AC-DC Medical 25 Watts THE XPERTS I N POWER PBM25 Series Small Size 2.50 x 4.25 x 0.96 Low Leakage Current Medical Safety Agency Approvals Both Line and Neutral Fused Meets Class B, EN55011 Conducted Fixed Switching Frequency Specification Input General Input Voltage 85-264 VAC Efficiency 68% minimum for 5 V & multi output models, 70% minimum all others Input Frequency 47-63 Hz Isolation 4000 VAC Input to Output Input Current 0.7 A (rms) at 115 VAC, 1500 VAC Input to Ground 0.4 A (rms) at 230 VAC 500 VAC Output to Ground Inrush Current 10 A max at 115 VAC, Switching Frequency 42 KHz 5 KHz 25 A max at 230 VAC MTBF 500,000 hrs per min MIL-HDBK-217F Earth 90 A max at 115 VAC 60 Hz Leakage Current 150 A max at 230 VAC 50 Hz Size 4.25 x 2.50 x 0.96 107.95 x 63.50 x 24.38 mm Output Weight Single output models 163 g, Output Voltage 5.1 to 28 VDC, See Table Multi output models 175 g Output Power 25/28 Watts, See Table Environmental Minimum Load No minimum for single outputs models See Table for multi outputs models Operating 0 C to +70 C, Temperature derate linearly from 100% load at Start Up Delay 3 s max +50C to 50% load at +70 C Start Up Rise TIme 12 ms Storage Temperature -40 C to +85 C Hold up Time 10 ms minimum at 110 VAC Cooling Convection cooled Initial Set Accuracy 1% Relative Humidity 5-95%, non-condensing Line Regulation 0.5% max at full load Operating Altitude 3000m Total Regulation See Table Shock 20 G in 3 directions, 3 times each Transient Response 4% max deviation, 500 s recovery time for a 25% load change Vibration 0.15mm over 10-55Hz and 1 octave/min. 3 axes for 30 min Ripple & Noise 1% max pk-pk (See Note 1) Overvoltage Provided on output 1 only set at EMC & Safety Protection 112-132% of its nominal output EMC EN55011 (CISPR11) Class B & FCC voltage Class B conducted Overcurrent 125-200% with auto recovery ESD EN61000-4-2 Level 2 Perf Criteria A Protection Radiated Immunity EN61000-4-3 3 V/m Perf Criteria A Short Circuit Trip & restart with auto recovery Protection EFT/Burst EN61000-4-4 Level 3 Perf Criteria A Temperature 0.04%/C Surge EN61000-4-5 Level 3 Perf Criteria A Coefficient Safety Approvals UL2601-1, CSA C22.2 No 601.1, EN60601-1, IEC 601-1 www.xp-medical.com sales xp-medical.comOUTPUT VOLTAGE & CURRENT RATINGS PBM25 Output Power Output 1 Output 2 Output 3 Model (2) (2) (2) Maximum Vnom Imin Imax Reg. Vnom Imin Imax Reg. Vnom Imin Imax Reg. Number 28 W 5.1 V 0.0 A 5.5 A 2% PBM25US05 28 W 12.0 V 0.0 A 2.3 A 1% PBM25US12* 28 W 15.0 V 0.0 A 1.9 A 1% PBM25US15* 28 W 24.0 V 0.0 A 1.2 A 1% PBM25US24* 28 W 28.0 V 0.0 A 1.0 A 1% PBM25US28* 25 W 5.1 V 0.4 A 2.5 A 3% +12 V 0.2 A 1.5 A 5% PBM25UD01* 25 W 5.1 V 0.4 A 2.5 A 3% +24 V 0.1 A 0.8 A 5% PBM25UD02* 25 W 5.1 V 0.4 A 2.5 A 3% +12 V 0.2 A 1.5 A 5% -12 V 0.05 A 0.2 A 4% PBM25UT01* 25 W 5.1 V 0.4 A 2.5 A 3% +15 V 0.1 A 1.0 A 5% -15 V 0.05 A 0.2 A 4% PBM25UT02* 25 W 5.1 V 0.4 A 2.5 A 3% +24 V 0.1 A 0.8 A 5% -12 V 0.05 A 0.2 A 4% PBM25UT03* Notes 1. Ripple and noise measured with 20 MHz bandwidth with 10 F electrolytic in parallel with 0.1 F ceramic. 2. Regulation includes line and load regulation and set point accuracy. All multiple output models may be operated at no-load, but output regulation will increase to 10% on outputs V2 & V3. * Also available from Farnell InOne, see pages 350 & 351. Mechanical Details - Single Output Models PIN CONNECTIONS Pin Function 1 Return 2 Return 3 Output 1 4 Output 1 Notes: 1. Dimensions shown in inches (mm). 2. Tolerance 0.03 (0.76) maximum. 3. Input connector P1 mates with Molex housing 09-50-3031 and Molex 2878 series crimp terminal. 4. Output connector P2 mates with Molex housing 09-50-3041 and Molex 2878 series crimp terminal. 5. Weight 163 g. 6. Covers available, order part number PBM25 COVER*, or add suffix -C to have cover fitted (78.8 x 12.0 x 34.8 mm). 7. Mating connector kit available, order part number PBM25 CONKT-S*. Mechanical Details - Multi Output Models PIN CONNECTIONS Pin Dual Output Triple Output 1 Output 2 Output 2 2 Output 1 Output 1 3 Output 1 Output 1 4 Common Return Common Return 5 Common Return Common Return 6 No Connection Output 3 Notes: 1. Dimensions shown in inches (mm). 2. Tolerance 0.03 (0.76) maximum. 3. Input connector P1 mates with Molex housing 09-50-3031 and Molex 2878 series crimp terminal. 4. Output connector P2 mates with Molex housing 09-50-3061 and Molex 2878 series crimp terminal. 5. Weight 175 g. 6. Covers available, order part number PBM25 COVER*, or add suffix -C to have cover fitted (78.8 x 120.0 x 34.8 mm). 7. Mating connector kit available, order part number PBM25 CONKT-D/T*. www.xp-medical.com sales xp-medical.com Rev 3.1