CUS200M Series Specifications Model CUS200M Input Input Voltage range V 85 - 264VAC (1) Input Frequency Hz 47 - 63Hz Input Current (115/230VAC) A 200W output: 2.2 / 1.1A 250W output: 3.0 / 1.5A Inrush Current (typ) (Cold Start) A 35A at 115VAC, 70A at 230VAC Leakage Current uA <300uA at 265VAC, 60Hz Power Factor (115/230VAC) - 0.99 / 0.95 Harmonic Compliance - Meets EN61000-3-2 Class A No Load Power Consumption W <0.5W at 230VAC (Remote on/off activated and no load on 5Vsb) Hold Up Time (typ) at 115 / 230VAC Input ms 16ms at 200W output, 12ms at 250W Efficiency at 115 / 230VAC input 200W % 12V: 92 / 93, 18 to 48V: 92 / 94 Conducted & Radiated EMI - EN55011-B, EN55022-B, FCC Class B Immunity - IEC61000-4-2, IEC60601-1-2 Ed4. See immunity table Insulation Class - Construction suitable for Class I installation Safety Agency Certifications - IEC/UL/ES/CSA/EN60601-1, IEC/UL/CSA/EN62368-1, 60950-1, CE Mark Immunity Test Standard Test Level Criteria Notes (the power stated below is total power (main power + Vsb output)) ESD EN61000-4-2 2 & 3 A Level 3 air discharge only Radiated Susceptibility EN61000-4-3 3 A Electrical Fast Transient Burst EN61000-4-4 3 A Surge EN61000-4-5 3 & 4 A Level 4 common mode only Conducted Susceptibility EN61000-4-6 3 A - Magnetic fields EN61000-4-8 4 A - Voltage Dips EN61000-4-11 Class 3 B 100 - 120VAC B 200 - 240VAC Short Interruptions EN61000-4-11 Class 3 B - Voltage Dips EN60601-1-2:2015 Class 2 A 100 - 120VAC or >160W B 200 - 240VAC and <160W Short Interruptions EN60601-1-2:2015 Class 2 B - SEMI F47 Line Dip SEMI F47 - - At input voltages > 200VAC 2 CUS200M Series