Technical Data Sheet TECHNICAL DATA SHEET RRC-SMB-MBC-M RRC-SMB-MBC RRC-SMB-MBC-M Medical standard battery charger for smart batteries P/N: 110104 Features: 50W single bay charger For the standard battery form factors RRC203x, RRC204x and RRC205x or smart batteries with a similar footprint Optimized charging process for RRC batteries: Longer cycle life Faster charging Simple operation Plug and Play External medical power supply for worldwide use Country specific AC input cables available Applications: Standard charging station for mobile devices used in medical market DS RRC-SMB-MBC-M E Subject to technical changes without notice * Released document Copy is not subject to revision service 1 / 3 Technical Data Sheet TECHNICAL DATA SHEET RRC-SMB-MBC-M RRC-SMB-MBC Specification Input Voltage range 19.00 26.00VDC Current 2.80A max. Power 50.00W Output Voltage range 0 - 17.40VDC Current range 0 - 4.80A Voltage tolerance 1% max. Current tolerance 10% max. 1.00A, 3% max. 4.00A Charge power 50.00W max. Protection Short circuit, over temperature shutdown, input-/output over current Environmental Operation: Storage: Ambient Temperature 0C to 40C -20C to 80C Pressure and Altitude 1060hPa to 540hPa 1060hPa to 540hPa -382m to 5000m -382m to 5000m Humidity (non-condensing) 20 to 80% r.H. 10 to 90% r.H. General Efficiency ~95% at 100% load Indicator Multi-color LED (green, red, orange) Battery types Standard battery form factors RRC203x, RRC204x and RRC205x or smart batteries with a similar footprint Green procurement RoHS 2011/65/EU + 2015/863/EU REACH 1907/2006 WEEE 2012/19/EU Chinese RoHS Cooling convection cooled LED Indications One time Red/Orange/Green Self-test: Charger is ready for use. Red/Green blinking Battery recognition and initialization. Orange light The inserted battery is of the correct type and is currently being charged. Green light The battery is charged and can be removed for use. Red blinking The battery is too hot or too cold to be charged without damage. If the battery is too cold it will be charged as soon as it has warmed up sufficiently. If the battery is too hot it should be removed to cool down. Red light The battery is damaged or it is a conventional battery which cannot be recharged. Charger Mechanical Details Housing dimensions (LxWxH) 120 x 64 x 43mm Weight 110g (excluding power supply) Safety & EMC In combination with included external AC/DC power supply Regulatory approvals Europe CE International CB Electromagnetic Emissions Europe EN55011, EN55032, level B USA FCC15 class B Electromagnetic Immunity ESD immunity EN/IEC61000-4-2 Electromagnetic field immunity EN/IEC61000-4-3 EFT / Burst EN/IEC61000-4-4 Surge EN/IEC61000-4-5 Conducted Immunity EN/IEC61000-4-6 Magnetic Fields EN/IEC61000-4-8 Voltage dips, short instrumentations & EN/IEC61000-4-11 voltage variations Immunity characteristics EN55024 DS RRC-SMB-MBC-M E Subject to technical changes without notice * Released document Copy is not subject to revision service 2 / 3