30W Single Output GB30 Family Medical/Industrial Grade FEATURES AND BENEFITS Medical rd 30W Open Frame and PCB-mount Power Supply Approved to CSA/EN/IEC/UL60601-1 3 Edition 1.9 x 4.0 x 1.0 Package E-cap Life of >8 Years Industrial Universal Input 90VAC264VAC >1,000,000 Hours MTBF <0.1W No Load Input Power 3 Years Warranty Meets Class B Radiated & Conducted EMI, with Approved to CSA/EN/IEC/UL62368-1 Mar gin Meets Heavy Industrial and IEC60601-1-2 th 4 Edition Levels of EMC Notes: 1. *Consult factory for compliance information. MODEL SELECTION Model Rated Output Ripple & Line Load Input Output Volts 2 1 Number Current Power Noise Regulation Regulation Class/Termination Termination 75mV GB30S05K01 5.0V 4.0A 20.0W 1% 2% Class I (Grounded) pk-pk input, 3-pin AMP/ 4-pin AMP/Molex type 75mV Molex type connector connector for K and GB30S07K01 7.5V 3.0A 22.5W 1% 2% pk-pk Change K to C for C versions Class II input 90mV GB30S09K01 9.0V 3.0A 27.0W 1% 2% pk-pk 120mV GB30S12K01 12.0V 2.5A 30.0W 1% 2% pk-pk Change K to P for 120mV PCB mount pins, Class GB30S15K01 15.0V 2.0A 30.0W 1% 2% pk-pk I input PCB mount pins for P Change K to PCB and V versions 240mV GB30S24K01 24.0V 1.33A 30.0W 1% 2% mount pins, Class II pk-pk input 480mV GB30S48K01 48.0V 0.63A 30.0W 1% 2% pk-pk Notes: 1. Ripple & noise are measured at 20MHz of bandwidth by using a 12 twisted pair-wire terminated with a 0.1F & 47F parallel capacitor. 2. Other output voltages available, consult factory. 3. All specifications are typical at 230VAC, full load, at 25C ambient unless noted. GB30 Datasheet v0719 Copyright 2019 SL Power Electronics Corp. All rights reserved. Page 130W Single Output GB30 Family Medical/Industrial Grade INPUT OUTPUT Input Voltage and Output Voltage See models chart 100VAC240VAC, 10%, 47Hz63Hz, 1 Frequency 20W30W continuous - See models chart for Output Power Input Current 115VAC: 1.2A, 230VAC: 0.6A specific voltage model ratings Turn On Time <700mS Inrush Current 264VAC, cold start: will not exceed 40A peak Hold-up Time 20mS/100VAC at full load Input Fuses 3.15A, 250VAC fuse in both line and neutral 500s resp. time for return to w/in 0.5% of final Earth Leakage Current <500A 264VAC, 60Hz, NC value for any 50% load step from 5% to 100% of (Input to Earth) <1mA 264VAC, 60Hz, SFC Transient Response rated load, i/t<0.2A/s Earth Leakage Current <100A 264VAC, 60Hz, NC Max voltage deviation is 3.5% (Output to Earth) <500A 264VAC, 60Hz, SFC Total Load Regulation 2% Efficiency >88%, typical 0.9min., 230VAC, 80%100% load vector, 25C Power Factor Notes: ambient 1. All specifications are typical at 230VAC input, full load, at 25C ambient unless noted. Notes: 1. All specifications are typical at 230VAC input, full load, at 25C ambient unless noted. ISOLATION RELIABILITY Input-Output: 4000VAC (2 MOPP) Isolation Input-Ground: 1500VAC (1 MOPP) >1,000,000 hours, full load, 110VAC & 220VAC input, Output-Ground: 1500VAC (1 MOPP) MTBF 25C amb., per telcordia 332 issue 6, stress method Isolation Resistance I/P-O/P, I/P-FG, O/P-FG: TBD >8 year life based on calculations at 115VAC/60Hz E-cap Life & 230VAC/50Hz, ambient 25C at 24 hrs/day, 365 days/year, 6 power up cycles/day PROTECTION Will shutdown upon an overtemperature Overtemperature Protection SAFETY condition, Auto-recovery 130% to 160% of rated output current value, ITE/Industrial Safety EN/IEC/UL62368-1 Overload Protection Hiccup mode rd Medical Safety EN/IEC/UL60601-1 3 Edition Short Circuit Protection Hiccup mode 120% to 150% of nominal output voltage, Overvoltage Protection Hiccup Mode GB30 Datasheet v0719 Copyright 2019 SL Power Electronics Corp. All rights reserved. Page 2