MINT1022
22 Watt Single Output Series
Medical - Universal Input Power Supply
90-264 VAC Input
Overvoltage and Short Circuit Protection
Compact Size, Low Profile 2.0 x 3.5 x 0.67
Approved to UL, CSA and EN Standards
CE Compliant (LVD, EMC, RoHS)
Low Standby Power Loss (<0.5 W)
International Safety Standard Approvals
Specifications All specifications are typical at nominal input, full load at 25C unless otherwise stated
Output Specifications
Environmental Specifications
Output Power Natural Convection 2 2 Watts continuous,
Thermal Performance Operatingambient 0-70C
24 W peak for 60 sec-
(seeratings chart)
onds, 10% duty cycle
Non-operating -40 to +85 C
0-50 C Fullload. See table.
Minimum Load Not required
50-70 Cambient derate to 50%
Total Regulation (Line and Load) 1.5%, see table
Relative Humidity Non-condensing 0-95% RH
Rise Time At turn-on 100 ms max.
Maximum Altitude Operating 10,000 ft./40,000 ft.max.
Transient Response 50% step load, 0.2A/s 4 % max. dev., 1.5 msec
/non-operating
max. recovery to 0.5%
Vibration 5Hz-500Hz 2.5 grms
Temperature Coefcfi ient +/-0.03%/C typical
Shock PerMIL-STD-810E 516.4part IV
Overvoltage Protection Standard - See chart
Voltage Adjustability Factory set with xfi ed
resistors to maximize
General Specifications
reliability
Hold-upT ime 120VAC,60Hz 15ms min. at 22Woutput
Efficiency 120VAC, 22W output 70-80% typical depending
on outputvoltage
Input Specifications
SwitchingFrequency Fixed 15 to 100kHz, varies with line
Voltage UniversalInput 90-240VAC(Nominal
andload
100-240VAC)
Safety Approvals Medical UL/EN/IEC 60601-1 CSA22.2
Frequency 47-63Hz
No.601
InputSurge Current Thermistorlimited 32Amax.
MaximumW eight 100Grams (0.20 lbs)
Earth leakagecurrent 264V, 60Hz 125A max. (200 A
MTBF 25 C,ground benign, 300,000hours
singlefault)
Telcordia
InputCurrent 120VAC 0.45A
Overloadand short Fullyprotected Cyclingtype powerlimit
230VAC 0.25A
circuit protection
InputFuse F1 2.0 A
SL Power Electronics Corp. 6050 King Drive Ventura, CA 93003 Phone:805.486.4565 Fax:858.712.2040 Email:info@slpower.com www.slpower.comMINT1022
22 Watt Single Output Series
Medical - Universal Input Power Supply
Medical Voltage Output Total Initial Setpoint Ripple/ OVP
Model Output Current Regulation Tolerance Noise Setpoint
MINT1022A0505I01 5V 4.40A +/-1.5% +/-2.5% 100mV 6.2+/-0.6V
MINT1022A1205I01 12V 1.83A +/-1.5% +/-2.5% 120mV 14.1+/-1.1V
MINT1022A2405I01 24V 0.92A +/-1.5% +/-2.5% 240mV 28+/-2.5V
Notes
1. Other output voltages available by special order
2. Regulation from initial setpoint for 50% loadchange from 50% to0% or 100%
3. Peak output ratingis 24 W atts for1 minute
4. 20MHz Bandwidth scope, differential mode. Measured with scope probe directly acrossoutput terminals of the power supply with load terminated with 0.1 F
5. Heatsink temperatures shouldnot be allowed to exceed 90 C
6.Installation data is online at www.slpower.com
Condor Part Number Key
M I N T 1 022 A 05 05 I 01
Medical Internal Standard Factory No.of Output Initial Design Output Output Input Standard
Product Designation Outputs Wattage of the Product Voltage Connector Connector Configuration
General Specifications (continued from p. 1) Mechanical Notes
INPUT J1 MOLEX 22-43-8040
EMC Information
INPUT J1 MOLEX 22-43-8040
Conducted emissions EN55011 Level B
PIN1 ACL INE
Radiated emissions EN55011 LevelB
PIN4 ACN EUTRAL
Linefreq. harmonics EN61000-3-2 Complies
GROUND SOLDER HOLE ONB OARD
Voltage fluctuations EN61000-3-3 Complies
OUTPUT J2M OLEX 22-43-8040
ESD Contact EN61000-4-2 6 kVcontact
PIN 1 OUTPUT1
ESDAir EN61000-4-2 8kV air
PIN2 O UTPUT 1
Radiated immunity EN61000-4-3 3 V/m
PIN 3C OMMON
FastT ransients(EFT) EN61000-4-4 2 kV/5kHz
PIN4 C OMMON
Linesurge immunity EN61000-4-5 1 kV diff/2 kV
common mode MATINGC ONNECTOR (MOLEX)
Conducted immunity EN61000-4-6 3 Vrms INPUT HOUSING5 0-37-5043 CONTACTS 08-70-1040
Powerfreq. mag.field EN61000-4-8 3A/m OUTPUT HOUSING5 0-37-5043 CONTACTS0 8-70-1040
Voltagedip immunity EN61000-4-11 Consult Factory
SL Power Electronics Corp 6050 King Drive Ventura, CA 93003 Phone:805.486.4565 Fax:805.487.8911 Email:info@slpower.com www.slpower.com
Data Sheet 2008SL Power Electronics Corp. The information and specificationscontained in this data sheet are believedto be correct at time of publication.
However, SL Power accepts no responsibility for consequences arising from reproduction errors or inaccuracies. Specificationsare subject to change without notice.
Rev.9/08/08