XPFM200A+ Medical Power Supply
FEATURES:
Medical version of ATX200 power supply
Approved to IEC601-1 and UL2601
Includes power factor correction to IEC1000-3-2
Conducted emissions to CISPR11 class B/IEC601-1-2
marked to LVD
Remote Sense
Provided as a standard feature on V2 (+) lead of all
models. Capable of compensating for 0.25 V total of
cabling losses in output voltages.
SPECIFICATIONS:
Output Noise
0.5% RMS, 1% Pk-Pk, 20 MHz Bandwidth, differential
Ac Input
mode. Measured using a differential noise probe. Probe
90-264 Vac, 47-63 Hz single phase.
should be placed directly across the power supply output
terminals terminated with a 1uF low Z capacitor.
Input Current
Maximum input current at minimum input voltage and full
Transient Response
load is 3.5 A. Active Power Factor Correction circuitry
All outputs stay within their specified regulation limits with
assures compliance with IEC1000-3-2, Class A.
a 20% load change. i / t < 0.2 A / s.
Inrush Current
Hold-Up Time
Inrush is limited by internal thermistors. The inrush at 240
25 ms total from loss of ac input at full load until loss of
Vac, averaged over the first ac half-cycle under cold start
regulation.
conditions will not exceed 37 A. Worst case current under
hot start conditions will be 60 Amps.
Inhibit
Inhibit signal is pulled to the dc output common to inhibit
Input Protection
V1 - 5 voltages. (V = < 0.4V, sink > 1.5mA)
Internal ac fuse provided on all units. The fuse is de-
signed to blow only if a catastrophic failure occurs in the
Power Good
unit. The fuse does not blow on unsustained overload or
TTL / CMOS compatible output rises high 100 to 400 ms
short circuit.
after V2 reaches regulation and should assure that
sufficient energy is stored in the input section to provide
Efficiency
normal power fail / shutdown.
75% typical at full rated load, nominal input voltage,
depending on load distribution.
Temperature Coefficient
0.03% / C typical on all outputs.
Output Power
Continuous output power 200 W.
Turn-On Time
Less than 2 s at 120 Vac 25C
Overload Protection
Fully protected against short circuit and output overload.
Storage
Short circuit protection is cycling type power limit on V1, 2
-40 to +85 C.
& 3; thermal foldback type on V4, 5, 6. Power limit factory
set to cycle outputs off at 240 W typical. Recovery after
Leakage Current
removal of fault is automatic.
90 A under normal conditions (120 Vac @ 60 Hz).
Maximum under single-fault conditions (264 Vac @ 50
Over Voltage Protection
Hz) is 300 A.
Built in on V1, 2 and 6 with firing point set per Table 1.
OVP firing shuts down power supply.
EMI/EMC Compliance
All models include built-in EMI filtering to meet the
Voltage Adjustment
following emissions requirements:
Factory set to specified voltage. Potentiometer on V1 not
considered user accessible.
EMI SPECIFICATIONS COMPLIANCE LEVEL
Conducted Emissions EN55011, Class B; FCC Class B
No Load Turn-On/Standby
Static Discharge EN61000-4-2, 6 kV contact 8 kV air
A minimum load of 1 A on V2 is required for proper
RF Field Susceptibility EN61000-4-3, 3V/meter
regulation. If not met, no degradation of reliability will
Fast Transients / Bursts EN61000-4-4, 2 kV, 5 kHz
occur.
Surge Susceptibility EN61000-4-5, 1 kV diff., 2 kV com.
Conducted RF Susceptibility EN61000-4-6, 3V
Output Regulation
Voltage Sags & Surges EN61000-4-11
Regulation for all outputs is the maximum deviation from
initial set point under all line and load conditions. Initial
Safety Agency
set tolerance is measured with all outputs at 50 % of full
All models are certified to be in compliance with the
rated load.
applicable requirements of UL2601-1, CSA-C22.2 No.
601.1, IEC601-1/60601-1.
Condor D.C. Power Supplies, Inc., 2311 Statham Parkway, Oxnard, CA 93033
800-235-5929 805-486-4565 FAX 805-487-8911 www.condorpower.com
110Minimum Maximum Peak (E) OVP Ripple and
Model OutputNo. Output TotalRegulation Notes
Current Current Current Trip Noise
XPFM200A+1V(C)+A3.30A.1 1A4 1%6 V4 4%V 0.3 2D
2V(C)+A5.1 1A2A2 2%8 V4 6%.2 V 0.5 1
3V(A)+A12 0A.1 7A9%%5 1
4V(B)-A12 0A0A.8 0%.8%8 1
5V-A5 0A0A.5 0%.5%6 1
6V(E) +B5SA0A1-.5 -%V4 6%.2 V 0.8 2
XPFM200B+1V(C)+A2.70A.1 1A4 1%6 V4 4%V 0.3 2D
2V(C)+A5 1A2A2 2%8 V4 6%.2 V 0.5 1
3V(A)+A12 0A.1 7A9%%5 1
4V(B)-A12 0A0A.8 0%.8%8 1
5V-A5 0A0A.5 0%.5%6 1
6B(E)+A5 VS0A1-.5 -%V4 6%.2 V 0.8 2
A. To maintain regulation on V3, the +5.1 V current must be at least 1/5 of V3 and not greater than 5 times the V3 current. Required +5 V to be adjusted
to within 1% with at least a 1 A load to maintain regulation on this output.
B. Combined current of V4, 5 must not exceed 0.8 A.
C. Combined load for V1, 2 must not exceed 125 W.
D. 200 Series has external fan, for internal fan replace 200 in model number with 201.
E. V6 maximum current is 1.25 A at initial start up.
XPFM200+ MECHANICAL SPECIFICATIONS:
XPFM200 XPFM201
External Fan Internal Fan
Environmental
Operating Non-operating
Specification
TCemperature (A)0Cto 40 -40 to +85
A. Units should be allowed to warm up/operate under non-condensing conditions
before application of power. Cooling provided by internal fan heatsink
HHumidity (A)0Hto 95% R 0 to 95% R
temperatures should not exceed 90C for extended periods in the installation.
Sghock (B) 20 40 g
pk pk
B. Random vibration10 to 2000Hz, 6dB/octave roll-off from 350 to 2000Hz, 3
orthogonal axes. Tested for 10 min./axis operating and 1 hr./axis non-operating.
Atltitude-t500 to 10,000 f -500 to 40,000 f
5 g , 0.026
2 C. Shock testinghalf-sinusoidal, 10 3 ms duration, direction, 3 orthogonal axes,
rms
Vgibration (C) 1.5 , 0.003 g /Hz
2
rms
g /Hz total 6 shocks.