NPC-100 and NPC-120 Series NovaSensor Disposable Medical Pressure Sensors Features Applications Solid state, high reliability Medical instrumentation Media compatibility Blood pressure measurement High performance Infusion pumps Factory filled with dielectric gel Kidney dialysis machines Small size Fully tested Temperature compensated Low cost disposable design Designed to AAMI specifications Amphenol Advanced Sensors0.415 in +0.010/-0.002 (10.54 mm +0.25/-0.05) Gel height +0.001 in/-0.035 in (+0.025 mm/-0.89 mm) 0.393 in to 0.397 in 0.392 in to 0.397 in (9.98 mm to 10.08 mm) (9.96 mm to 10.08 mm) Parameter Value Units Notes NPC-100 and NPC-120 General Pressure Ranges 30 to 300 mmHg 0.58 psi to 5.8 psi Series Specifications Overpressure 125 psi minimum The NovaSensor NPC-100 and NPC-120 Series Electrical 72F (22C) Unless Otherwise Stated pressure sensor are specifically designed for use in Input Excitation 1 to 10 VDC Calibrated for 6 VDC disposable medical applications. The devices are Dielectric Breakdown 10,000 VDC 5 compensated and calibrated per the Association for Risk Current 2 A Maximum (per AAMI), 5 the Advancement of Medical Instrumentation (AAMI) Input Impedance 1800 to 3300 guidelines for industry acceptability. The sensors Output Impedance 285 to 315 integrate a high-performance, pressure sensor die with Environmental temperature compensation circuitry and gel protection in a small, low-cost package. Temperature Compensated 15C to 40C C (59 to 104F) The NPC-100 and NPC-120 Series are manufactured Operating 15C to 40C C (59 to 104F) in a class 1000 clean-room to minimize possible Storage 25C to 70C C (13 to 158F) sources of contamination. A specially designed silicon Humidity 10 to 90 % micromachined sensing element is used to meet or Light Sensitivity 1 mmHg maximum (per AAMI BP22) exceed all industry requirements while minimizing Operating Product Life 168 hours assembly and test cost for maximum customer Shelf Life 3 years value. Thick-film laser-trimming is employed for Mechanical final compensation and calibration. Sensitivity is Weight <0.0044 lb (<2 g) maintained to 1% and linearity is better than 1% in the physiological operating pressure range. Media Interface Medical grade, dielectric gel Gel Tube Interface Polycarbonate Material The NPC-100 and NPC-120 Series are batch- manufactured in ceramic plate form and shipped as Parameter Units Minimum Type Maximum Notes an intact array or in Tape and Reel for easy customer Performance Parameters* automation. This assembly method draws from well- Offset mmHg 25 0 25 established manufacturing techniques used in the Sensitivity V/V/mmHg 4.95 5 5.05 electronics industry in order to produce a quality, high Calibration mmHg 97.5 100 102.5 2 volume product. Symmetry % 5 0.236 in to 0.246 in Linearity (30 to 100 mmHg) mmHg 1 6 0.0875 in to 0.0905 in (5.99 mm to 6.25 mm) (2.22 mm to 2.30 mm) Linearity (100 to 200 mmHg) % output 1 6 0.049 in to 0.052 in Linearity (200 to 300 mmHg) % output 1.5 6 (1.24 mm to 1.32 mm) Thermal Coefficient Offset mmHg/C 0.3 3 0.189 in to 0.192 in (4.80 mm to 4.88 mm) Thermal Coefficient Span %/C 0.1 3 Frequency Response Hz 1200 5 Radius 0.005 in Phase Shift degrees 5 5 (0.13 mm) maximum Offset Drift mmHg/8 hrs 1 4,5 0.142 in to 0.145 in (3.61 mm to 3.68 mm) * 1. All values measured at 6 VDC and 71.6F (22C) and after five second warm up unless 0.282 in to 0.290 in otherwise specified. (7.16 mm to 7.37 mm) 2. Output of sensor with no pressure applied and a 150 kW resistor shorted across Material is clear Polycarbonate. +IN to +OUT. NPC-100 Series pressure sensor port interface 3. Over a temperature range of 59F to 104F (15C to 40C). 4. Normalized offset/bridge voltage8 hours after 20 seconds. warm-up. 5. Previously qualified, not tested in production. 0.236 in to 0.246 in 0.0875 in to 0.0905 in (5.99 mm to 6.25 mm) 6. Deviation from straight line drawn through zero and 100 mmHg data points. (2.22 mm to 2.30 mm) 7. These sensors are compatible with ETO (ethylene oxide) sterilization. The user is responsible for establishing and validating suitable process conditions to prevent damage to the sensor. 0.049 in to 0.052 in (1.24 mm to 1.32 mm) 0.160 in to 0.170 in 0.320 in +0.010/-0.002 1 (4.06 mm to 4.32 mm) 0.124 in to 0.126 in (8.13 mm +0.25/-0.05) (3.14 mm to 3.20 mm) Pressure port Dielectric gel Radius 0.005 in (0.13 mm) maximum Vent hole Polycarbonate gel tube2 0.142 in to 0.145 in Contact pads (3.61 mm to 3.68 mm) 0.282 in to 0.290 in (7.16 mm to 7.37 mm) Material is clear Polycarbonate. 0.109 in to 0.119 in 0.017 in (2.77 mm to 3.02 mm) 1 NPC-120 Series pressure sensor por t interface (0.43 mm) 0.010 in (0.25 mm) 1. Tolerances reflect typical flaring during the singulation process. NPC-100 Series package diagram