AC-DC Power Supplies
EML15 Series
15 Watts
Compact Size
Medical Approvals
Single Outputs from 3.3 to 48 V
PCB Mount, Open Frame & Chassis Mount
Encapsulated PCB & Chassis Mount
Class II
Peak Load Capability
3 Year Warranty
Dimensions:
The EML15 is a series of open frame and encapsulated AC-DC single output power supplies
EML15:
designed for medical applications. The series provides a number of flexible mechanical options (-P): 2.44 x 1.21 x 0.95 (62.0 x 30.7 x 24.1 mm)
(-T): 3.10 x 1.25 x 0.91 (78.7 x 31.8 x 23.1 mm)
including PCB mount, open frame, chassis mount with screw terminals and a DIN Rail option. With
(-E): 2.56 x 1.31 x 0.96 (65.0 x 33.3 x 24.4 mm)
approvals to world-wide medical safety standards, compliance with class B for both conducted and
(-S): 3.30 x 1.36 x 1.04 (84.0 x 34.5 x 26.4 mm)
radiated emissions and a 130%, 30s peak load capability, these class II isolation parts benefit
system designers with easy integration into the latest healthcare products and applications.
Models & Ratings
Output Current
(2,3)
Output Power Output Voltage Efficiency Model Number
(1)
Nominal Peak
10 W 3.3 VDC 3.00 A 3.90 A 75% EML15US03
15 W 5.0 VDC 3.00 A 3.90 A 78% EML15US05
15 W 9.0 VDC 1.67 A 2.17 A 80% EML15US09
15 W 12.0 VDC 1.25 A 1.62 A 80% EML15US12
15 W 15.0 VDC 1.00 A 1.30 A 80% EML15US15
15 W 24.0 VDC 0.63 A 0.82 A 82% EML15US24
15 W 36.0 VDC 0.42 A 0.54 A 82% EML15US36
15 W 48.0 VDC 0.32 A 0.41 A 82% EML15US48
Notes
1. Peak load lasting <30 s with a maximum duty cycle of 10%, average output power not to exceed nominal.
2. Add suffix to model number to define type: add -P for PCB mount, add -T for chassis mount, add -E for encapsulated, add -S for screw terminals.
3. A Screw terminal version (-S) is available with DIN clip attached, add suffix D, e.g. EML15US24-SD, DIN rail mounting kit is available as a seperate item,
order code ECL15 DIN CLIP.
Summary
Characteristic Minimum Typical Maximum Units Notes & Conditions
85 264 VAC
Input Voltage Range
120 370 VDC
No Load Input Power 0.3 W
Efficiency 75 80 82 % See Models & Ratings table
Operating Temperature -20 +70 C Derate linearly from 100% at +50 C to 50% at +70 C
EMC EN55032 Level B Conducted & Radiated, EN61000-4, EN61000-3, EN60601-1-2
Safety Approvals EN60601-1, ANSI/AAMI ES60601-1. IEC60601-1
www.xppower.com
1AC-DC Power Supplies
EML15 Series
Input
Characteristic Minimum Typical Maximum Units Notes & Conditions
85 264 VAC
Input Voltage Range
120 370 VDC
Input Frequency 47 63 Hz
Input Current - Full Load 0.32/0.16 A rms At 115/230 VAC
No Load Input Power 0.3 W
Inrush Current 20/40 A At 115/230 VAC
Earth Leakage Current Class II construction no earth
Input Protection F2.0A/250 V internal fuse fitted in line and neutral
Output
Characteristic Minimum Typical Maximum Units Notes & Conditions
Output Voltage 3 48 VDC See Models and Ratings table
Initial Set Accuracy 1 %
Output Voltage Adjustment 5 % Not encapsulated units
Minimum Load 0A No minimum load required
Start Up Delay 2s
Start Up Rise Time 14 ms
Hold Up Time 12 ms at full load and 115 VAC
Line Regulation 0.5 %
Load Regulation 1 %
Deviation, recovery within 1% in less than 500 s
Transient Response 4%
for a 25% load change
50 mV pk-pk 3.3-5 V versions, 20 MHz bandwidth
Ripple & Noise 120 mV pk-pk 12-15 V versions, 20 MHz bandwidth
200 mV pk-pk 24-48 V versions, 20 MHz bandwidth
195 216 % Vnom 3.3 V versions, recycle input to reset
Overvoltage Protection
115 130 % Vnom All other versions, recycle input to reset
Overload Protection 120 150 %
Short Circuit Protection Trip & Restart (hiccup mode)
Temperature Coefficient 0.05 %/C
General
Characteristic Minimum Typical Maximum Units Notes & Conditions
Efficiency 75 80 82 % See Models & Ratings table
Isolation: Input to Output 4000 VAC
Switching Frequency 70 kHz
3
Power Density 4.8 W/in PCB Mount version
Mean Time Between Failure >400 kHrs MIL-HDBK-217F, +25 C GB
0.07 (35) lb (g) Open frame versions
Weight 0.20 (90) lb (g Encapsulated version
0.24 (110) lb (g Screw terminal version
Environmental
Characteristic Minimum Typical Maximum Units Notes & Conditions
Operating Temperature -20 +70 C Derate linearly from 100% at +50 C to 50% at +70 C
Storage Temperature -40 +85 C
Cooling Convection-cooled
Humidity 95 %RH Non-condensing
Operating Altitude 3048 m
Shock IEC68-2-27, 30 g, 11 ms half sine, 3 times in each of 6 axes
Vibration IEC68-2-6, 2 g, 10 Hz to 500 kHz, 10 mins/cycle, 60 mins each cycle
www.xppower.com