Catalog: 1654001
Corcom Product Guide Issue Date: 06.2011
High Performance RFI Power Line Filters for Medical Equipment
HT Series
6EHT3
UL Recognized
CSA Certied
VDE Approved
6EHT7
6EHT1
HT Series Specications
Designed to provide signicant attenuation of RFI Maximum leakage current each Line to Ground:
@ 120 VAC 60 Hz: 2 A
noise in the frequency range from 10kHz to 30MHz
@250 VAC 50 Hz: 5 A
Size and cost-effective
Hipot rating (one minute):
Line to Ground: 2250 VDC
Line to Line: 1450 VDC
Rated Voltage (max): 250 VAC
Operating Frequency: 50/60 Hz
Rated Current: 3 to 15A
Operating Ambient Temperature Range
(at rated current I ): -10C to +40C
r
In an ambient temperature (T ) higher than +40C
a
the maximum operating current (I ) is calculated as
o
follows: I = I (85-Ta)/45
o r
Ordering Information
6 E HT 7 M
Available Part Numbers
Sux
3EHT1 6EHT7
M Metric inserts (style 7 only)
3EHT3 6EHT7M
Omit for standard or none
3EHT7 10EHT1
3EHT7M 10EHT3
Input / Output Style
6EHT1 15EHT1
1 .250 [6.3] spade terminals
6EHT3 15EHT6
3 Wire leads
6 Threaded bolt
7 IEC inlet* (line side)
.250 [6.3] spade terminals
Accessories
(load side)
GA400: NEMA 5-15P to IEC 60320-1 C-13 line cord
HT Series
Leakage current designation
E Low leakage (<0.5 mA)
Current Rating
3, 6, 10 or 15A
*IEC 60320-1 C14 inlet mates with C13 connector
40
Dimensions are in inches and millimeters unless otherwise specied. Values in italics For email, phone or live chat, please go to
te.com/help
are metric equivalents. Dimensions are shown for reference purposes only.
corcom.com
Specications subject to change.RFI Power Line Filters
Catalog: 1654001
Corcom Product Guide Issue Date: 06.2011
High Performance Power Line Filters for Medical Equipment (continued)
HT Series
Electrical Schematics
HT3
3 & 6EHT
LINE
LOAD
Blue
Blue
Green/Yellow
1
B
Brown
Brown
10EHT
LINE LOAD A
C
D
E
15EHT
LINE LOAD
HT6
Case Styles
HT1 (3, 6, 10A)
E DB
E
2.00 0.01
B
50.80 0.25
A
C
C
D
A
HT7 & HT7M
HT1 (15A)
D
B
E DB
E
C
A
2.00 0.01
C
50.80 0.25
A
41
Dimensions are in inches and millimeters unless otherwise specied. Values in italics For email, phone or live chat, please go to
te.com/help
are metric equivalents. Dimensions are shown for reference purposes only.
corcom.com
Specications subject to change.