 15Vout 40mVp-p 4.7F/50V X7R MLCC for 24Vout 50mVp-p Efficiency Table1: Efficiency Crosstable Graph1: Efficiency Factor vs. Load 1.2 Efficiency Crosstable (%) full load Input Voltage 1.0 5 12 24 48 24W 48W 0.8 3.3S 81.5 83.5 83 82.5 83 82.5 05S 86 86 86 86.5 86 86.5 0.6 12S 86 89 89 88 89 88 0.4 15S 87.5 88.5 89 88.5 89 88.5 24S 87 88.5 88.5 88 88.5 88 0.2 05D 84 85 85 85 85 85 12D 86.5 89 88.5 88 88.5 88 0.0 0 10 20 30 40 50 60 70 80 90 100 15D 87.5 88 88.5 87 88.5 87 Load (%) continued on next page MED-2 REV.: 05/2016 www.recom-power.com Output Voltage Efficiency Factor