Open Frame or Enclosed Versions Available
UL / IEC / EN 60601 3.1 Edition & UL / IEC / EN
60950 AM2 Safety Approvals
th
4 Edition UL / IEC / EN 60601 EMC Compliant
4000VAC Input to Output Isolation (2x MOPP)
Class I and Class II Input Configurations
Suitable for BF Application with appropriate
system consideration
High Efficiency up to 94%
<500mW No Load Input Power
Electrical Specifications
Input Environmental
Operating Temperature -30C to +70C (with derating)
Storage Temperature -30C to +85C
Input Voltage 90-264 VAC
Humidity 20%-90% RH
Input Frequency 47-63 Hz
Operating altitude <5000 meters
Input Current (RMS) <3.0A at 115 VAC; <1.5A at 230 VAC
MTBF: >250k hours per MIL-HDBK-217F at full load
Power Factor >0.9 at Full load (230 VAC)
and 25C ambient temperature
Inrush Current (<2ms) <45A at 115 VAC; <90A at 230 VAC
Leakage Current <100A max. (Input-Output)
Compliance
Safety Approvals
USA/Canada UL60601-1 3.1 Edition , UL/cUL60950-1 AM2
Output
UL62368-1 (Pending)
Output Power 240W max. See table for details
Europe IEC/EN60601-1 3rd edition, TUV EN60950-1
Output Voltage See table
CB Report
Hold Up Time 10ms min (Vout = 90% Vnominal)
EN62368-1 (Pending)
Efficiency Up to 94%. See table for details
Isolation: 4000VAC input to output, 2 x MOPP
Minimum Load No Minimum Load 2000VAC input to ground,1 x MOPP
1500VAC output to ground,1 x MOPP
EMC (IEC60601-1-2:2014) : FCC Class B Radiated & Conducted
EN55011/55022 Class B Radiated &
Conducted (Class A Radiated for Class II
Configuration)
Harmonic Currents IEC 61000-3-2
Voltage Flicker IEC 61000-3-3
Electrostatic Discharge IEC 61000-4-2: 15kV Air, 8kV contact
Radiated Immunity IEC 61000-4-3: 10V/m
Protection
EFT/Burst IEC 61000-4-4: +/-2kV
Surge Immunity IEC 61000-4-5: 2005 1kV diff, 2kV com
Overvoltage Auto recovery
Conducted Immunity IEC 61000-4-6: 10Vrms
Magnetic Field IEC 61000-4-8: 30A/m
Overload Auto recovery
Dips / Interruptions IEC 61000-4-11: 30% reduction for 500ms,
100% reduction for 10ms.
Short Circuit Auto recovery
General
Dimensions 2 .05W x 4 .10L x 1 .0 8 7H
Weight 234g Typical
Your Partners in Power ... Power Partners, Inc. | 43 Broad Street | Hudson, MA 01749
1
Tel: (978)567-9600 | Fax: (978)567-9601
Specifications subject to change.
Website: www.powerpartners-inc.com
PDAM240: November 12, 2018
Medical
AC/DC Power
Models and Ratings
Output Max Load Max Load 10CFM Output Ripple &
Model Efficiency Fan Output
Voltage Convection Forced Air Regulation Noise
PDAM240-12A-H 12V 13.33A 20A +/-2% 120mV 92.5% 12V/0.5A
PDAM240-14A-H 24V 6.67A 10A +/-2% 240mV 93% 12V/0.5A
PDAM240-18A-H 48V 3.33A 5A +/-2% 480mV 94% 12V/0.5A
Notes:
1. All models are available in an enclosed version (e.g. PDAM240-12A would be PDAM240-12C)
2. All models are available with terminal block type output. Remove the H extension (indicating header type output) when ordering.
3. We strongly recommend conducting isolation testing with a DC voltage.
4. Hold-up Time measured at 90% Vout.
5. Main load must be greater than 50% to drive the fan output.
6. Ripple and noise measured at 20MHz bandwidth with a 47uF electrolytic and 0.1uF ceramic capacitor in parallel with the output, at the
DC connector.
7. Please secure the PSU to your assembly using the four mounting holes in the corners for Class I and Class II equipment.
Block Diagram
Your Partners in Power ... Power Partners, Inc. | 43 Broad Street | Hudson, MA 01749
2
Tel: (978)567-9600 | Fax: (978)567-9601
Specifications subject to change.
Website: www.powerpartners-inc.com
PDAM240: November 12, 2018