xx Connecting the Probe to Your Circuit Safety Summary Attachtheprobetothecircuitusingtheconnectorsandadapters Connect and Disconnect Properly. Connect the probe output shownonthebackoftheseinstructions. Selectthebestmethod to the measurement instrument before connecting the probe foryour needs, andthenproceedtoSetting upthe Probe. to the circuit under test. Disconnect the probe input and the probe ground from the circuit under test before disconnecting Setting up the Probe the probe from the measurement instrument. Press the D15-D0 button to set and view the following Observe All Terminal Ratings. To avoidre or shock hazard, parameters of each digital channel: observe all ratings and markings on the product. Consult the Threshold voltage and vertical position product manual for further ratings information before making connections to the product. Signal height (set once for all 16 channels) Channel label Do not Operate Without Covers. Do not touch exposed connections and components when power is present. Thedefaultsettingsare1.4Vthresholdswithnochannellabels. Avoid Exposed Circuitry. Do not touch exposed connections and components when power is present. DoNotOperateWithSuspectedFailures. Ifyoususpectthereis damage to this product, have it inspected by qualied service personnel. P6316 Use the B1 and B2 buttons to set and view bus characteristics DoNotOperateinWet/DampConditions. DoNotOperateinan such as: Digital Probe Explosive Atmosphere. Clock type Instructions Keep Product Surfaces Clean and Dry. Bus type (Serial or Parallel) Safety Terms and SymbolsTermsin This Manual. Bus width These terms may appear in this manual: Display format (Hex, Binary, ASCII) WARNING. Warning statements identify conditions or practices that could result in injury or loss of life. CAUTION. Caution statements identify conditions or practices that could result in damage to this product or other property. Symbols on the Product. These symbols may appear on the product: Parallel bus setup information is resident on the MSO Family of oscilloscopes. However, for other buses such as CAN and I2C, you must have the appropriate application key. See your oscilloscope manual or product data sheet for nomenclature and ordering details. Functional Check Logic activity immediately displays on all connected, active channels. If you do not see an active signal: Typical Application 1. Use the P6316 probe to view digital signals on a system 1. Press Trigger. bus. 2. Select Edge for trigger type. 2. Use analog probes, such as the P6139A or P2221 probes 3. Selectthechannelthatyouare settingupas the source. to view analog waveform information. 4. Press Autoset. 1 Ifyoudonotsee anactivesignal,tryanotherprobechannel(or *P071234500* analog probe) toverify circuitactivity atthe testpoint. 071-2345-00 Specications Product Description Table 1: Electrical and mechanical specications The P6316 digital probe connects the Tektronix MSO2000 Characteristic Description Seriesofmixed-signaloscilloscopestodigitalbusesandsignals on your target system. The probe breaks out 16 data channels Input channels 16 over two 2x8-pin headers (GROUP 1 and GROUP 2). Each Threshold voltage range 20 V header includes eight signals on one row and eight grounds Threshold accuracy 100mV+3%ofthresholdsetting along the opposite row. Minimumsignalswing 500 mV p-p A pair of interchangableying lead sets are included with the probe. The lead sets bring out one ground connection on each Maximum signal swing 20 V centered around the threshold voltage end and eight signal leads for connecting to individual test points. Maximum nondestructive 40 V p-p input signal to probe Input resistance 101 k Input capacitance 8.0 pF Minimum detectable input 5ns pulse Probe length 0.9m(3ft) Table 2: Environmental specications Characteristics Description Temperature +50 C (+122 F) Maximum operating 0C(+32F) Minimum operating 40Cto+71C Nonoperating Connecting the Probe to the Oscilloscope (40 F to +160 F) Connectthe probe as shown in the illustration below. Humidity 5% to 95% relative humidity at up to +30 C (+86 F) 1. Insertthe probe label-side up into the receptacle. 5% to 60% RH above +30 C 2. To remove the probe, squeeze the buttons onthe side and (+86 F) up to +50 C (+122 F), pull out the probe. non-condensing Altitude 3 km (9,843 ft) maximum Operating 12 km (39,370 ft) maximum Nonoperating Electrostatic Immunity The probe is not static sensitive Equipment Recycling. This product complies with the European Unions requirements according to Directive 2002/96/EC on waste electrical and electronic equipment (WEEE). For more information about recycling options, check the Support/Service section of the Tektronix Web site (www.tektronix.com). xConnecting the Probe to the Circuit Standard Accessories Contacting Tektronix Web site: www.tektronix.com The followingstandardaccessories shipwith the probe andare Warranty Information shown in the illustration above. Phone: 1-800-833-9200 Forwarrantyinformation,gotowww.tektronix.com/warranty Address: Tektronix, Inc. Item Part Department or name (if known) number Description Quantity number 14200 SW Karl Braun Drive P.O. Box 500 1 8-Channel leadset 2 ea 196-3508-XX Beaverton, OR 97077 2 Micro grabber clip 2kitsof10ea 020-2896-XX USA 3Probetip 2kitsof5ea 020-2897-XX Email: techsupport tektronix.com Instructions (English, 1 ea 071-2345-XX Japanese,&Simplied Chinese languages) CopyrightTektronix,Inc. Allrightsreserved. www.tektronix.com