MINT1065D
65W Single Output Series
rd
Medical, 3 Edition Certified Universal Input Power Supply 3 Year Warranty
65W Convection Cooled AC-DC Power Supply
Universal Input 90-264 VAC
Class I or Class II input versions
Less 0.5W No Load Power Dissipation
Rugged EMC Compliant Design
2 x 4 x 1.2 Package
ANSI AAMI ES/CSA/EN/IEC 60601-1, 3rd Edition
Certified
Meets Efficiency Level IV Efficiency
Compliant (LVD, RoHS)
Optional Chassis/Cover
RoHS
Specifications All Specifications are typical at nominal input, full load at 25C unless otherwise stated.
AC Input Turn On Time
100-240VAC +/- 10%, 47-63 Hz single phase. Less than 1 second under all rated load conditions.
Input Current Hold-up Time
115VAC: 0.7A, 230VAC: 1.5A. <17 ms at 65W, nominal input (120VAC).
Inrush Current Overvoltage Protection
The inrush at 264VAC, at a cold start, will not exceed 40A. Built-in
Input Fuses Overload Protection
F1, F2: 2.5A, 250VAC fuses provided on all models. Hiccup Mode
Efficiency Short Circuit Protection
Up to 90% typical. Hiccup Mode
Earth Leakage Current Isolation
<300A @ 264Vac, 60Hz, normal; <500A SFC. Touch Current is Input-Output: 4000Vac/2 MOOP
<100A NC, <500A SFC @ 264Vac, 60Hz, measured from output to Input-Ground: 1500Vac/1 MOOP
ground
Output Power Medical Safety Standards
65W continuous. ANSI AAMI ES/CSA/EN/IEC 60601-1, 3rd Edition
Ripple and Noise Switching Frequency
1% pk-pk max., 20MHz BW, differential mode. Measured with noise Variable 50-75kHz typical.
probe directly across output terminals.
Output Voltage Operating Temperature
See chart. 0 to 70C (derate from full rated power above 40 C).
Voltage Adjustability Cooling
Voltage pre-set by fixed resistance. Convection
Minimum Load Output Power Derating
Not required. See derating chart
Total Regulation Storage Temperature
+/- 2%. See chart.
-40 to +85 C.
Transient Response Relative Humidity
500s max., 50% load step, typical. 5% to 95%, non-condensing
______________________________________________________________________________________________________________________________ 1 of 3
SL Power Electronics Corp 6050 King Drive Ventura, CA 93003 Phone 805.486.4565 Fax 858.712.2040 Email: info@slpower.com www.slpower.comMINT1065D
65W Single Output Series
rd
Medical 3 Edition Certified Universal Input Power Supply 3 Year Warranty
Output Parameters
Model Output Current Ripple Total OVP
Number Volts (V) Open Frame w/cover & Noise Regulation Threshold
MINT1065D1275K01 12V 5.25A 4.00A 120mV 2% 15.0 1.5V
MINT1065D1375K01 13.2V 4.80A 3.63A 140mV 2% 16.0 1.5V
MINT1065D1575K01 15V 4.33A 3.20A 150mV 2% 18.0 1.5V
MINT1065D1875K01 18V 3.50A 2.66A 180mV 2% 21.0 2.0V
MINT1055D2075K01 20V 3.25A 2.40A 200mV 2% 23.0 2.0V
MINT1065D2475K01 24V 2.70A 2.00A 240mV 2% 27.0 2.0V
MINT1065D4875K01 48V 1.35A 1.00A 480mV 2% 55.0 4.0V
Specifications (continued)
Vibration Shock
2
Operating: 0.003 g /Hz, 1.5 g overall, 3 axes, 10 min. / axis Operating: Half-sine, 20 g , 10 ms, 3 axes, 6 shocks total
rms pk
2
Non-Operating: 0.026 g /Hz, 5.0 g overall, 3 axes, 1 hr. / axis Non-Operating: Half-sine, 40 g , 10 ms, 3 axes, 6 shocks total
rms pk
Pollution Degree Overvoltage Category
2 OVCII (2500V for Mains), OVCI (1500V for Secondary)
Dimensions Operating Altitude
W 2.0 x L 4.0 x H 1.17. See outline drawing. Up to 4000m.
Weight Non-operating Altitude
150 grams -500 to 40,000 ft.
EMI/EMC Compliance
Conducted Emissions EN55011 Class B, FCC Part 15, Class B.
Radiated Emissions EN55011 Class A, FCC Part 15, Class A.
Static Discharge Immunity EN61000-4-2, 6kV Contact Discharge, 8kV air discharge
Radiated RF Immunity EN61000-4-3, 3V/m.
EFT/Burst Immunity EN61000-4-4, 2kV/5kHz..
Line Surge Immunity EN61000-4-5, 1kV differential, 2kV common-mode
Conducted RF Immunity EN61000-4-6, 3Vrms
Power Frequency Magnetic Field Immunity EN61000-4-8, 3A/m
Voltage Dip Immunity EN61000-4-11 240VAC, 0%/0.5 cycle, 40%/5cycles, 70%/25
cycles
Model Number Key
MINT 1 065 X 12 75 K 01
Model:01" = Standard Model, 02 and higher indicates a modified model.
Input Connector: = 3 pin Header (class I input); = 2 pin Header (class II input).
Output Connector:75" = Output Connector, 4 pin header. Other options available, consult factory.
Output Voltage:12" = 12V,24" = 24V, etc.
Configuration: = EN60601-1 3rd Edition Certified, = Optional chassis & cover (class I only).
Output Pow er:065" = 65W
# of Outputs:1" = Single output
Product Family: = Medical, = Internal,N = New Technology
______________________________________________________________________________________________________________________________ 2 of 3
SL Power Electronics Corp 6050 King Drive Ventura, CA 93003 Phone 805.486.4565 Fax 858.712.2040 Email: info@slpower.com www.slpower.com