1500 Watts xppower.com HPU1K5 Series Medical Safety Approvals (-M Versions) Variable Fan Speed To Reduce Audible Noise -20 C to +70 C Operation AC OK, DC OK, Inhibit & 5 V Standby Supply Fault & Overtemperature Signals SEMI F47 Compliant 3 Year Warranty Specification Input General Input Voltage 85-264 VAC, see derating curve Efficiency 90% typical Isolation 4000 VAC Input to Output 2 x MOPP, Input Frequency 47-63 Hz 1500 VAC Input to Ground 1 x MOPP Input Current 13 A/6.5 A typical at 115/230 VAC 500 VDC Output to Ground Inrush Current 35 A maximum at 264 VAC Switching Frequency 70 kHz (PFC), 130 kHz (main converter) Power Factor >0.9 typical Earth Leakage Current 1.1 mA max at 264 VAC 60Hz, 3 Power Density 18 W/in <300 A max at 264 VAC (-M version) Signals AC OK, DC OK, Inhibit, Fault Input Protection Internal T20 A/250 VAC fuse in line and (see Signals page) neutral MTBF 470 kHrs to Telecordia SR-332 at 25 C, GB Output Environmental Output Voltage See model table Operating Temperature -20 C to +70 C, derate linearly from Output Voltage Trim Via potentiometer or external voltage, +50 C at 2.5 %/C to 50% load at +70C see model tables Cooling Internal load dependant variable speed fans Initial Set Accuracy 1% of nominal with 50% load Operating Humidity 95% RH, non-condensing Minimum Load No minimum load required Storage Temperature -40 C to +85 C Line Regulation 0.5% maximum Operating Altitude 3000 m Load Regulation V1: 0.5%, V2: 5% Shock 3 shocks in each axis (total 18 shocks) Start Up Delay 1 s typical 30 g 11 ms (half sine). Compliant with EN60068-2-27. Over/Undershoot 0.5% typical Vibration 2 g 10-500 Hz 10 sweeps. Compliant with Transient Response 4% deviation, recovery to within 2% in EN60068-2-6. 500 s for 50-75-50% load change EMC & Safety Ripple & Noise 24-48 V models: 1% max pk-pk 12 V models: 2% max pk-pk Emissions EN55011 level A conducted & radiated, V Standby: 3% max pk-pk, EN55032 level A conducted & radiated 20 MHz bandwidth Immunity Compliant with EN61204-3:2000 Overvoltage Protection 115-140% of V1 nominal, recycle input high severity levels AC to reset Harmonic Currents EN61000-3-2 class A, Overtemperature Protects the unit against overtemperature. EN61000-3-2 class C for loads 10% Protection Auto restart Voltage Flicker EN61000-3-3 Overcurrent Protection 110 - 140% V1, V Standby power limited ESD Immunity EN61000-4-2, level 3, Perf Criteria A Short Circuit Protection Continuous, trip and restart (hiccup mode) Radiated Immunity EN61000-4-3, level 3, Perf Criteria A Temperature 0.02%/C (after 20 minute warm up) EFT/Burst EN61000-4-4, level 3, Perf Criteria A Coefficient Surge EN61000-4-5, installation class 3 Remote Sense Compensates for 0.5 V total drop Perf Criteria A, SEMI F47 Current Share Share upto 8 units maximum, units Conducted Immunity EN61000-4-6, level 3, Perf Criteria A share current within 10% of each other Dips & Interruptions EN61000-4-11 , 30% 10 ms, 60% 100 ms, at full load. 100% 5000 ms, Perf Criteria A, B, B , EN60601-1-2, 30% 500 ms, 60% 100 ms, 100% 10 ms, 100% 5000 ms, Perf Criteria A, A, A, B Safety Approvals EN60601-1, ANSI/AAMI ES60601-1, CSA22.2 No.60601-1 per cUL, Including Risk Management M Versions, IEC60950- 1:2005 Ed 2 / IEC62368-1:2014 UL 62368-1 & CAN/CSA C22.2 No. 62368- 1-14, EN62368-1:2014/A11:2017 AC-DCAC-DC HPU1K5 Models and Ratings Output Current V1 Output Voltage Voltage Adj Standby Supply (1) Output Power Model Number V1 V1 V2 <180 VAC >180 VAC 1200 W 12.0 VDC 11-14 V 100 A 100 A 5 V/1 A HPU1K5PS12 1500 W 24.0 VDC 22-28 V 50 A 63 A 5 V/1 A HPU1K5PS24 1500 W 48.0 VDC 45-52 V 25 A 31 A 5 V/1 A HPU1K5PS48 Notes 1. See derating curves. 2. For medical version, add suffix -M to model number. Mechanical Details 1.70 (43.2) 0.10 (2.5) 1.55 Voltage Adj. 12.75 (323.9) (39.4) +V BUSS BAR -V BUSS BAR 4.00 (101.6) E L N Logic J3 Connector 19 20 1.00 (25.4) 10.50 (266.7) 4X M3 THD on both side faces 0.11 (2.8) max screw penetration 0.25 (6.4) 1 2 1.00 (25.4) 0.35 (8.9) 0.25 (6.4) 2.90 8.00 (203.2) (73.7) 0.53 (13.3) AIR FLOW 2.95 (74.9) AIR FLOW 4X M3 THD 0.11 (2.8) Max Screw Penetration Logic Connector: J3, JST, PN S20B-PHDSS (LF) SN) Pin Function Pin Function Pin Function 1 + Sense 8 NC 15 DC OK 2 + Sense 9 Inhibit 16 NC 3 - Sense 10 NC 17 Signal GND 4 - Sense 11 Fault 18 NC 5 Current Share 12 NC 19 5 V Standby Rtn (V2) 6 Current Share 13 AC OK 20 5 V Standby (V2) 7 V Trim 14 NC Mates with JST PN PHDR-20VS, Crimp contacts JST PN SPHD-00IT-P0.5 Notes 1. All dimensions are in inches (mm). 2. Weight 5.2 lb (2.35 kg)