DC/DC Converter THM 10 Series, 10 Watt Wide 2:1 input voltage 10 W DC/DC converter in a compact DIP-24 plastic case I/O isolation 5000 VACrms rated for 250 VACrms working voltage Certification according to IEC/EN/ES 60601-1 3rd edition for 2xMOPP CB Scheme Risk management process according to IEC 606010-1 ES 60601-1 ISO 14971 including risk management file The THM-10 series is a range of medical 10 Watt DC/DC converters in DIP-24 plastic package and with wide 2:1 input voltage range. They provide a reinforced Acceptance criteria for electronic isolation system for 5000 VACrms isolation and a very low leakage current of assemblies according to IPC-A-610 Level 3 less than 2 A. The units are approved to IEC/EN/ES 60601-1 3rd edition for 2 Low leakage current < 2A x MOPP (Means Of Patient Protection) and come along with an ISO 14971 risk Extended operating temperature range management file. Design and production conform to the quality management 40C to 90C. system ISO 13485. With a high efficiency of up to 89% and highest grade com - EMC compliance to IEC 60601-1-2 4th ponents the converters can reliably operate in an ambient temperature range edition and EN55032 class A of 40C up to +90C. They constitute a reliable solution not only for medical Operating up to 5000m altitude equipment but also for demanding ranges of application such as transportation, 5 year product warranty control & measurement or IGBT drivers. Models Order code* Input voltage range Output voltage Output current max. Efciencfi y typ. THM 10-0510 3.3 VDC 2500 mA 80.0 % THM 10-0511 5.0 VDC 2000 mA 84.0 % THM 10-0512 12 VDC 830 mA 86.5 % THM 10-0513 15 VDC 670 mA 87.0 % 4.5 9 VDC THM 10-0515 (5 VDC nominal) 24 VDC 416 mA 85.5 % THM 10-0521 5.0 VDC 1000 mA 83.0 % THM 10-0522 12 VDC 416 mA 85.5 % THM 10-0523 15 VDC 333 mA 86.5 % THM 10-1210 3.3 VDC 2500 mA 83.0 % THM 10-1211 5.0 VDC 2000 mA 85.5 % THM 10-1212 12 VDC 830 mA 88.0 % THM 10-1213 15 VDC 670 mA 89.0 % 9. 0 18 VDC THM 10-1215 (12 VDC nominal) 24 VDC 416 mA 89.0 % THM 10-1221 5.0 VDC 1000 mA 84.0 % THM 10-1222 12 VDC 416 mA 89.0 % THM 10-1223 15 VDC 333 mA 88.0 % THM 10-2410 3.3 VDC 2500 mA 83.0 % THM 10-2411 5.0 VDC 2000 mA 86.5 % THM 10-2412 12 VDC 830 mA 89.0 % THM 10-2413 15 VDC 670 mA 89.0 % 18 36 VDC THM 10-2415 (24 VDC nominal) 24 VDC 416 mA 89.0 % THM 10-2421 5.0 VDC 1000 mA 85.0 % THM 10-2422 12 VDC 416 mA 89.0 % THM 10-2423 15 VDC 333 mA 88.0 % THM 10-4810 3.3 VDC 2500 mA 82.5 % THM 10-4811 5.0 VDC 2000 mA 86.5 % THM 10-4812 12 VDC 830 mA 89.0 % THM 10-4813 15 VDC 670 mA 89.0 % 36 75 VDC THM 10-4815 (48 VDC nominal) 24 VDC 416 mA 88.5 % THM 10-4821 5.0 VDC 1000 mA 85.0 % THM 10-4822 12 VDC 416 mA 88.0 % THM 10-4823 15 VDC 333 mA 88.0 % * suffix -B2 for trim option with adjustable output suffix -B3 for remote control option suffix -B4 for trim + remote-control option www.tracopower.com Page 1 of 5 THM 10 Series, 10 Watt Input Specifications Input current no load 5 Vin models: 20 mA typ. 12 Vin models: 10 mA typ. 24 Vin models: 6 mA typ. 48 Vin models: 4 mA typ. Surge voltage (3 s max.) 5 Vin models: 16 V max. 12 Vin models: 25 V max. 24 Vin models: 50 V max. 48 Vin models: 100 V max. Start-up voltage 5 Vin models: 4.5 VDC (or lower) 12 Vin models: 9 VDC (or lower) 24 Vin models: 18 VDC (or lower) 48 Vin models: 36 VDC (or lower) Startup time 30 ms typ. Under voltage shut down 5 Vin models: 3 4.4 VDC 12 Vin models: 7 8.8 VDC 24 Vin models: 15 17.5 VDC 48 Vin models: 31 34.5 VDC Conducted noise Conducted & Radiated input suppression EN 55011 limits to IEC 60601-1-2 4th editon EN 55032 class A (internal filter) EN 55032 class B (with external components) Filter proposal for compling to EN 55032 class B www.tracopower.com/overview/thm10 EMC immunity Generic for Medical equipment IEC/EN 60601-1-2 4th edition ESD (electrostatic discharge) EN 61000-4-2, air 8 kV, contact 6 kV, perf. criteria A Radiated immunity EN 61000-4-3, 10 V/m, perf. criteria A Fast transient / surge EN 61000-4-4, 2 kV, perf. criteria A (with external input capacitor / diode) EN 61000-4-5, 2 kV perf. criteria A 5 Vin models: Nippon chemi-con KY 1000 F/ 25 V and reverse diode (Vishay V10P45) in parallel 12 & 24 Vin models: Nippon chemi-con KY 470 F/ 50 V 48 Vin models: Nippon chemi-con KY 330 F/ 100 V Conducted immunity EN 61000-4-6, 10 Vrms, perf. criteria A Magnetic field immunity EN 61000-4-8 100 A/m, continuous, perf. criteria A 1000 A/m, 1 sec., perf. criteria A External input fuse required 5 Vin models: 5 A (recommended values, slow blow type) 12 Vin models: 2 A 24 Vin models: 1 A 48 Vin models: 0.5 A www.tracopower.com Page 2 of 5